Electron beam therapy at extended source-to-surface distance: a Monte Carlo investigation

被引:12
|
作者
O'Shea, Tuathan P. [1 ]
Foley, Mark J. [1 ]
Rajasekar, David [2 ]
Downes, Patrick A. [3 ]
van der Putten, Wil [2 ]
Moore, Margaret [2 ]
Shearer, Andrew [3 ]
机构
[1] Natl Univ Ireland Galway, Sch Phys, Med Phys Grp, Galway, Ireland
[2] Univ Hosp Galway, Dept Med Phys & Bioengn, Galway, Ireland
[3] Natl Univ Ireland Galway, Dept Informat Technol, Galway, Ireland
来源
关键词
electron beam; extended SSD; output factor; Monte Carlo; BEAMnrc;
D O I
10.1120/jacmp.v9i4.2811
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Electron-beam therapy is used to treat superficial tumors at a standard 100 cm source-to-surface distance (SSD). However, certain clinical situations require the use of an extended SSD. In the present study, Monte Carlo methods were used to investigate clinical electron beams, at standard and non-standard SSDs, from a Siemens Oncor Avant Garde (Siemens Healthcare, Erlangen, Germany) linear accelerator (LINAC). The LINAC treatment head was modeled in BEAMnrc for electron fields 5 cm in diameter and 10 x 10 cm, 15 x 15 cm, and 20 x 20 cm; for 6 MeV, 9 MeV, and 12 MeV; and for 100 cm, 110 cm, and 120 cm SSD. The DOSXYZnrc code was used to calculate extended SSD factors and dose contributions from various parts of the treatment head. The main effects of extended SSD on water phantom dose distributions were verified by Monte Carlo methods. Monte Carlo-calculated and measured extended SSD factors showed an average difference of +/- 1.8%. For the field 5 cm in diameter, the relative output at extended SSD declined more rapidly than it did for the larger fields. An investigation of output contributions showed this decline was mainly a result of a rapid loss of scatter dose reaching the d(max) point from the lower scrapers of the electron applicator. The field 5 cm in diameter showed a reduction in dose contributions; the larger fields generally showed an increased contribution from the scrapers with increase in SSD. Angular distributions of applicator-scattered electrons have shown a large number of acute-angle electron tracks contributing to the output for larger field sizes, explaining the shallow output reduction.
引用
收藏
页码:57 / 67
页数:11
相关论文
共 50 条
  • [1] Monte Carlo Calculations of Cerrobend to Brass Conversions for Extended Source-To-Surface Distance Total-Body Irradiation Treatments
    Li, A.
    Yang, Y. M.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [2] Verification of the monitor units calculations of the new commercial Monte Carlo based treatment planning system for electron beams at extended source-to-surface distance
    Daskalov, G
    Cygler, J
    MEDICAL PHYSICS, 2003, 30 (06) : 1518 - 1518
  • [3] Electron beam characteristics at extended source-to-surface distances for irregular cut-outs
    Arunkumar, T.
    Supe, Sanjay S.
    Ravikumar, M.
    Sathiyan, S.
    Ganesh, M.
    JOURNAL OF MEDICAL PHYSICS, 2010, 35 (04) : 207 - 214
  • [4] Moving gantry method for electron beam dose profile measurement at extended source-to-surface distances
    Fekete, Gabor
    Fodor, Emese
    Pesznyak, Csilla
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (02): : 403 - 411
  • [5] Validation of monitor unit calculations at arbitrary source-to-surface distance for the new commercial Monte Carlo based treatment planning system for electron beams
    Cygler, J
    Daskalov, G
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S42 - S43
  • [6] DOSE DISTRIBUTION CONSIDERATIONS OF MEDIUM-ENERGY ELECTRON-BEAMS AT EXTENDED SOURCE-TO-SURFACE DISTANCE
    SHAW, CB
    AYYANGAR, KM
    PAWLICKI, T
    KORB, LJ
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 32 (01): : 159 - 164
  • [7] Monte Carlo Investigation of the Mobetron to Perform Modulated Electron Beam Therapy
    Emam, I.
    Eldib, A.
    Hosini, M.
    AlSaeed, E.
    Ma, C.
    MEDICAL PHYSICS, 2014, 41 (06) : 110 - 110
  • [8] Monte Carlo simulation of electron cones used in electron beam therapy
    Williams, T.
    Ebert, M.
    Keall, E.
    Haskard, D.
    Australasian Physical and Engineering Sciences in Medicine, 1998, 21 (03): : 112 - 119
  • [9] Electron beam characteristics at extended source to surface distance for a Clinac-DHX linear accelerator
    Arunkumar, T.
    Supe, Sanjay S.
    Ravikumar, M.
    Sathiyan, S.
    Ganesh, K. M.
    Varatharaj, C.
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2009, 15 (03) : 121 - 133
  • [10] A MONTE-CARLO INVESTIGATION OF ELECTRON-BEAM APPLICATOR SCATTER
    EBERT, MA
    HOBAN, PW
    MEDICAL PHYSICS, 1995, 22 (09) : 1431 - 1435