Effect of respiration on Korotkoff sounds and oscillometric cuff pressure pulses during blood pressure measurement

被引:21
|
作者
Zheng, Dingchang [1 ]
Di Marco, Luigi Yuri [1 ]
Murray, Alan [1 ]
机构
[1] Newcastle Univ, Inst Cellular Med, Cardiovasc Phys & Engn Res Grp, Newcastle Upon Tyne NE2 4HH, Tyne & Wear, England
基金
英国工程与自然科学研究理事会;
关键词
Blood pressure measurement; Oscillometric pulses; Korotkoff sounds; Respiratory modulation; HEART-RATE; INTERVAL; OSCILLATIONS; FREQUENCY; PATTERN;
D O I
10.1007/s11517-014-1150-1
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Blood pressure (BP) measurement accuracy depends on consistent changes in Korotkoff sounds (KorS) for manual measurement and oscillometric pulses for automated measurement, yet little is known about the direct effect of respiration on these physiological signals. The aim of this research was to quantitatively assess the modulation effect of respiration on Korotkoff sounds and oscillometric pulses. Systolic and diastolic blood pressures were measured manually from 30 healthy subjects (age 41 +/- A 12 years). Three static cuff pressure conditions were studied for two respiratory rates. Cuff pressure [with oscillometric pulses (OscP)], ECG, chest motion respiration [respiration signal (Resp), from magnetometer] and Korotkoff sounds (KorS, from digital stethoscope) were recorded twice for 20 s. The physiological data were evenly resampled. Respiratory frequency was calculated from Resp (f(R)), OscP (f(O)) and KorS (f(K)) from peak spectral frequency. There was no statistically significant difference between f(R) and f(O) or f(K). Respiratory modulation was observed in all subjects. OscP amplitude modulation changed significantly between the two respiratory rates (p < 0.05) and between the three cuff pressures (p < 0.0001), and decreased significantly with decreasing cuff pressure (p < 0.05). The phase shift between Resp and modulation of OscP was statistically significant with respiratory rates (p < 0.05), but not with cuff pressures. It is accepted that BP in individuals is variable and that this relates to respiration; we now show that this respiration modulates oscillometric pulse and Korotkoff sound amplitudes from which BP is measured.
引用
收藏
页码:467 / 473
页数:7
相关论文
共 50 条
  • [1] Effect of respiration on Korotkoff sounds and oscillometric cuff pressure pulses during blood pressure measurement
    Dingchang Zheng
    Luigi Yuri Di Marco
    Alan Murray
    [J]. Medical & Biological Engineering & Computing, 2014, 52 : 467 - 473
  • [2] Respiratory modulation of oscillometric cuff pressure pulses and Korotkoff sounds during clinical blood pressure measurement in healthy adults
    Chen, Diliang
    Chen, Fei
    Murray, Alan
    Zheng, Dingchang
    [J]. BIOMEDICAL ENGINEERING ONLINE, 2016, 15
  • [3] Respiratory modulation of oscillometric cuff pressure pulses and Korotkoff sounds during clinical blood pressure measurement in healthy adults
    Diliang Chen
    Fei Chen
    Alan Murray
    Dingchang Zheng
    [J]. BioMedical Engineering OnLine, 15
  • [4] A Method for Extracting Respiratory Frequency during Blood Pressure Measurement, from Oscillometric Cuff Pressure Pulses and Korotkoff Sounds Recorded during the Measurement
    Chen, Diliang
    Chen, Fei
    Murray, Alan
    Zheng, Dingchang
    [J]. 2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2016, : 4268 - 4271
  • [5] Systolic blood pressure measurement by detecting the photoplethysmographic pulses and electronic Korotkoff-sounds during cuff deflation
    Shalom, Eran
    Hirshtal, Elad
    Slotki, Itzchak
    Shavit, Linda
    Yitzhaky, Yitzhak
    Engelberg, Shlomo
    Nitzan, Meir
    [J]. PHYSIOLOGICAL MEASUREMENT, 2020, 41 (03)
  • [6] Phase Difference between Respiration Signal and Respiratory Modulation Signal from Oscillometric Cuff Pressure Pulses during Blood Pressure Measurement
    Chen, Diliang
    Chen, Fei
    Murray, Alan
    Zheng, Dingchang
    [J]. 2016 COMPUTING IN CARDIOLOGY CONFERENCE (CINC), VOL 43, 2016, 43 : 1013 - 1016
  • [7] Blood pressure measurement using Korotkoff sounds
    Mahajan, Rutvij
    Thakurdesai, Shantanu
    Modak, Gaurav
    Jotrao, Sumit
    Bhat, Ajinkya
    [J]. 2015 INTERNATIONAL CONFERENCE ON INDUSTRIAL INSTRUMENTATION AND CONTROL (ICIC), 2015, : 635 - 639
  • [8] Novel Method of Automatic Auscultation for Blood Pressure Measurement Using Pulses in Cuff Pressure and Korotkoff Sound
    Park, D. K.
    Oh, H. S.
    Kang, J. H.
    Kim, I. Y.
    Chee, Y. J.
    Lee, J. S.
    [J]. COMPUTERS IN CARDIOLOGY 2008, VOLS 1 AND 2, 2008, : 181 - +
  • [9] SIMULTANEOUS COMPARISON OF KOROTKOFF AND OSCILLOMETRIC METHODS FOR AMBULATORY BLOOD-PRESSURE MEASUREMENT
    BRIGDEN, GS
    CASHMAN, PM
    BROADHURST, P
    LAHIRI, A
    RAFTERY, EB
    [J]. JOURNAL OF HYPERTENSION, 1989, 7 : S354 - S355
  • [10] Extracting oscillometric pulses from the cuff pressure: does it affect the pressures determined by oscillometric blood pressure monitors?
    Amoore, John N.