Laboratory tests for measurement of von Willebrand factor show poor agreement among different centers: Results from the United Kingdom National External Quality Assessment Scheme for Blood Coagulation

被引:83
|
作者
Kitchen, Steve [1 ]
Jennings, Ion [1 ]
Woods, Tim A. L. [1 ]
Kitchen, Dianne P. [1 ]
Walker, Isobel D. [1 ]
Preston, Francis E. [1 ]
机构
[1] United Kingdom Natl External Qual Assessment Sche, Sheffield, S Yorkshire, England
来源
SEMINARS IN THROMBOSIS AND HEMOSTASIS | 2006年 / 32卷 / 05期
关键词
von Willebrand factor : antigen (VWF : Ag); von Willebrand factor : ristocetin cofactor activity (vWF : RCo); laboratory assays; von Willebrand disease (vWD);
D O I
10.1055/s-2006-947863
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In recognition of the importance of von Willebrand factor (vWF) testing in the diagnosis of von Willebrand disease (vWD), the United Kingdom National External Quality Assessment Scheme for Blood Coagulation regularly distributes samples for determination of v W:antigen (vWF:Ag). Data from 10 separate surveys performed between 2001 and 2005 are reviewed. These include results from similar to 200 different centers, of which 55% are within the United Kingdom and the remainder are from other countries. During the period of the surveys, the use of immunoelectrophoresis for determination of vWF:Ag practically disappeared and was largely replaced by latex agglutination assays. The coefficient of variation (CV) of results in different centers was similar to 15 to 20% for most vWF:Ag techniques, with CVs of similar to 7% for a fluorescence-based assay. Several different techniques were used for determination of vWF ristocetin cofactor activity (vWF:RCo), all of which were associated with poor agreement among centers as indicated by CVs of 40 to 50%. Several centers calculated the ratio of vWF:Ag/vWF:RCo but with variable success. Ratios compatible with either type 1 or type 2 vWD were obtained on samples from subjects with type 1 vWD, as well as on samples from subjects with genetically confirmed type 2 vWD. Overall, our data show that laboratory testing for vWD remains problematic. It remains to be seen whether newer techniques will offer consistently improved precision.
引用
收藏
页码:492 / 498
页数:7
相关论文
共 12 条
  • [1] External Quality Assessment Data for Investigation of von Willebrand Disease: Focus on Relative Utility of Contemporary Functional von Willebrand Factor Assays. The United Kingdom National External Quality Assessment Scheme (UK NEQAS) Experience
    Jennings, Ian
    Reilly-Stitt, Chris
    Lowe, Anna
    Kitchen, Steve
    Walker, Isobel
    SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2022, 48 (06): : 732 - 738
  • [2] Multilaboratory testing in thrombophilia through the United Kingdom National External Quality Assessment Scheme (blood coagulation) Quality Assurance Program
    Jennings, I
    Kitchen, S
    Woods, TAL
    Preston, FE
    SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2005, 31 (01): : 66 - 72
  • [3] ERRORS IN BACTERIOLOGICAL TECHNIQUES - RESULTS FROM THE UNITED-KINGDOM NATIONAL EXTERNAL QUALITY ASSESSMENT SCHEME FOR MICROBIOLOGY
    SNELL, JJS
    DEMELLO, JV
    PHUA, TJ
    MEDICAL LABORATORY SCIENCES, 1986, 43 (04): : 344 - 355
  • [4] The United Kingdom National External Quality Assessment Scheme (blood transfusion laboratory practice): trends in proficiency and practice between 1985 and 2000
    Knowles, SM
    Milkins, CE
    Chapman, JF
    Scott, M
    TRANSFUSION MEDICINE, 2002, 12 (01) : 11 - 23
  • [5] Pre-analytical variables in haemostasis: Findings from the United Kingdom National External Quality Assessment scheme for Blood Coagulation (UK NEQAS BC) haemolysis exercise
    Brown, Lilia
    Jennings, Ian
    Kitchen, Steve
    Kitchen, Dianne P.
    Woods, Tim A. L.
    Walker, Isobel D.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2021, 43 (05) : 1198 - 1206
  • [6] Local calibration of International Normalised Ratio improves between laboratory agreement: Results from the UK National External Quality Assessment Scheme
    Kitchen, S
    Jennings, I
    Woods, TAL
    Preston, FE
    THROMBOSIS AND HAEMOSTASIS, 1999, 81 (01) : 60 - 65
  • [7] Laboratory international normalized ratios determined with commercial thromboplastins in >450 centers before and after the establishment of the International Standard for human thromboplastin-rTF 16: data from United Kingdom National External Quality Assessment Scheme for Blood Coagulation
    Kitchen, Steve
    Jennings, Ian
    Kitchen, Dianne P.
    Reilly-Stitt, Christopher
    Walker, Isobel D.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2023, 21 (05) : 1385 - 1387
  • [8] COMPARISON OF RESULTS FROM 2 ANTIBIOTIC SUSCEPTIBILITY TESTING TRIALS THAT FORMED PART OF THE UNITED-KINGDOM NATIONAL EXTERNAL QUALITY ASSESSMENT SCHEME
    SNELL, JJS
    BROWN, DFJ
    GARDNER, PS
    JOURNAL OF CLINICAL PATHOLOGY, 1984, 37 (03) : 321 - 328
  • [9] Interlaboratory Variation in Factor VIII:C Inhibitor Assay Results Is Sufficient to Influence Patient Management: Data from the UK National Quality External Assessment Scheme for Blood Coagulation
    Kitchen, Steve
    Jennings, Ian
    Preston, F. Eric
    Kitchen, Dianne P.
    Woods, Tim A. L.
    Walker, Isobel D.
    SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2009, 35 (08): : 778 - 785
  • [10] Potentially clinically important inaccuracies in testing for the lupus anticoagulant: an analysis of results from three surveys of the UK National External Quality Assessment Scheme (NEQAS) for blood coagulation
    Jennings, I
    Kitchen, S
    Woods, TAL
    Preston, FE
    Greaves, M