Evaluation of a high resolution positron emission tomography (PET) scanner for positron emission mammography (PEM)

被引:0
|
作者
White, LL [1 ]
Ott, RJ [1 ]
Carnochan, P [1 ]
Duxbury, DM [1 ]
Bruyndonckx, P [1 ]
Tavernier, S [1 ]
机构
[1] Inst Canc Res, Sutton, Surrey, England
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A small animal scanner (VUB-PET), built at the Vrije Universiteit, Brussels has been used to test the feasibility of building a dedicated positron emission mammography (PEM) scanner. VUB-PET is not ideal for breast scanning due to a small field of view (FOV) and an inactive region before the active detector. It does however have a resolution of 3-4mm and high sensitivity, properties needed in a PEM scanner to detect small tumours when using F-18 fluorodeoxyglucose (FDG). A cylindrical, water-filled phantom (69mm diameter by 65mm long) was used to simulate a small breast and small spherical inserts simulated tumours. Various parameters were investigated including 'tumour' size and position in phantom, counts acquired, activity in the phantom and tumour to background ratio (TBR). Background activity was added in some studies, to simulate the uptake of FDG in the heart by attaching Na-22 sources to the back of the phantom. Region of interest (ROI) analysis was used to determine the mean counts in the 'tumour' and the cylinder and the ratio of the means (ROM), was determined The larger the ROM of an image, the more likely the 'tumour' is to be detected The main conclusions of these studies are that the 'tumour'size and the TBR are the main contributors to the visibility of the 'tumour. The level of activity in the breast seems to have little effect on the detection of the 'tumour, due to the small randoms rate of the scanner. This data will be used to help in the development of a clinical PEM scanner. The results indicate the activity in the heart should not be a problem if the detectors resolving time is kept short. Additionally, the activity injected may be less than for a normal PET scan reducing the dose to the patient. However it seems unlikely that small 'tumours' (<5-6mm) will be visible unless the TBR is very high (>10).
引用
收藏
页码:1236 / 1240
页数:5
相关论文
共 50 条
  • [1] Design and evaluation of the Clear-PEM scanner for positron emission mammography
    Abreu, MC
    Aguiar, JD
    Almeida, FG
    Almeida, P
    Bento, P
    Carrico, B
    Ferreira, M
    Ferreira, NC
    Gonçalves, F
    Leong, C
    Lopes, F
    Lousa, P
    Martins, MV
    Matela, N
    Mendes, PR
    Moura, R
    Nobre, J
    Oliveira, N
    Ortigao, C
    Peralta, L
    Pereira, R
    Rego, J
    Ribeiro, R
    Rodrigues, P
    Sampaio, J
    Santos, AI
    Silva, L
    Silva, JC
    Sousa, P
    Teixeira, IC
    Teixeira, JP
    Trindade, A
    Varela, J
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) : 71 - 77
  • [2] Positron emission mammography (PEM)
    Garcia-Gomez, Jaime I.
    Villasenor-Navarro, Yolanda
    [J]. GACETA MEXICANA DE ONCOLOGIA, 2012, 11 (04): : 281 - 287
  • [3] Positron emission tomography (PET) and mammography (PEM) for breast cancer: Importance to surgeons
    Tafra, Lorraine
    [J]. ANNALS OF SURGICAL ONCOLOGY, 2007, 14 (01) : 3 - 13
  • [4] Positron Emission Tomography (PET) and Mammography (PEM) for Breast Cancer: Importance to Surgeons
    Lorraine Tafra
    [J]. Annals of Surgical Oncology, 2007, 14 : 3 - 13
  • [5] AN ULTRA-HIGH RESOLUTION PRECLINICAL POSITRON EMISSION TOMOGRAPHY SCANNER
    Martins, P.
    Blanco, A.
    arques, Crespo P.
    Marques, M. F. F.
    Marques, R. F.
    Gordo, P. M.
    Kajetanowicz, M.
    Korcyl, G.
    Lopes, L.
    Michel, J.
    Palka, M.
    Traxler, M.
    Abrunhosa, A. J.
    Couceiro, M.
    Fonte, P.
    [J]. 2015 IEEE 4TH PORTUGUESE MEETING ON BIOENGINEERING (ENBENG), 2015,
  • [6] Novel high resolution detectors for Positron Emission Tomography (PET)
    Belcari, N
    Camarda, M
    Del Guerra, A
    Herbert, DJ
    Vaiano, A
    Di Domenico, G
    Zavattini, G
    [J]. NUCLEAR PHYSICS B-PROCEEDINGS SUPPLEMENTS, 2003, 125 : 48 - 52
  • [7] Development of a Commercial Positron Emission Mammography (PEM)
    Miyake, Masayasu
    Yamamoto, Seiichi
    Itoh, Masatoshi
    Kumagai, Kazuaki
    Sasaki, Takehisa
    dos Santos, Targino Rodrigues
    Tashiro, Manabu
    Baba, Mamoru
    [J]. 13TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING, VOLS 1-3, 2009, 23 (1-3): : 2248 - +
  • [8] Numerical Evaluation of a Positron Emission Mammography Scanner using GATE
    Fernando Torres-Urzua, Luis
    Alva-Sanchez, Hector
    Martinez-Davalos, Arnulfo
    Rodriguez-Villafuerte, Mercedes
    [J]. MEDICAL PHYSICS: Fourteenth Mexican Symposium on Medical Physics, 2016, 1747
  • [9] Imaging performance of PEM-1: A high resolution system for positron emission mammography
    Thompson, CJ
    Murthy, K
    Clancy, RL
    Robar, JL
    Bergman, A
    Lisbona, R
    Loutfi, A
    Gagnon, JH
    Weinberg, IN
    Mako, R
    [J]. 1995 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD, VOLS 1-3, 1996, : 1074 - 1078
  • [10] Positron Emission Tomography (PET)
    Oltmanns, Doerte
    [J]. CHEMIE IN UNSERER ZEIT, 2015, 49 (01) : 7 - 7