An Implantable Sensor for Arterial Pressure Monitoring with Minimal Loading: Design and Finite Element Validation

被引:0
|
作者
Beyaz, Mustafa Ilker [1 ]
机构
[1] Antalya Bilim Univ, Dept Elect & Elect Engn, Antalya, Turkey
来源
关键词
pressure sensor; in-vivo; artery pressure; flexible sensor; MEMS;
D O I
10.1109/SENSORS52175.2022.9967260
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents an in-vivo flexible pressure sensor that can be implanted around an artery for continuous measurement of blood pressure. Currently, in-vivo arterial sensor architectures rely on circular geometries fully encasing the artery perimeter, leading to significant reduction in blood pressure and flow rate inside the artery. The novelty of the sensor presented here lies in its geometrical structure and application that minimizes such disadvantages. The design consists of two arc-shaped flexible piezoelectric sensors located across each other and fixed together using medical-grade sutures on an artery periphery. Radial expansion of the artery during blood pulse wave is converted to an electrical signal through piezoelectric transduction. Finite element simulations were performed on laminar flow and solid mechanics domains to evaluate the effect of sensor presence on blood pressure and flow. Compared to previously reported sensor designs, it has been demonstrated that the pressure and flow values can be raised from 95% to 98% and from 66% to 88%, respectively. The improvement in flow rate is specifically important and vital for heart- and organ implantation-related medical operations.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] A minimal invasive implantable sensor for telemetric intraocular pressure monitoring
    Mariacher, Siegfried
    Ebner, Martina
    Januschowski, Kai
    Szurman, Peter
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (07)
  • [2] Implantable low power integrated pressure sensor system for minimal invasive telemetric patient monitoring
    Hierold, C
    Clasbrummel, B
    Behrend, D
    Scheiter, T
    Steger, M
    Oppermann, K
    Kapels, H
    Landgraf, E
    Wenzel, D
    Etzrodt, D
    MICRO ELECTRO MECHANICAL SYSTEMS - IEEE ELEVENTH ANNUAL INTERNATIONAL WORKSHOP PROCEEDINGS, 1998, : 568 - 573
  • [3] Development and Validation of a Dynamic Abdominal Pressure Twin Sensor Finite Element Model
    Yang, Peiyu
    Katangoori, Divya
    Noll, Scott
    Stammen, Jason
    Suntay, Brian
    Carlson, Michael
    Moorhouse, Kevin
    SAE INTERNATIONAL JOURNAL OF TRANSPORTATION SAFETY, 2023, 11 (03) : 345 - 355
  • [4] OPTIMAL DESIGN AND EXPERIMENTAL VALIDATION OF A NEW NO-CUFF BLOOD PRESSURE SENSOR BASED ON A NEW FINITE ELEMENT MODEL
    Tu, Tse-Yi
    Chao, Paul C. -P.
    Lee, Yung-Pin
    Kao, Yung-Hua
    PROCEEDINGS OF THE ASME INTERNATIONAL DESIGN ENGINEERING TECHNICAL CONFERENCES AND COMPUTERS AND INFORMATION IN ENGINEERING CONFERENCE, 2014, VOL 8, 2014,
  • [5] Design and fabrication of implantable pressure sensing resistor sensor for human bladder monitoring system
    Jonghyun Kim
    Bumkyoo Choi
    Microsystem Technologies, 2011, 17 : 1453 - 1458
  • [6] Design and fabrication of implantable pressure sensing resistor sensor for human bladder monitoring system
    Kim, Jonghyun
    Choi, Bumkyoo
    MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 2011, 17 (09): : 1453 - 1458
  • [7] Monitoring Pulmonary Arterial Hypertension Using an Implantable Hemodynamic Sensor
    Benza, Raymond L.
    Doyle, Mark
    Lasorda, David
    Parikh, Kishan S.
    Correa-Jaque, Priscilla
    Badie, Nima
    Ginn, Greg
    Airhart, Sophia
    Franco, Veronica
    Kanwar, Manreet K.
    Murali, Srinivas
    Raina, Amresh
    Agarwal, Rahul
    Rajagopal, Sudarshan
    White, Jason
    Biederman, Robert
    CHEST, 2019, 156 (06) : 1176 - 1186
  • [8] Subcutaneous blood pressure monitoring with an implantable optical sensor
    Michael Theodor
    Dominic Ruh
    Jens Fiala
    Katharina Förster
    Claudia Heilmann
    Yiannos Manoli
    Friedhelm Beyersdorf
    Hans Zappe
    Andreas Seifert
    Biomedical Microdevices, 2013, 15 : 811 - 820
  • [9] Novel implantable pressure and acceleration sensor for bladder monitoring
    Soebadi, Mohammad Ayodhia
    Weydts, Tristan
    Brancato, Luigi
    Hakim, Lukman
    Puers, Robert
    De Ridder, Dirk
    INTERNATIONAL JOURNAL OF UROLOGY, 2020, 27 (06) : 543 - 550
  • [10] Subcutaneous blood pressure monitoring with an implantable optical sensor
    Theodor, Michael
    Ruh, Dominic
    Fiala, Jens
    Foerster, Katharina
    Heilmann, Claudia
    Manoli, Yiannos
    Beyersdorf, Friedhelm
    Zappe, Hans
    Seifert, Andreas