Ultrasound fusion image error correction using subject-specific liver motion model and automatic image registration

被引:7
|
作者
Yang, Minglei [1 ,3 ]
Ding, Hui [1 ,3 ]
Zhu, Lei [2 ,4 ]
Wang, Guangzhi [1 ,3 ]
机构
[1] Tsinghua Univ, Sch Med, Dept Biomed Engn, Beijing 100084, Peoples R China
[2] Beijing Shen Mindray Med Elect Technol Res Inst C, Beijing 100085, Peoples R China
[3] Tsinghua Univ, Sch Med, Room C249, Beijing 100084, Peoples R China
[4] 3F Bldg5,8 Chuangye Rd, Beijing 100085, Peoples R China
基金
中国国家自然科学基金;
关键词
Ultrasound fusion imaging; Respiration-induced fusion error; Subject-specific motion model; Automatic image registration; Liver;
D O I
10.1016/j.compbiomed.2016.10.008
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: Ultrasound fusion imaging is an emerging tool and benefits a variety of clinical applications, such as image-guided diagnosis and treatment of hepatocellular carcinoma and unresectable liver metastases. However, respiratory liver motion-induced misalignment of multimodal images (i.e., fusion error) compromises the effectiveness and practicability of this method. The purpose of this paper is to develop a subject-specific liver motion model and automatic registration-based method to correct the fusion error. Methods: An online-built subject-specific motion model and automatic image registration method for 2D ultrasound-3D magnetic resonance (MR) images were combined to compensate for the respiratory liver motion. The key steps included: 1) Build a subject-specific liver motion model for current subject online and perform the initial registration of pre-acquired 3D MR and intra-operative ultrasound images; 2) During fusion imaging, compensate for liver motion first using the motion model, and then using an automatic registration method to further correct the respiratory fusion error. Evaluation experiments were conducted on liver phantom and five subjects. Results: In the phantom study, the fusion error (superior-inferior axis) was reduced from 13.90 +/- 2.38 mm to 4.26 +/- 0.78 mm by using the motion model only. The fusion error further decreased to 0.63 +/- 0.53 mm by using the registration method. The registration method also decreased the rotation error from 7.06 +/- 0.21 degrees to 1.18 +/- 0.66. In the clinical study, the fusion error was reduced from 12.90 +/- 9.58 nun to 6.12 +/- 2.90 mm by using the motion model alone. Moreover, the fusion error decreased to 1.96 +/- 0.33 mm by using the registration method. Conclusions: The proposed method can effectively correct the respiration-induced fusion error to improve the fusion image quality. This method can also reduce the error correction dependency on the initial registration of ultrasound and MR images. Overall, the proposed method can improve the clinical practicability of ultrasound fusion imaging.
引用
下载
收藏
页码:99 / 109
页数:11
相关论文
共 50 条
  • [1] Automatic motion correction in dynamic radionuclide renography using image registration
    Lee, KJ
    Barber, DC
    NUCLEAR MEDICINE COMMUNICATIONS, 1998, 19 (12) : 1159 - 1167
  • [2] Automated subject-specific, hexahedral mesh generation via image registration
    Ji, Songbai
    Ford, James C.
    Greenwald, Richard M.
    Beckwith, Jonathan G.
    Paulsen, Keith D.
    Flashman, Laura A.
    McAllister, Thomas W.
    FINITE ELEMENTS IN ANALYSIS AND DESIGN, 2011, 47 (10) : 1178 - 1185
  • [3] Population-based prediction of subject-specific prostate deformation for MR-to-ultrasound image registration
    Hu, Yipeng
    Gibson, Eli
    Ahmed, Hashim Uddin
    Moore, Caroline M.
    Emberton, Mark
    Barratt, Dean C.
    MEDICAL IMAGE ANALYSIS, 2015, 26 (01) : 332 - 344
  • [4] A subject-specific technique for respiratory motion correction in image-guided cardiac catheterisation procedures
    King, A. P.
    Boubertakh, R.
    Rhode, K. S.
    Ma, Y. L.
    Chinchapatnam, P.
    Gao, G.
    Tangcharoen, T.
    Ginks, M.
    Cooklin, M.
    Gill, J. S.
    Hawkes, D. J.
    Razavi, R. S.
    Schaeffter, T.
    MEDICAL IMAGE ANALYSIS, 2009, 13 (03) : 419 - 431
  • [5] Error Correction in Image Registration using POCS
    Duraisamy, Prakash
    Alam, Mohammed S.
    Jackson, Stephen C.
    OPTICAL PATTERN RECOGNITION XXII, 2011, 8055
  • [6] Motion Correction Using Subpixel Image Registration
    HajiRassouliha, Amir
    Taberner, Andrew J.
    Nash, Martyn P.
    Nielsen, Poul M. F.
    RECONSTRUCTION, SEGMENTATION, AND ANALYSIS OF MEDICAL IMAGES, 2017, 10129 : 14 - 23
  • [7] Automatic error correction using adaptive weighting for vessel-based deformable image registration
    Ren J.
    Green M.
    Huang X.
    Abdalbari A.
    Biomedical Engineering Letters, 2017, 7 (2) : 173 - 181
  • [8] Respiratory motion correction for liver contrast-enhanced ultrasound by automatic selection of a reference image
    Zhang, Ji
    Zhang, Yanrong
    Chen, Juan
    Ling, Gonghao
    Wang, Xiangyu
    Xu, Haibo
    MEDICAL PHYSICS, 2019, 46 (11) : 4992 - 5001
  • [9] Subject-specific and respiration-corrected 4D liver model from real-time ultrasound image sequences
    Li, Cheng
    Wu, Jiaze
    Gogna, Apoorva
    Tan, Bien Soo
    Teo, Jin Yao
    Ooi, London Lucien
    Liu, Jimin
    Yu, Haoyong
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING-IMAGING AND VISUALIZATION, 2018, 6 (01): : 7 - 16
  • [10] Subject-specific four-dimensional liver motion modeling based on registration of dynamic MRI
    Noorda, Yolanda H.
    Bartels, Lambertus W.
    Viergever, Max A.
    Pluim, Josien P. W.
    JOURNAL OF MEDICAL IMAGING, 2016, 3 (01)