Towards optimization of probe placement for radio-frequency ablation

被引:0
|
作者
Altrogge, Inga [1 ]
Kroeger, Tim
Preusser, Tobias
Bueskens, Christof
Pereira, Philippe L.
Schmidt, Diethard
Weihusen, Andreas
Peitgen, Heinz-Otto
机构
[1] Univ Bremen, Ctr Complex Syst & Visualizat, CeVis, D-2800 Bremen 33, Germany
[2] Univ Bremen, Ctr Ind Math, ZETEM, D-2800 Bremen, Germany
[3] Univ Tubingen, Dept Diagnost Radiol, Tubingen, Germany
[4] Ctr Med Diagnost Syst & Visualizat, MeVis, Bremen, Germany
关键词
D O I
暂无
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We present a model for the optimal placement of mono- and bipolar probes in radio-frequency (RF) ablation. The model is based on a numerical computation of the probe's electric potential and of the steady state of the heat distribution during RF ablation. The optimization is performed by minimizing a temperature based objective functional under these constraining equations. The paper discusses the discretization and implementation of the approach. Finally, applications of the optimization to artificial data and a comparison to a real RF ablation are presented.
引用
收藏
页码:486 / 493
页数:8
相关论文
共 50 条
  • [1] Role of Radio-Frequency Ablation
    de Baere, T.
    [J]. EUROPEAN JOURNAL OF CANCER, 2011, 47 : S88 - S89
  • [2] Finite-element analysis of hepatic multiple probe radio-frequency ablation
    Haemmerich, D
    Tungjitkusolmun, S
    Staelin, ST
    Lee, FT
    Mahvi, DM
    Webster, JG
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (08) : 836 - 842
  • [3] Design and Optimization of Radio-Frequency (RF) Ablation Probes for Pain Therapy
    Sharma, Shashwat
    Ludwig, Alon
    Sarris, Costas D.
    [J]. 2015 USNC-URSI RADIO SCIENCE MEETING (JOINT WITH AP-S SYMPOSIUM) PROCEEDINGS, 2015, : 328 - 328
  • [4] Radio-frequency endometrial ablation for menorrhagia
    French, L
    [J]. AMERICAN FAMILY PHYSICIAN, 2005, 71 (03) : 557 - 558
  • [5] Complications of radio-frequency thermal ablation
    Francica, G
    [J]. RADIOLOGY, 2001, 220 (02) : 554 - 554
  • [6] Tumor ablation with radio-frequency energy
    Gazelle, GS
    Goldberg, SN
    Solbiati, L
    Livraghi, T
    [J]. RADIOLOGY, 2000, 217 (03) : 633 - 646
  • [7] PLASMA RESONANCE IN A RADIO-FREQUENCY PROBE
    TAKAYAMA, K
    IKEGAMI, H
    MIYAZAKI, S
    [J]. PHYSICAL REVIEW LETTERS, 1960, 5 (06) : 238 - 240
  • [8] A tunable radio-frequency magnetic probe
    Sun, B.
    Yuan, G. Y.
    Huo, W. G.
    Ding, Z. F.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (05):
  • [9] Placement optimization of integrated circuits for reduced radio-frequency interferences in mobile devices
    Song, Eakhwan
    Park, Hark-Byeong
    Hwang, Chulsoon
    Park, Hyun Ho
    [J]. MICROWAVE AND OPTICAL TECHNOLOGY LETTERS, 2016, 58 (01) : 31 - 37
  • [10] Comparison Temperature Distribution between Microwave and Radio-Frequency Ablation Probe in Hepatic Cancer
    Chaichanyut, M.
    Tungjitkusolmun, S.
    Potejanasaja, I.
    Lertprasert, P.
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 4187 - +