Long-term structural failure of coaxial polyurethane implantable cardioverter defibrillator leads

被引:36
|
作者
Hauser, RG
Cannom, D
Hayes, DL
Parsonnet, V
Hayes, J
Ratliff, N
Tyers, GFO
Epstein, AE
Vlay, SC
Furman, S
Gross, J
机构
[1] Minneapolis Heart Inst Fdn, Minneapolis, MN USA
[2] Good Samaritan Hosp, Los Angeles, CA USA
[3] Mayo Clin, Rochester, MN USA
[4] Newark Beth Israel, Newark, NJ USA
[5] Marshfield Clin Fdn Med Res & Educ, Marshfield, WI USA
[6] Univ British Columbia, Vancouver, BC V5Z 1M9, Canada
[7] Univ Alabama Birmingham, Birmingham, AL USA
[8] Univ Hosp Stony Brook, Stony Brook, NY USA
[9] Montefiore Med Ctr, Bronx, NY 10467 USA
来源
关键词
implantable cardioverter defibrillator; leads; complications; registry;
D O I
10.1046/j.1460-9592.2002.t01-1-00879.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Transvene models 693616966, a coaxial polyurethane ICD lead, may be prone to structural failure. These models comprise 54% of ICD lead failures in the authors' Multicenter Registry database. Because ICD leads perform a vital function, the clinical features, causes, and probability of Transvene 693616966 lead failure were determined. The Registry and United States Food and Drug Administration databases were queried for the clinical features and structural causes of the Transvene 693616966 lead failure, and a five-center substudy estimated the survival probability for 521 Transvene 6936/6966 implants. The mean time to failure was 4.8 +/- 2.1 years, and the estimated survival at 60 and 84 months after implant were 92% and 84%, respectively. Oversensing was the most common sign of failure (76%), and 24 patients experienced inappropriate shocks. The manufacturer's reports indicated that high voltage coil fracture and 80A polyurethane defects were the predominant causes of lead failure. Transvene models 6936 and 6966 coaxial polyurethane ICD leads are prone to failure over time. Patients who have these leads should be evaluated frequently, Additional studies are needed to identify safe management strategies.
引用
收藏
页码:879 / 882
页数:4
相关论文
共 50 条
  • [1] Long-term Outcome of Implantable Cardioverter/Defibrillator Lead Failure
    Davidsson, Gustav Arnar
    Jonsdottir, Gudlaug M.
    Oddsson, Hjortur
    Lund, Sigrun H.
    Arnar, David O.
    [J]. JAMA INTERNAL MEDICINE, 2020, 180 (02) : 322 - 324
  • [2] Long-Term Evaluation of Biotronik Linox and Linoxsmart Implantable Cardioverter Defibrillator Leads
    Good, Eric D.
    Cakulev, Ivan
    Orlov, Michael V.
    Hirsh, David
    Simeles, John
    Mohr, Kelly
    Moll, Phil
    Bloom, Heather
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2016, 27 (06) : 735 - 742
  • [3] Prospective long-term follow-up of silicone-polyurethane-insulated implantable cardioverter-defibrillator leads
    Cairns, John A.
    Healey, Jeff S.
    Epstein, Andrew E.
    Themeles, Ellison
    Balasubramanian, Kumar
    Connolly, Stuart J.
    [J]. HEART RHYTHM O2, 2022, 3 (01): : 57 - 64
  • [4] Impact of fracture-prone implantable cardioverter defibrillator leads on long-term patient mortality
    Koike, Toshiharu
    Shoda, Morio
    Ejima, Koichiro
    Yagishita, Daigo
    Suzuki, Atsushi
    Hasegawa, Shun
    Kataoka, Shohei
    Yazaki, Kyoichiro
    Higuchi, Satoshi
    Kanai, Miwa
    Yamaguchi, Junichi
    [J]. JOURNAL OF ARRHYTHMIA, 2023, 39 (03) : 454 - 463
  • [5] Risk of Failure of Transvenous Implantable Cardioverter Defibrillator Leads
    Borleffs, C. Jan Willem
    van Erven, Lieselot
    van Bommel, Rutger J.
    van der Velde, Enno T.
    van der Wall, Ernst E.
    Bax, Jeroen J.
    Rosendaal, Frits R.
    Schalij, Martin J.
    [J]. CIRCULATION, 2008, 118 (18) : S926 - S926
  • [6] LONG-TERM OUTCOME WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    WINKLE, RA
    MEAD, RH
    RUDER, MA
    GAUDIANI, VA
    SMITH, NA
    BUCH, WS
    SCHMIDT, P
    SHIPMAN, T
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (06) : 1353 - 1361
  • [7] LONG-TERM COMPLICATIONS OF IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LEAD SYSTEMS
    ALMASSI, GH
    OLINGER, GN
    WETHERBEE, JN
    FEHL, G
    [J]. ANNALS OF THORACIC SURGERY, 1993, 55 (04): : 888 - 892
  • [8] Long-term evaluation of a novel extravascular implantable cardioverter defibrillator
    Gebran, Anthony
    Refaat, Marwan M.
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2021, 32 (09) : 2379 - 2380
  • [9] In Vivo Durability of Polyurethane Insulated Implantable Cardioverter Defibrillator (ICD) Leads
    Salih, Anmar
    Goswami, Tarun
    [J]. POLYMERS, 2024, 16 (12)
  • [10] Increased failure rate of a polyurethane implantable cardioverter defibrillator lead
    Maisel, WH