Urea Biosensor Development using Immobilized Urease and Light Dependent Resistor

被引:0
|
作者
Fatoni, Amin [1 ,2 ]
Zusfahair [1 ]
Nurfiah, Siti [1 ]
Anggraeni, Mekar Dwi [3 ]
Aziz, Abdullah Nur [4 ]
机构
[1] Univ Jenderal Soedirman, Fac Math & Nat Sci, Dept Chem, Purwokerto 53123, Jawa Tengah, Indonesia
[2] Univ Jenderal Soedirman, Ctr Maritime Biosci Studies, Purwokerto 53123, Jawa Tengah, Indonesia
[3] Univ Jenderal Soedirman, Fac Hlth Sci, Dept Nursing, Purwokerto 53123, Jawa Tengah, Indonesia
[4] Univ Jenderal Soedirman, Fac Math & Nat Sci, Dept Phys, Purwokerto 53123, Jawa Tengah, Indonesia
关键词
D O I
10.1063/1.5097490
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The existence of urea in human's body becomes an important biological parameter for kidney disease identification. Development of urea analysis methods are widely studied, including the using a biosensor. This study was developed a urea biosensor using immobilized urease in alginate matrix and a light dependent (LDR) as continuous color change detector. The color detection was based in the reaction of urea with urease to produce ammonia which finally identified using bromothymol blue as color indicator. The LDR, a cheap electronic component was controlled by Arduino microcontroller to monitor the color change related to urea concentration. The result showed the optimum reaction time was 4 minutes with the best linearity was obtained by red color change with urea concentration (y = 0.42x + 39.34, R-2 = 0.991). Furthermore, the limit of detection of this developed urea biosensor was 0.35 mM and the limit of quantification was 1.17 mM. The linear range of the urea biosensor was 1.17 mM to 42.65 mM. The immobilized urease in alginate bead could be used up to 11 times. Urea detection in blood samples showed no significant differences between the results of the clinical analysis with the results of developed urea biosensor with a correction value of 1.3.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Urease immobilized potentiometric biosensor for determination of urea
    Sihombing, Kawan
    Tamba, Monica Chrisdayanti
    Marbun, Winry Sulistiani
    Situmorang, Manihar
    INDIAN JOURNAL OF CHEMISTRY SECTION A-INORGANIC BIO-INORGANIC PHYSICAL THEORETICAL & ANALYTICAL CHEMISTRY, 2018, 57 (02): : 175 - 180
  • [2] Development of a urea potentiometric biosensor based on gelatin-immobilized urease
    Panpae, K
    Krintrakul, S
    Chaiyasit, A
    Proceedings of 44th Kasetsart University Annual Conference: AGRICULTURAL SCIENCE: CARRYING FORWARD THE ROYAL BIO-ENERGY INITIATIVE, 2006, : 166 - 174
  • [3] Continuous analysis of urea in blood by immobilized urease using a conductive biosensor
    Shu, HC
    Wu, NP
    JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 2000, 47 (4B) : 987 - 991
  • [4] Potentiometric biosensor for determination of urea in milk using immobilized Arthrobacter creatinolyticus urease
    Ramesh, Rajendran
    Puhazhendi, Puhazhselvan
    Kumar, Jitendra
    Gowthaman, Marichetti Kuppuswami
    D'Souza, Stanislaus Francis
    Kamini, Numbi Ramudu
    MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2015, 49 : 786 - 792
  • [5] A novel urea conductometric biosensor based on zeolite immobilized urease
    Kirdeciler, Salih Kaan
    Soy, Esin
    Ozturk, Seckin
    Kucherenko, Ivan
    Soldatkin, Oleksandr
    Dzyadevych, Sergei
    Akata, Burcu
    TALANTA, 2011, 85 (03) : 1435 - 1441
  • [6] Urea potentiometric biosensor based on urease immobilized on chitosan membranes
    Magalhaes, JMCS
    Machado, AASC
    TALANTA, 1998, 47 (01) : 183 - 191
  • [7] Potentiometric Urea Biosensor Based on a Urease-Immobilized Polypyrrole
    Prissanaroon-Ouajai, Walaiporn
    Sirivat, Anuvat
    Pigram, Paul J.
    Brack, Narelle
    MACROMOLECULAR SYMPOSIA, 2015, 354 (01) : 334 - 339
  • [8] Fabrication of an amperometric urea biosensor using urease and metal catalysts immobilized by a polyion complex
    Hao, Wu
    Das, Gautam
    Yoon, Hyon Hee
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2015, 747 : 143 - 148
  • [9] Development of Amperometric Biosensor Immobilized by Entrapment of Urease Enzyme in Polypiroll Film for the Determination of Blood Urea
    Busono, Pratondo
    PROCEEDINGS OF THE 2014 INTERNATIONAL CONFERENCE ON PHYSICS AND ITS APPLICATIONS, 2015, : 123 - 126
  • [10] Characterization of gelatin-immobilized pigeonpea urease and preparation of a new urea biosensor
    Srivastava, PK
    Kayastha, AM
    Srinivasan
    BIOTECHNOLOGY AND APPLIED BIOCHEMISTRY, 2001, 34 (01) : 55 - 62