Brachytherapy dosimetry parameters calculated for a new 103Pd source

被引:18
|
作者
Rivard, MJ
Melhus, CS
Kirk, BL
机构
[1] Tufts Univ, Sch Med, Tufts New England Med Ctr, Dept Radiat Oncol, Boston, MA 02111 USA
[2] Oak Ridge Natl Lab, Nucl Computat Informat Grp, Nucl Sci & Technol Div, Oak Ridge, TN 37831 USA
关键词
Monte Carlo calculations; brachytherapy dosimetry; TG-43; MCNP; Pd-103;
D O I
10.1118/1.1781552
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new brachytherapy source having Pd-103 adsorbed onto silver beads has been designed. The dose distributions of this source have been characterized using version 5 of the MCNP Monte Carlo radiation transport code available from Oak Ridge National Laboratory. These results are presented in terms of the updated AAPM Task Group No. 43 (TG-43U1) formalism, dosimetry parameters, and recommended calculation methodology. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:2466 / 2470
页数:5
相关论文
共 50 条
  • [1] Monte Carlo calculations and experimental measurements of dosimetry parameters of a new 103Pd source
    Li, ZF
    Palta, JR
    Fan, JJ
    [J]. MEDICAL PHYSICS, 2000, 27 (05) : 1108 - 1112
  • [2] Dosimetric parameters of the new design 103Pd brachytherapy source based on Monte Carlo study
    Saidi, Pooneh
    Sadeghi, Mahdi
    Shirazi, Alireza
    Tenreiro, Claudio
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (01): : 13 - 18
  • [3] First Report of a New 103PD Line Source for Prostate Brachytherapy
    Beyer, D. C.
    Shishkov, I.
    Ge, W.
    Kaminetsky, J.
    Stock, R. G.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E203 - E204
  • [4] Radiological characterization of a 103Pd ocular brachytherapy source
    Hearn, R
    Soares, CG
    Bergman, J
    Millage, K
    Napolatino, M
    Rodgers, J
    [J]. MEDICAL PHYSICS, 2004, 31 (06) : 1914 - 1914
  • [5] Relative dosimetry analysis of a high dose-rate 103Pd ocular brachytherapy source
    Millage, K
    Napolitano, M
    Hearn, R
    Rodgers, J
    [J]. MEDICAL PHYSICS, 2004, 31 (06) : 1914 - 1914
  • [6] ROPES eye plaque brachytherapy dosimetry for two models of 103Pd seeds
    Pooneh Saidi
    Mahdi Sadeghi
    Alireza Shirazi
    Claudio Tenreiro
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2011, 34 : 223 - 231
  • [7] ROPES eye plaque brachytherapy dosimetry for two models of 103Pd seeds
    Saidi, Pooneh
    Sadeghi, Mahdi
    Shirazi, Alireza
    Tenreiro, Claudio
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2011, 34 (02) : 223 - 231
  • [8] 103Pd strings: Monte Carlo assessment of a new approach to brachytherapy source design
    Rivard, Mark J.
    Reed, Joshua L.
    DeWerd, Larry A.
    [J]. MEDICAL PHYSICS, 2014, 41 (01)
  • [9] 103Pd brachytherapy dose prescription
    Wallace, RE
    [J]. MEDICAL PHYSICS, 1999, 26 (09) : 2034 - 2035
  • [10] Dosimetry Parameters and TPS Commissioning for the CivaSheet Directional Pd-103 Brachytherapy Source
    Rivard, M. J.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3472 - 3472