Elecsys® and Kryptor immunoassays for the measurement of sFlt-1 and PlGF to aid preeclampsia diagnosis: are they comparable?

被引:24
|
作者
Stepan, Holger [1 ]
Hund, Martin [2 ]
Dilba, Peter [3 ]
Sillman, Johanna [2 ]
Schlembach, Dietmar [4 ]
机构
[1] Univ Leipzig, Dept Obstet, Liebigstr 20a, D-04103 Leipzig, Germany
[2] Roche Diagnost Int Ltd, Rotkreuz, Switzerland
[3] Roche Diagnost GmbH, Penzberg, Germany
[4] Clinicum Neukoelln, Clin Obstet, Vivantes Network Hlth, Berlin, Germany
关键词
immunoassay; preeclampsia; sFlt-1/PlGF; PLACENTAL GROWTH-FACTOR; TYROSINE KINASE 1; FACTOR RATIO; PREDICTION;
D O I
10.1515/cclm-2018-1228
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: For pregnant women with suspected preeclampsia, the soluble fms-like tyrosine-kinase 1 (sFlt-1)/placental growth factor (PlGF) ratio is a biomarker to aid diagnosis. We performed method comparisons between Elecsys (R) and Kryptor sFlt-1 and PlGF immunoassays and assessed the diagnostic performance for preeclampsia. Methods: Serum samples from a case-control study involving 113 pregnant women with preeclampsia/elevated liver enzymes and low platelet count (HELLP) and 270 controls were analyzed. sFlt-1 and PlGF were measured using Roche Elecsys (R) and BRAHMS Kryptor sFlt-1/PlGF immunoassays. The sFlt-1/PlGF ratios were calculated, and Passing-Bablok regression/Bland-Altman plots were performed. Gestation-specific cut-offs, <= 33 and >= 85/>= 110, were assessed. Results: Mean (+/- 2 standard deviation [SD]) differences between the Elecsys (R) and Kryptor values were: sFlt-1, 173.13 pg/mL (6237.66, -5891.40); PlGF, -102.71 pg/mL (186.06, -391.48); and sFlt-1/PlGF, 151.74 (1085.11, -781.63). The Elecsys (R) and Kryptor immunoassays showed high correlation: Pearson's correlation coefficients were 0.913 (sFlt-1) and 0.945 (PlGF). Slopes were 1.06 (sFlt-1) and 0.79 (PlGF), resulting in similar to 20% lower values for Kryptor PlGF. Sensitivities and specificities using the sFlt-1/PlGF >= 85 cut-off for early-onset preeclampsia (20 + 0 to 33 + 6 weeks) were 88.1%/100.0% (Elecsys (R)) and 90.5%/96.2% (Kryptor), respectively, and using the >= 110 cut-off for late-onset preeclampsia (>= 34 + 0 weeks) were 51.3%/96.5% (Elecsys (R)) and 78.9%/90.1% (Kryptor), respectively. Using Elecsys (R) and Kryptor sFlt-1/PlGF, 0% and 3.8% of women, respectively, were falsely ruled-in for early-onset, and 3.5% and 9.9%, respectively, for late-onset preeclampsia. Conclusions: Despite high correlation between the Elecsys (R) and Kryptor immunoassays, we observed significant differences between sFlt-1/PlGF and PlGF results. Therefore, sFlt-1/PlGF cut-offs validated for Elecsys (R) immunoassays are not transferable to Kryptor immunoassays.
引用
收藏
页码:1339 / 1348
页数:10
相关论文
共 50 条
  • [1] Correlation of Kryptor and Elecsys® immunoassay sFlt-1/PlGF ratio on early diagnosis of preeclampsia and fetal growth restriction: A case-control study
    Simon, Elisa
    Herraiz, Ignacio
    Villalain, Cecilia
    Isabel Gomez-Arriaga, Paula
    Soledad Quezada, Maria
    Ana Lopez-Jimenez, Elena
    Galindo, Alberto
    PREGNANCY HYPERTENSION-AN INTERNATIONAL JOURNAL OF WOMENS CARDIOVASCULAR HEALTH, 2020, 20 : 44 - 49
  • [2] Efficiency of sFlt-1/PlGF Ratio in Preeclampsia Diagnosis
    Tataru-Copos, Anca-Florina
    Popescu, Mircea Ioachim
    Murvai, Romina
    El Kharoubi, Amin
    MEDICINA-LITHUANIA, 2022, 58 (09):
  • [3] Update on the Diagnosis and Prognosis of Preeclampsia with the Aid of the sFlt-1/PlGF Ratio in Singleton Pregnancies
    Herraiz, Ignacio
    Llurba, Elisa
    Verlohren, Stefan
    Galindo, Alberto
    FETAL DIAGNOSIS AND THERAPY, 2018, 43 (02) : 81 - 89
  • [4] sFlt-1, PlGF, sFlt-1/PlGF ratio and uterine artery Doppler for preeclampsia diagnostics
    Taraseviciene, Viktorija
    Grybauskiene, Regina
    Maciuleviciene, Regina
    MEDICINA-LITHUANIA, 2016, 52 (06): : 349 - 353
  • [5] Clinical implementation of sFlt-1 and PLGF assays for preeclampsia diagnosis
    Yip, P. M.
    Miller, J.
    Higgins, V.
    Fu, L.
    CLINICA CHIMICA ACTA, 2024, 558
  • [6] Biochemical Markers, sFLT-1/PlGF, Aid in the Diagnostic Odyssey of Preeclampsia
    Strickland, Sydney Webb
    Zhu, Yusheng
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2023, 8 (03): : 440 - 442
  • [7] Evaluation of the angiogenic factors sFlt-1, PlGF, and the sFlt-1/PlGF ratio in preeclampsia and associated features
    Gannoun, Marwa Ben Ali
    Mehdi, Meriem
    Zitouni, Hedia
    Boussabah, Manel
    Zouari, Ines
    Jlali, Ali
    Almawi, Wassim Y.
    AMERICAN JOURNAL OF REPRODUCTIVE IMMUNOLOGY, 2023, 90 (01)
  • [8] Longitudinal assessment of PlGF, sFlt-1 and sFlt-1/PlGF in pregnant women with risk factors for preeclampsia
    Mendes, F. S.
    Lucena, F. C.
    Lage, E. M.
    Teixeira, P. G.
    Diniz, R.
    Alpoim, P. N.
    Dusse, L. M. S.
    THROMBOSIS RESEARCH, 2019, 175 : S12 - S13
  • [9] Angiogenic balance (sFlt-1/PlGF) and preeclampsia
    Lecarpentier, Edouard
    Tsatsaris, Vassilis
    ANNALES D ENDOCRINOLOGIE, 2016, 77 (02) : 97 - 100
  • [10] ON THE ORIGIN AND DISAPPEARANCE OF SFLT-1 AND PLGF IN PREECLAMPSIA
    Saleh, L.
    Van den Meiracker, A. H.
    Geensen, R.
    Van Lennep, J. E. Roeters
    Duvekot, J. J.
    Verdonk, K.
    Steegers, E. A. P.
    Russcher, H.
    Danser, A. H. J.
    Visser, W.
    JOURNAL OF HYPERTENSION, 2017, 35 : E176 - E176