The closing behavior of mechanical aortic heart valve prostheses

被引:22
|
作者
Lu, PC [1 ]
Liu, JS
Huang, RH
Lo, CW
Lai, HC
Hwang, NHC
机构
[1] Tamkang Univ, Dept Water Resources & Environm Engn, Taipei 251, Taiwan
[2] Tamkang Univ, Dept Mech & Electromech Engn, Taipei 251, Taiwan
关键词
D O I
10.1097/01.MAT.0000130678.59655.c0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Mechanical artificial heart valves rely on reverse flow to close their leaflets. This mechanism creates regurgitation and water hammer effects that may form cavitations, damage blood cells, and cause thromboembolism. This study analyzes closing mechanisms of monoleaflet (Medtronic Hall 27), bileaflet (Carbo-Medics 27; St. Jude Medical 27; Duromedics 29), and trileaflet valves in a circulatory mock loop, including an aortic root with three sinuses. Downstream flow field velocity was measured via digital particle image velocimetry (DPIV). A high speed camera (PIVCAM 10-30 CCD video camera) tracked leaflet movement at 1000 frames/s. All valves open in 40-50 msec, but monoleaflet and bileaflet valves close in much less time (< 35 msec) than the trileaflet valve (> 75 msec). During acceleration phase of systole, the monoleaflet forms a major and minor flow, the bileaflet has three jet flows, and the trileaflet produces a single central flow like physiologic valves. In deceleration phase, the aortic sinus vortices hinder monoleaflet and bileaflet valve closure until reverse flows and high negative transvalvular pressure push the leaflets rapidly for a hard closure. Conversely, the vortices help close the trileaflet valve more softly, probably causing less damage, lessening back flow, and providing a washing effect that may prevent thrombosis formation.
引用
收藏
页码:294 / 300
页数:7
相关论文
共 50 条
  • [1] APPLICATION OF THE NATURAL AORTIC-VALVE CLOSING MECHANISM TO THE DESIGN OF HEART-VALVE PROSTHESES
    VANSTEENHOVEN, AA
    LIU, SH
    VEENSTRA, PC
    ARTIFICIAL ORGANS, 1979, 3 (03) : 315 - 315
  • [2] Closing behavior of the mechanical heart valve in a total artificial heart
    H. S. Lee
    T. Tsukiya
    A. Homma
    Y. Taenaka
    E. Tatsumi
    H. Takano
    Journal of Artificial Organs, 2003, 6 (1) : 0037 - 0041
  • [3] INFLUENCE OF THE AORTIC ROOT COMPLIANCE ON HEMOLYSIS AND CLOSING DYNAMICS OF HEART VALVE PROSTHESES: AN IN VITRO STUDY
    Linde, T.
    Mager, I.
    Candan, S. I.
    Schmitz, C.
    Utzenrath, M.
    Schmitz-Rode, T.
    Steinseifer, U.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2010, 33 (07): : 430 - 430
  • [4] On the closing sounds of a mechanical heart valve
    Wu, CF
    Herman, BA
    Retta, SM
    Grossman, LW
    Liu, JS
    Hwang, NHC
    ANNALS OF BIOMEDICAL ENGINEERING, 2005, 33 (06) : 743 - 750
  • [5] On the Closing Sounds of a Mechanical Heart Valve
    Changfu Wu
    Bruce A. Herman
    Stephen M. Retta
    Laurence W. Grossman
    Jia-Shing Liu
    Ned H. C. Hwang
    Annals of Biomedical Engineering, 2005, 33 : 743 - 750
  • [6] INVITRO COMPARISON OF AORTIC HEART-VALVE PROSTHESES .1. MECHANICAL VALVES
    KNOTT, E
    REUL, H
    KNOCH, M
    STEINSEIFER, U
    RAU, G
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1988, 96 (06): : 952 - 961
  • [7] CLOSING BEHAVIOR OF NATURAL AORTIC-VALVE
    VANSTEENHOVEN, AA
    VANDERSCHAAR, PJ
    RENEMAN, RS
    FEDERATION PROCEEDINGS, 1978, 37 (03) : 217 - 217
  • [8] INVITRO PULSATILE FLOW HEMODYNAMICS OF 5 MECHANICAL AORTIC HEART-VALVE PROSTHESES
    WALKER, PG
    YOGANATHAN, AP
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 : S113 - S123
  • [9] Mechanical heart valve prostheses - Identification and evaluation
    Butany, J
    Ahluwalia, MS
    Munroe, C
    Fayet, C
    Ahn, C
    Blit, P
    Kepron, C
    Cusimano, RJ
    Leask, RL
    CARDIOVASCULAR PATHOLOGY, 2003, 12 (01) : 1 - 22
  • [10] Numerical comparison of the closing dynamics of a new trileaflet and a bileaflet mechanical aortic heart valve
    Chi-Pei Li
    Po-Chien Lu
    Journal of Artificial Organs, 2012, 15 : 364 - 374