A new Low Power Instrument for Impedance Measurements in Biomedicine based on FFT. Application to Interleukin-10 protein detection

被引:0
|
作者
Palacio, F. [1 ]
Prades, J. D. [1 ]
Gomez, J. M. [1 ]
Martinez, M. [2 ]
Errachid, A. [3 ]
Errachid, A. [3 ]
Lopez, M. [1 ]
机构
[1] Univ Barcelona, Dept Elect, Barcelona, Spain
[2] Univ Barcelona, Dept C Mat & Met, Barcelona, Spain
[3] Univ Claude Bernard Lyon, Lab Sci Analyt, Villeurbanne, France
关键词
Electrochemical sensors; impedance measurements; portable electronic system;
D O I
10.1016/j.proeng.2014.11.670
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
We describe a low power, easy to use and portable device for measuring the electrochemical impedance of different types of biosensors. The measurement technique is based on the direct synthesis of sinusoidal signals followed by a discrete time Fourier transform analysis, carried out by an integrated electronic stage that excites and collects the response from the sample. This kind of technique provides very good signal to noise ratio, high sensibility to small variations in the sensor impedance and high accuracy and reliability. The device was tested studying the sensor response at different frequencies of a gold microelectrode, fabricated through a combination of soft lithography tools, capable of monitoring the antibody-antigen concentration for the Interleukin 10 (IL-10). Our custom design makes possible extending the current range down to the pA, with accuracy better than 1%, and optimizing the power consumption to reach figures under 150 mu W per measurement. (c) 2014 The Authors. Published by Elsevier Ltd.
引用
收藏
页码:312 / 315
页数:4
相关论文
共 4 条
  • [1] Novel capacitance biosensor based on hafnium oxide for interleukin-10 protein detection
    Lee, Michael
    Zine, Nadia
    Baraket, Abdellatif
    Zabala, Miguel
    Campabadal, Francesca
    Trivella, Maria Giovanna
    Jaffrezic-Renault, Nicole
    Errachid, Abdelhamid
    EUROSENSORS XXV, 2011, 25
  • [2] A novel biosensor based on hafnium oxide: Application for early stage detection of human interleukin-10
    Lee, Michael
    Zine, Nadia
    Baraket, Abdellatif
    Zabala, Miguel
    Campabadal, Francesca
    Caruso, Raffaele
    Trivella, Maria Giovanna
    Jaffrezic-Renault, Nicole
    Errachid, Abdelhamid
    SENSORS AND ACTUATORS B-CHEMICAL, 2012, 175 : 201 - 207
  • [3] Low Power Ice Detection With Capacitive and Impedance Spectroscopy-based Measurements
    Felsberger, Richard
    Schweighofer, Bernhard
    Flatscher, Matthias
    Rath, Matthias
    Grubmuller, Michael
    Wegleiter, Hannes
    2018 IEEE 27TH INTERNATIONAL SYMPOSIUM ON INDUSTRIAL ELECTRONICS (ISIE), 2018, : 809 - 813
  • [4] Constitutive, but Not Challenge-Induced, Interleukin-10 Production Is Robust in Acute Pre-Pubescent Protein and Energy Deficits: New Support for the Tolerance Hypothesis of Malnutrition-Associated Immune Depression Based on Cytokine Production in vivo
    Monk, Jennifer M.
    Steevels, Tessa A. M.
    Hillyer, Lyn M.
    Woodward, Bill
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2011, 8 (01): : 117 - 135