Performance evaluation of a chemiluminescence microparticle immunoassay for CK-MB

被引:5
|
作者
Lin, Zhi-yuan [1 ,2 ]
Fang, Yi-zhen [1 ]
Jin, Hong-wei [1 ]
Lin, Hua-yue [1 ]
Dai, Zhang [1 ]
Luo, Qing [1 ]
Li, Hong-wei [1 ]
Lin, Yan-ling [1 ]
Huang, Shui-zhen [1 ]
Gao, Lei [3 ]
Xu, Fei-hai [3 ]
Zhang, Zhong-ying [1 ]
机构
[1] Xiamen Univ, Affiliated Zhongshan Hosp, Ctr Clin Lab, Xiamen, Peoples R China
[2] Fujian Med Univ, Union Hosp, Fuzhou, Fujian, Peoples R China
[3] Xiamen Innodx Biotech Co Ltd, Xiamen, Peoples R China
关键词
chemiluminescence microparticle immunoassay; CK-MB; evaluation; kit; performance;
D O I
10.1002/jcla.22426
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BackgroundTo verify and evaluate the performance characteristics of a creatine kinase phosphokinase isoenzymes MB (CK-MB) assay kit, which produced by Xiamen Innodx Biotech Co. Ltd. MethodsEvaluation was carried out according to Guidelines for principle of analysis performance evaluation of in vitro diagnostic reagent. The performance parameters included detection limit, linearity range, reportable range, recovery test, precision verification, interference test, cross-reactivity, matrix effect, and method comparison. ResultsThe detection limit was 0.1ng/mL. The assay had clinical linearity over range of 0.1ng/mL-500ng/mL. Reportable range was from 0.1ng/mL to 1000ng/mL. The average percent of recovery was 99.66%. The coefficient of variation (CV) for within-run and between-run of low CK-MB sample was 5.55% and 6.16%, respectively. As for high-level sample, it was 7.88% and 7.80%. In medical decision level, the relative deviation (Bias) of all interference tests was lower than 15%. When the sample had mild-hemolysis; hemoglobin 15g/L; triglyceride 17mmol/L; bilirubin 427.5mol/L; rheumatoid factor 206U/mL, there was no significant interference to be found. Moreover, assay kit had no cross-reaction with CK-MM and CK-BB. At last, total diagnostic accuracy of kit was 93.24%, when compared with refer kit. ConclusionOverall the results of the verification study indicated the performance of kit is met the requirements of the clinical test.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] EVALUATION OF ABBOTT IMX CK-MB IMMUNOASSAY
    SASSE, EA
    MADIEDO, G
    KOPINSKI, W
    [J]. CLINICAL CHEMISTRY, 1990, 36 (10) : 1858 - 1859
  • [2] EVALUATION OF BECTON-DICKINSON (BD) CK-MB IMMUNOASSAY - COMPARISON WITH MAGIC LITE CK-MB AND ELECTROPHORESIS
    PREESE, L
    BENNETT, R
    FREDRICKSON, A
    APPLE, F
    [J]. CLINICAL CHEMISTRY, 1988, 34 (06) : 1283 - 1284
  • [3] CLINICAL-EVALUATION OF AN ENZYME-IMMUNOASSAY FOR CK-MB
    FENTON, JJ
    GORDON, WC
    BELL, ML
    RIPPE, DF
    [J]. CLINICAL CHEMISTRY, 1983, 29 (06) : 1218 - 1218
  • [4] EVALUATION OF ABBOTT IMX CK-MB IMMUNOASSAY - COMPARISON WITH BAXTER STRATUS CK-MB, HYBRITECH TANDEM E-II-CK-MB, AND ELECTROPHORESIS
    PREESE, L
    RENNEKE, J
    APPLE, FS
    [J]. CLINICAL CHEMISTRY, 1990, 36 (06) : 1130 - 1130
  • [5] EVALUATION OF THE ABBOTT IMX CK-MB IMMUNOASSAY - COMPARISON WITH THE ROCHE ISOMUNE MB ASSAY
    GIACHERIO, D
    PYZIKSHULER, R
    [J]. CLINICAL CHEMISTRY, 1990, 36 (06) : 1132 - 1132
  • [6] CK-MB ASSAY FOR THE AURAFLEX(R) IMMUNOASSAY SYSTEM
    GILLIGAN, E
    CARDINAELS, C
    [J]. CLINICAL CHEMISTRY, 1995, 41 (06) : S62 - S62
  • [7] Electrochemical Quantification of CK-MB in Serum Based on Immunoassay
    Li, Yongxing
    Guo, Hua
    Hu, Yamin
    Qi, Xiangqian
    [J]. INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2018, 13 (01): : 287 - 295
  • [8] ENHANCED LUMINESCENCE IMMUNOASSAY FOR CREATINE-KINASE MB (CK-MB)
    BILLING, J
    HINCHY, V
    MONTAGUE, D
    [J]. CLINICAL CHEMISTRY, 1990, 36 (06) : 1130 - 1130
  • [9] PRELIMINARY ASSESSMENT OF THE MAGIC LITE CK-MB IMMUNOASSAY
    PETERSON, R
    [J]. CLINICAL CHEMISTRY, 1988, 34 (10) : 2164 - 2164
  • [10] EVALUATION OF AN IMMUNOCHEMICAL PROCEDURE FOR CK-MB
    PESCE, MA
    BODOURIAN, SH
    [J]. CLINICAL CHEMISTRY, 1982, 28 (07) : 1617 - 1617