Long-term changes of hydroxyapatite-coated dental implants

被引:0
|
作者
Baltag, L
Watanabe, K
Kusakari, H
Taguchi, N
Miyakawa, O
Kobayashi, M
Ito, N
机构
[1] Niigata Univ, Fac Dent, Dept Fixed Prosthodont, Niigata 9518514, Japan
[2] Niigata Univ, Fac Dent, Dept Dent Mat & Technol, Niigata, Japan
[3] Niigata Univ, EMX Lab, Niigata, Japan
来源
关键词
hydroxyapatite; retrieved blade implant; X-ray diffraction; FTIR; electron probe microanalysis;
D O I
10.1002/(SICI)1097-4636(2000)53:1<76::AID-JBM11>3.0.CO;2-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There are many controversies about the long-term prognosis of hydroxyapatite (HA)-coated implants. Failure mag be related to compositional and structural changes of the coating occurring during implantation. Two retrieved and two unused HA-coated blade-type implants were examined by stereomicroscopy, secondary electron microscopy, Fourier transform infrared spectroscopy, X-ray diffraction, and electron probe microanalysis. The objective was to investigate the HA morphology, composition, and structure, and to characterize the changes that occurred in the retrieved implant coatings. Retrieved implants presented partial Loss of the coating, especially at the apical and mesiodistal edges. Remaining HA was thick and Battened in the cervical and central areas and gradually thinner and rougher towards the apical and mesiodistal edges. Increase of CI and Mg, decrease of OH, and X-ray diffraction peak broadening mere found in the retrieved implant coatings, in comparison with the unused implants. Morphological changes of the retrieved implants seem to depend on stress values in the surrounding bone and on implant mobility, Compositional changes and increased amount of lattice imperfections appeared in the retrieved implant coatings, as a result of ion substitutions in the apatite lattice. However, the present study could not confirm the influence of these changes on implant failure. (C) 2000 John Wiley & Sons, Inc.
引用
收藏
页码:76 / 85
页数:10
相关论文
共 50 条
  • [1] Long-term follow-up of hydroxyapatite-coated dental implants - A clinical trial
    Binahmed, Abdulaziz
    Stoykewych, Andrew
    Hussain, Ali
    Love, Brock
    Pruthi, Vijay
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2007, 22 (06) : 963 - 968
  • [2] Variability of hydroxyapatite-coated dental implants
    Gross, KA
    Berndt, CC
    Iacono, VJ
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 1998, 13 (05) : 601 - 610
  • [3] The survival of hydroxyapatite-coated implants is questioned
    Nico Creugers
    Evidence-Based Dentistry, 2002, 3 (3) : 77 - 78
  • [4] Osteoinductive hydroxyapatite-coated titanium implants
    Ripamonti, Ugo
    Roden, Laura C.
    Renton, Louise F.
    BIOMATERIALS, 2012, 33 (15) : 3813 - 3823
  • [5] Hydroxyapatite-coated implants: A case for their use
    Morris, HF
    Ochi, S
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1998, 56 (11) : 1303 - 1311
  • [6] Comparative evaluation of dense hydroxyapatite implants and hydroxyapatite-coated implants.
    Ogiso, M
    Yamamura, M
    Kuo, PT
    Lee, RR
    Borgese, D
    Matsumoto, T
    JOURNAL OF DENTAL RESEARCH, 1996, 75 : 2808 - 2808
  • [7] Long-Term Outcomes of a New Model of Anatomical Hydroxyapatite-Coated Hip Prosthesis
    Herrera, Antonio
    Mateo, Jesus
    Lobo-Escolar, Antonio
    Jose Panisello, Juan
    Ibarz, Elena
    Gracia, Luis
    JOURNAL OF ARTHROPLASTY, 2013, 28 (07): : 1160 - 1166
  • [8] HYDROXYAPATITE-COATED SUBPERIOSTEAL DENTAL IMPLANTS - DESIGN RATIONALE AND CLINICAL-EXPERIENCE
    KAY, JF
    GOLEC, TS
    RILEY, RL
    JOURNAL OF PROSTHETIC DENTISTRY, 1987, 58 (03): : 339 - 343
  • [9] Fatigue properties of hydroxyapatite-coated dental implants after exposure to a periodontal pathogen
    Mukherjee, DP
    Dorairaj, NR
    Mills, DK
    Graham, D
    Krauser, JT
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2000, 53 (05): : 467 - 474
  • [10] Long-term results of a hydroxyapatite-coated femoral component in total hip replacement
    Rajaratnam, S. S.
    Jack, C.
    Tavakkolizadeh, A.
    George, M. D.
    Fletcher, R. J.
    Hankins, M.
    Shepperd, J. A. N.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (01): : 27 - 30