Evaluating image denoising methods in myocardial perfusion single photon emission computed tomography (SPECT) imaging

被引:9
|
作者
Skiadopoulos, S. [1 ]
Karatrantou, A. [1 ]
Korfiatis, P. [1 ]
Costaridou, L. [1 ]
Vassilakos, P. [2 ]
Apostolopoulos, D. [2 ]
Panayiotakis, G. [1 ]
机构
[1] Univ Patras, Sch Med, Dept Med Phys, Patras 26500, Greece
[2] Univ Patras, Dept Nucl Med, Sch Med, Patras 26500, Greece
关键词
SPECT imaging; myocardial perfusion; image denoising; multi-scale platelet analysis; Butterworth filter; attenuation correction; WIENER FILTER; NUCLEAR; ATTENUATION; RECONSTRUCTION; STRATEGIES; SCATTER; NOISE; METZ;
D O I
10.1088/0957-0233/20/10/104023
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The statistical nature of single photon emission computed tomography (SPECT) imaging, due to the Poisson noise effect, results in the degradation of image quality, especially in the case of lesions of low signal-to-noise ratio (SNR). A variety of well-established single-scale denoising methods applied on projection raw images have been incorporated in SPECT imaging applications, while multi-scale denoising methods with promising performance have been proposed. In this paper, a comparative evaluation study is performed between a multi-scale platelet denoising method and the well-established Butterworth filter applied as a pre- and post-processing step on images reconstructed without and/or with attenuation correction. Quantitative evaluation was carried out employing (i) a cardiac phantom containing two different size cold defects, utilized in two experiments conducted to simulate conditions without and with photon attenuation from myocardial surrounding tissue and (ii) a pilot-verified clinical dataset of 15 patients with ischemic defects. Image noise, defect contrast, SNR and defect contrast-to-noise ratio (CNR) metrics were computed for both phantom and patient defects. In addition, an observer preference study was carried out for the clinical dataset, based on rankings from two nuclear medicine clinicians. Without photon attenuation conditions, denoising by platelet and Butterworth post-processing methods outperformed Butterworth pre-processing for large size defects, while for small size defects, as well as with photon attenuation conditions, all methods have demonstrated similar denoising performance. Under both attenuation conditions, the platelet method showed improved performance with respect to defect contrast, SNR and defect CNR in the case of images reconstructed without attenuation correction, however not statistically significant (p > 0.05). Quantitative as well as preference results obtained from clinical data showed similar performance of the denoising methods studied. In conclusion, the multi-scale platelet denoising method applied on raw projection images provides more efficient noise reduction while preserving image quality in a myocardial phantom SPECT imaging as compared to the Butterworth filter applied either on projection or reconstructed images. However, this trend in favour of the platelet denoising method was not observed on clinical data reconstructed either without or with attenuation correction.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Myocardial Perfusion Single-Photon Emission Computed Tomography (SPECT) Image Denoising: A Comparative Study
    Rahimian, Abdurrahim
    Etehadtavakol, Mahanaz
    Moslehi, Masoud
    Ng, Eddie Y. K.
    [J]. DIAGNOSTICS, 2023, 13 (04)
  • [2] Single Photon Emission Computed Tomography (SPECT) Myocardial Perfusion Imaging Guidelines: Instrumentation, Acquisition, Processing, and Interpretation
    Sharmila Dorbala
    Karthik Ananthasubramaniam
    Ian S. Armstrong
    Panithaya Chareonthaitawee
    E. Gordon DePuey
    Andrew J. Einstein
    Robert J. Gropler
    Thomas A. Holly
    John J. Mahmarian
    Mi-Ae Park
    Donna M. Polk
    Raymond Russell
    Piotr J. Slomka
    Randall C. Thompson
    R. Glenn Wells
    [J]. Journal of Nuclear Cardiology, 2018, 25 : 1784 - 1846
  • [3] Single Photon Emission Computed Tomography (SPECT) Myocardial Perfusion Imaging Guidelines: Instrumentation, Acquisition, Processing, and Interpretation
    Dorbala, Sharmila
    Ananthasubramaniam, Karthik
    Armstrong, Ian S.
    Chareonthaitawee, Panithaya
    DePuey, E. Gordon
    Einstein, Andrew J.
    Gropler, Robert J.
    Holly, Thomas A.
    Mahmarian, John J.
    Park, Mi-Ae
    Polk, Donna M.
    Russell, Raymond, III
    Slomka, Piotr J.
    Thompson, Randall C.
    Wells, R. Glenn
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2018, 25 (05) : 1784 - 1846
  • [4] Attenuation correction for single photon emission computed tomography myocardial perfusion imaging
    Corbett J.R.
    Kritzman J.N.
    Ficaro E.P.
    [J]. Current Cardiology Reports, 2004, 6 (1) : 32 - 40
  • [5] PREDICTORS OF NORMAL SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY MYOCARDIAL PERFUSION IMAGING
    Koh, Angela
    Flores, Jennifer
    Keng, Felix
    Chua, Terrance S. J.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (10)
  • [6] Sequential single-photon emission computed tomography myocardial perfusion imaging
    Shaw, LJ
    Berman, DS
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2005, 96 (8A): : 28J - 39J
  • [7] Effect of correction methods on image resolution of myocardial perfusion imaging using single photon emission computed tomography combined with computed tomography hybrid systems
    Tantawy, Hazem M.
    Abdelhafez, Yasser G.
    Helal, Nadia L.
    Kany, Amr, I
    Saad, Ibrahim E.
    [J]. JOURNAL OF PHYSICS COMMUNICATIONS, 2020, 4 (01):
  • [8] Fusion imaging between myocardial perfusion single photon emission computed tomography and cardiac computed tomography
    Nakaura, T
    Utsunomiya, D
    Shiraishi, S
    Tomiguchi, S
    Kawanaka, K
    Honda, T
    Awai, K
    Yamashita, Y
    [J]. CIRCULATION, 2005, 112 (03) : E47 - E48
  • [9] Cost-Effectiveness of stress-only myocardial perfusion single photon emission computed tomography (SPECT) imaging
    Vallejo, Enrique
    Acevedo, Cesar
    Varela, Samuel
    Carlos Alburez, Jose
    Bialostozky, David
    [J]. GACETA MEDICA DE MEXICO, 2012, 148 (01): : 6 - 13
  • [10] Appropriateness of referrals for single-photon emission computed tomography myocardial perfusion imaging (SPECT-MPI) in Iran
    Gholamrezanezhad, Ali
    Shirafkan, Ahmadali
    Mirpour, Sahar
    [J]. IRANIAN JOURNAL OF NUCLEAR MEDICINE, 2011, 19 (01): : 1 - 5