Determination of tectoridin in rat plasma by high-performance liquid chromatography and its application to pharmacokinetic studies

被引:5
|
作者
Liu, Taiming [1 ]
Chen, Zhuo [1 ]
Guo, Yuyi [1 ]
Jiang, Xuehua [1 ]
Ma, Guo [2 ]
机构
[1] Sichuan Univ, W China Sch Pharm, Key Lab Drug Targeting, Chengdu 610041, Peoples R China
[2] Fudan Univ, Sch Pharm, Shanghai 200032, Peoples R China
关键词
tectoridin; pharmacokinetics; high-performance liquid chromatography; plasma; BELAMCANDA-CHINENSIS; TECTORIGENIN; METABOLITES; RHIZOMES; INJURY;
D O I
10.1002/bmc.1180
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple, specific and sensitive HPLC method with UV detection was developed and validated for the determination of tectoridin in rat plasma for the first time. Chromatographic separation was performed on a Welchrom (TM) C-18 column(150 x 4.6 mm, i.d., 5 mu m) at a flow rate of 1.0 mL min(-1), using a mixture of methanol-2% HAc aqueous solution (31:69, v/v) as the mobile phase with UV detection at 266 nm. The calibration curves for tectoridin were linear over the concentration range of 1.10-274.40 mu g mL(-1) in rat plasma. The intra- and inter-day accuracies (RE) were within -3.23% and 4.11%. The intra- and inter-day precisions (RSD) were not more than 2.74 and 4.72%, respectively. The present method was successfully applied to the pharmacokinetic studies of tectoridin in rats after intravenous administration of three different doses. Copyright (C) 2009 John Wiley & Sons, Ltd.
引用
收藏
页码:754 / 758
页数:5
相关论文
共 50 条
  • [1] Determination of phillygenin in rat plasma by high-performance liquid chromatography and its application to pharmacokinetic studies
    Ye, Liang-hong
    Li, Yun-xia
    Peng, Cheng
    Gong, Xiao-hong
    Zheng, Xin-guang
    [J]. EUROPEAN JOURNAL OF DRUG METABOLISM AND PHARMACOKINETICS, 2013, 38 (03) : 201 - 207
  • [2] Determination of phillyrin in rat plasma by high-performance liquid chromatography and its application to pharmacokinetic studies
    Li, Yun-Xia
    Jiang, Xue-Hua
    Huang, Liang
    Lu, Jing
    Zhou, Jing
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2006, 41 (02) : 628 - 632
  • [3] Determination of phillygenin in rat plasma by high-performance liquid chromatography and its application to pharmacokinetic studies
    Liang-hong Ye
    Yun-xia Li
    Cheng Peng
    Xiao-hong Gong
    Xin-guang Zheng
    [J]. European Journal of Drug Metabolism and Pharmacokinetics, 2013, 38 : 201 - 207
  • [4] Determination of forsythiaside in rat plasma by high-performance liquid chromatography and its application to pharmacokinetic studies
    Li, Yun-Xia
    Jiang, Xue-Hua
    Liang, Hu-Yi
    Li, Xue
    [J]. BIOMEDICAL CHROMATOGRAPHY, 2008, 22 (04) : 361 - 366
  • [5] Determination of matrine in rat plasma by high-performance liquid chromatography and its application to pharmacokinetic studies
    Wu, XN
    Yamasita, F
    Hashida, M
    Chen, XG
    Hu, ZD
    [J]. TALANTA, 2003, 59 (05) : 965 - 971
  • [6] Determination of tanshinone I in rat plasma by high-performance liquid chromatography and its application to pharmacokinetic studies
    Du, Wei
    Wei, Yu-Hui
    Zhao, Gang
    Qin, Hong-Yan
    Wu, Xin-An
    [J]. JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2008, 46 (08) : 730 - 734
  • [7] High-Performance Liquid Chromatography Determination of Praziquantel in Rat Plasma; Application to Pharmacokinetic Studies
    Dey, S.
    Ghosh, M.
    Rangra, N. K.
    Kant, K.
    Shah, S. R.
    Pradhan, P. K.
    Singh, S.
    [J]. INDIAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2017, 79 (06) : 885 - 892
  • [8] Determination of protoapigenone in rat plasma by high-performance liquid chromatography with UV detection and its application in pharmacokinetic studies
    Wei, Anhua
    Zhou, Daonian
    Wu, Guanghua
    [J]. BIOMEDICAL CHROMATOGRAPHY, 2013, 27 (11) : 1452 - 1456
  • [9] High-Performance Liquid Chromatography Method For The Determination Of Xanthone In Rat & Its Application In Pharmacokinetic Studies
    Anuar, Nurfazreen
    Jaafar, Mohd Hafiz Mohd
    Hamid, Khuriah Abdul
    Effendi, Tommy Julianto Bustami
    [J]. IEEE SYMPOSIUM ON BUSINESS, ENGINEERING AND INDUSTRIAL APPLICATIONS (ISBEIA 2012), 2012, : 331 - 336
  • [10] Determination of schizandrin in rat plasma by high-performance liquid chromatography-mass spectrometry and its application in rat pharmacokinetic studies
    Xu, MJ
    Wang, GJ
    Xie, HT
    Wang, R
    Wang, W
    Li, XY
    Li, H
    Zhu, DN
    Yue, L
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 828 (1-2): : 55 - 61