Space-Charge Limited Transport in CdTe-based X- and γ-ray Detectors

被引:0
|
作者
Maslyanchuk, O. L. [1 ]
Solovan, M. M. [1 ]
Kulchynsky, V. V. [1 ]
Gnatyuk, V. A. [2 ]
Aoki, T. [3 ]
机构
[1] Yuriy Fedkovych Chernivtsi Natl Univ, Chernovtsy, Ukraine
[2] Natl Acad Sci Ukraine, VE Lashkaryov Inst Semicond Phys, Kiev, Ukraine
[3] Shizuoka Univ, Res Inst Elect, Hamamatsu, Shizuoka, Japan
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The principle feature limiting spectrometric performance of CdTe and CdMnTe-based X/gamma-ray detectors is the impossibility to increase the operating voltage to provide sufficient collection of charge carriers because of raising leakage current. We have carried out the detailed analysis of electrical characteristics of CdTe and CdMnTe-based detectors with two Ohmic contacts. It is shown that a rapid increase in leakage current at applying high bias voltage in the CdTe and CdMnTe crystals is caused by the current limited by space charge (SCLC). The same activation energy for the current of equilibrium holes and the current caused by nonequilibrium holes at high voltages confirms the fact that SCLC in the Ni/CdMnTe/Ni detector formed by injection of majority carriers (holes) due to lowering the barrier at an imperfect Ohmic contacts. This fact distinguishes Ni/CdMnTe/Ni detector from Pt/CdTe/Pt detector where SCLC formed by tunnel transitions of minority carriers from the Fermi level in the metal (or slightly below it) to the semiconductor through a thin insulating film between the crystal and metal contact. Comparison the measured and calculated current-voltage characteristics of CdMnTe crystal at different temperatures taking into account SCLC according to the Mott-Gurney theory allowed to determine the density of discrete trapping centers (1.4 x 10(13) cm(-3)) and energy of hole traps (0.39 eV) in the Ni/CdMnTe/Ni detector.
引用
收藏
页数:2
相关论文
共 50 条
  • [1] Charge transport features of CdTe-based X- and γ-ray detectors with Ti and TiOx Schottky contacts
    Maslyanchuk, Olena
    Solovan, Mykhailo
    Brus, Viktor
    Maryanchuk, Pavlo
    Maistruk, Eduard
    Fodchuk, Ihor
    Gnatyuk, Volodymyr
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 988
  • [2] Crystal defects and charge collection in CdTe-based X- and gamma-ray detectors
    Maslyanchuk, O.
    Fodchuk, I
    Solovan, M.
    Boledzyuk, I
    Kuzmin, A.
    Gnatyuk, V
    Aoki, T.
    FIFTEENTH INTERNATIONAL CONFERENCE ON CORRELATION OPTICS, 2021, 12126
  • [3] Effect of the concentration of uncompensated impurities on the properties of CdTe-based X- and γ-ray detectors
    L. A. Kosyachenko
    V. M. Sklyarchuk
    S. V. Melnychuk
    O. L. Maslyanchuk
    E. V. Grushko
    O. V. Sklyarchuk
    Semiconductors, 2012, 46 : 374 - 381
  • [4] Effect of the concentration of uncompensated impurities on the properties of CdTe-based X- and γ-ray detectors
    Kosyachenko, L. A.
    Sklyarchuk, V. M.
    Melnychuk, S. V.
    Maslyanchuk, O. L.
    Grushko, E. V.
    Sklyarchuk, O. V.
    SEMICONDUCTORS, 2012, 46 (03) : 374 - 381
  • [5] Capabilities of CdTe-Based Detectors With MoOx Contacts for Detection of X- and γ-Radiation
    Maslyanchuk, O. L.
    Solovan, M. M.
    Brus, V. V.
    Kulchynsky, V. V.
    Maryanchuk, P. D.
    Fodchuk, I. M.
    Gnatyuk, V. A.
    Aoki, T.
    Potiriadis, C.
    Kaissas, Y.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2017, 64 (05) : 1168 - 1172
  • [6] Possibilities of CdTe-based X/γ-ray Detectors with MoOx Contacts
    Maslyanchuk, O. L.
    Solovan, M. M.
    Kulchynsky, V. V.
    Brus, V. V.
    Maryanchuk, P. D.
    Fodchuk, I. M.
    Gnatyuk, V. A.
    Aoki, T.
    Potiriadis, C.
    Kaissas, Y.
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [7] Reasons of low charge collection efficiency in CdTe-based X/γ-ray detectors with ohmic contacts
    Aoki, Toru
    Maslyanchuk, O. L.
    Kosyachenko, L. A.
    Gnatyuk, V. A.
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS XV, 2013, 8852
  • [8] CdTe-based detectors for recording X-ray pulses with a subnanosecond resolution
    S. L. El’yash
    A. V. Rodigin
    T. V. Loiko
    A. I. Polyakov
    S. V. Kapitanov
    Instruments and Experimental Techniques, 2011, 54 : 555 - 557
  • [9] CdTe-based detectors for recording X-ray pulses with a subnanosecond resolution
    El'yash, S. L.
    Rodigin, A. V.
    Loiko, T. V.
    Polyakov, A. I.
    Kapitanov, S. V.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2011, 54 (04) : 555 - 557
  • [10] CdTe-based Schottky diode X-ray detectors for medical imaging
    Gnatyuk, V. A.
    Vlasenko, O. I.
    Kosyachenko, L. A.
    Maslyanchuk, O. L.
    Sklyarchuk, V. M.
    Grushko, E. V.
    Aoki, T.
    8TH INTERNATIONAL CONFERENCE ON CORRELATION OPTICS, 2008, 7008