Computer-assisted evoked electromyography with stimulating surgical instruments for recurrent/external laryngeal nerve identification and preservation in thyroid and parathyroid operation

被引:51
|
作者
Dackiw, APB
Rotstein, LE
Clark, OH
机构
[1] Univ Toronto, Toronto, ON, Canada
[2] Univ Calif San Francisco, San Francisco, CA 94143 USA
关键词
D O I
10.1067/msy.2002.128483
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background. The reported incidence of recurrent laryngeal nerve (RLN) palsy/paralysis in thyroid and parathyroid operation ranges from 2% to 13%. Injury to the external branch of the superior laryngeal nerve (EBSLN) is less clearly documented. We hypothesized that a novel evoked electromyography system using an audio warning alarm might be beneficial for detection and preservation of the RLN and EBSLN. Methods. A total of 117 thyroid/parathyroid operations were performed using a nerve locator/monitor (Neurovision SE, RLN Systems Inc, Jefferson City, Mo). Dissection was performed using a stimulating hemostat with conduction to an endotracheal surface electrode. Results. A total of 97 thyroidectomies (50 total, 4 7 lobectomies) and, 20 parathyroidectomies (16 directed, 4 bilateral) were performed representing 176 RLN and 152 EBSLN at risk. Of 176 RLN, 161 were correctly identified by the nerve stimulator alarm including 2 nonrecurrent nerves. The cricothyroid space and the superior pole vessels were scanned to identify the EBSLN by observing for cricothyroideus contraction or an alarm. Fourteen of 152 (8.9%) cases of type 2 anatomy were suggested where meticulous dissection of superior pole vessels prevented EBSLN injury. Conclusions. Computer-assisted evoked electromyography with stimulating surgical instruments is a useful surgical tool. This technology may be especially useful in reoperation in dense scar tissue and preserving the EBSLN in thyroid operation.
引用
收藏
页码:1100 / 1106
页数:7
相关论文
共 10 条