Volume guaranteed? Accuracy of a volume-targeted ventilation mode in infants

被引:12
|
作者
Farrell, Olivia [1 ,2 ]
Perkins, Elizabeth J. [1 ]
Black, Don [1 ]
Miedema, Martijn [1 ,3 ]
Paul, Joel Don [1 ,2 ]
Pereira-Fantini, Prue M. [1 ,2 ]
Tingay, David Gerald [1 ,2 ,4 ]
机构
[1] Murdoch Childrens Res Inst, Neonatal Res, Melbourne, Vic, Australia
[2] Univ Melbourne, Dept Paediat, Melbourne, Vic, Australia
[3] Acad Med Ctr, Neonatol, Amsterdam, Netherlands
[4] Royal Childrens Hosp, Neonatol, Melbourne, Vic, Australia
基金
英国医学研究理事会;
关键词
ARTERIAL CARBON-DIOXIDE; TIDAL VOLUME; NEWBORN-INFANTS; LUNG INJURY;
D O I
10.1136/archdischild-2017-312640
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objectives Volume-targeted ventilation (VTV) is widely used and may reduce lung injury, but this assumes the clinically set tidal volume (V-Tset) is accurately delivered. This prospective observational study aimed to determine the relationship between V-Tset, expiratory V-T (V-Te) and endotracheal tube leak in a modern neonatal volume-targeted ventilator (VTV) and the resultant partial arterial pressure of carbon dioxide (PaCO2) relationship with and without VTV. Design Continuous inflations were recorded for 24 hours in 100 infants, mean (SD) 34 (4) weeks gestation and 2483 (985) g birth weight, receiving synchronised mechanical ventilation (SLE5000, SLE, UK) with or without VTV and either the manufacturer's V4 (n=50) or newer V5 (n=50) VTV algorithm. The V-Tset, V-Te and leak were determined for each inflation (maximum 90 000/infant). If PaCO2 was sampled (maximum of 2 per infant), this was compared with the average V-Te data from the preceding 15 min. Results A total of 7 497 137 inflations were analysed. With VTV enabled (77 infants), the V-Tset V-Te bias (95% CI) was 0.03 (-0.12 to 0.19) mL/kg, with a median of 80% of V-Te being +/- 1.0 mL/kg of V-Tset. Endotracheal tube leak up to 30% influenced V-Tset-V-Te bias with the V4 (r(2)=-0.64, p<0.0001; linear regression) but not V5 algorithm (r(2)=0.04, p=0.21). There was an inverse linear relationship between V Te and PaCO2 without VTV (r(2)=0.26, p=0.004), but not with VTV (r(2)=0.04, p=0.10), and less PaCO2 within 40-60 mm Hg, 53% versus 72%, relative risk (95% CI) 1.7 (1.0 to 2.9). Conclusion VTV was accurate and reliable even with moderate leak and PaCO2 more stable. VTV algorithm differences may exist in other devices.
引用
收藏
页码:F120 / F125
页数:6
相关论文
共 50 条
  • [1] Volume Not Guaranteed: Closed Endotracheal Suction Compromises Ventilation in Volume-Targeted Mode
    Kiraly, Nicholas J.
    Tingay, David G.
    Mills, John F.
    Dargaville, Peter A.
    Copnell, Beverley
    NEONATOLOGY, 2011, 99 (01) : 78 - 82
  • [2] Volume-Targeted Ventilation
    Belteki, Gusztav
    Morley, Colin J.
    CLINICS IN PERINATOLOGY, 2021, 48 (04) : 825 - 841
  • [3] Volume-targeted ventilation
    Keszler M.
    Journal of Perinatology, 2005, 25 (Suppl 2) : S19 - S22
  • [4] Volume-targeted ventilation
    Keszler, Martin
    EARLY HUMAN DEVELOPMENT, 2006, 82 (12) : 811 - 818
  • [5] Volume-targeted ventilation
    Keszler, Martin
    Abubakar, M. Kabir
    SEMINARS IN PERINATOLOGY, 2024, 48 (02)
  • [6] Volume-targeted ventilation in infants born at or near term
    Chowdhury, Olie
    Rafferty, Gerrard F.
    Lee, Silke
    Hannam, Simon
    Milner, Anthony D.
    Greenough, Anne
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2012, 97 (04): : F264 - F266
  • [7] Methods and evidence on volume-targeted ventilation in preterm infants
    Claure, Nelson
    Bancalari, Eduardo
    CURRENT OPINION IN PEDIATRICS, 2008, 20 (02) : 125 - 131
  • [8] Volume-targeted ventilation of newborns
    Singh, Jaideep
    Sinha, Sunil K.
    Donn, Steven M.
    CLINICS IN PERINATOLOGY, 2007, 34 (01) : 93 - +
  • [9] Volume-Limited and Volume-Targeted Ventilation
    Morley, Colin J.
    CLINICS IN PERINATOLOGY, 2012, 39 (03) : 513 - +
  • [10] Volume-targeted versus pressure-limited ventilation for preterm infants
    Chen, Lih-Ju
    Chen, Jia-Yuh
    JOURNAL OF THE CHINESE MEDICAL ASSOCIATION, 2019, 82 (10) : 791 - 794