Preparation of a reference material for microplastics in water-evaluation of homogeneity

被引:39
|
作者
Seghers, John [1 ]
Stefaniak, Elzbieta A. [1 ]
La Spina, Rita [2 ]
Cella, Claudia [2 ]
Mehn, Dora [2 ]
Gilliland, Douglas [2 ]
Held, Andrea [1 ]
Jacobsson, Ulf [1 ]
Emteborg, Hakan [1 ]
机构
[1] European Commiss, Joint Res Ctr, B-2440 Geel, Belgium
[2] European Commiss, Joint Res Ctr, I-21027 Ispra, Italy
关键词
Reference material; Microplastics; Water; Homogeneity; PET; Harmonisation;
D O I
10.1007/s00216-021-03198-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Validation of analytical methods for measurements of microplastics (MP) is severely hampered because of a general lack of reference materials, RM. There is a great need to develop such reference materials. This study presents a concept of three-component kit with immobilised MP in solid NaCl, a surfactant and clean water that can be applied for the production of many types of MP RMs. As proof of concept, an RM for polyethylene terephthalate (PET) particles in water was prepared and evaluated for its homogeneity. The particles ranged from 30 mu m (Feret(min)) to about 200 mu m adapted by wet sieving. A specific number of PET particles were immobilized in about 0.29 g of solid NaCl by freeze-drying 1 mL of a NaCl suspension. By using manual and automated counting, twenty reconstituted 1-L water samples were evaluated for homogeneity with respect to number of PET particles from 30 mu m to > 200 mu m/L of water. The number of particles was 730 +/- 120 (mean +/- one standard deviation (SD); n = 10) and 865 +/- 155 particles (n = 10) obtained by optical microscopy in two independent laboratories. This corresponded to relative SDs of 16.4 and 17.9% and a mean of 797 +/- 151 particles (18.9% RSD, for n = 20). Homogeneity studies of the NaCl carrier without reconstitution resulted in 794 +/- 60 particles (7.5% RSD). The homogeneity of PET in the salt carrier was also evaluated directly with respect to mass of PET per vial using an ultra-micro balance. An average mass of 293 +/- 41 mu g of PET was obtained (14, % RSD for n = 14). Micrographs were recorded to demonstrate the absence of major sources of contamination of the RM components. Information about the particle size distribution and particle shapes was obtained by laser diffraction (LD) and dynamic image analysis (DIA). In addition, the identity of the PET polymer was confirmed by Raman and FT-IR spectroscopy. The RM was developed for a large-scale inter-laboratory comparison of PET particles in water (ILC). Based on the homogeneity results, the material was found to be sufficiently homogeneous to be of meaningful use in the ILC. In a 3-day process, more than 500 samples of PET particles in the NaCl carrier were prepared with good potential for further upscaling with respect to the number of vials or with other kinds of polymers. The stability of PET was not evaluated but it was deemed to be stable for the duration of the ILC.
引用
收藏
页码:385 / 397
页数:13
相关论文
共 50 条
  • [1] Preparation of a reference material for microplastics in water—evaluation of homogeneity
    John Seghers
    Elzbieta A. Stefaniak
    Rita La Spina
    Claudia Cella
    Dora Mehn
    Douglas Gilliland
    Andrea Held
    Ulf Jacobsson
    Håkan Emteborg
    Analytical and Bioanalytical Chemistry, 2022, 414 : 385 - 397
  • [2] PREPARATION AND EVALUATION OF THE HOMOGENEITY OF MILK AS A CANDIDATE REFERENCE MATERIAL FOR C-14
    RAO, RR
    COOPER, EL
    MILTON, GN
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1995, 352 (1-2): : 136 - 142
  • [3] Homogeneity evaluation of liquid reference material developed for nuclear decommissioning
    Mazanova, Monika
    Kacur, Martin
    APPLIED RADIATION AND ISOTOPES, 2023, 197
  • [4] PREPARATION AND HOMOGENEITY STUDY OF A MUSSEL CANDIDATE REFERENCE MATERIAL FOR CERTIFICATION OF ORGANOTIN COMPOUNDS
    QUEVAUVILLER, P
    CHIAVARINI, S
    CREMISINI, C
    MORABITO, R
    BIANCHI, M
    MUNTAU, H
    MIKROCHIMICA ACTA, 1995, 120 (1-4) : 281 - 287
  • [5] Use of INAA in the homogeneity evaluation of a bovine kidney candidate reference material
    Liliana Castro
    Edson G. Moreira
    Marina B. A. Vasconcellos
    Journal of Radioanalytical and Nuclear Chemistry, 2017, 311 : 1291 - 1298
  • [6] Use of INAA in the homogeneity evaluation of a bovine kidney candidate reference material
    Castro, Liliana
    Moreira, Edson G.
    Vasconcellos, Marina B. A.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2017, 311 (02) : 1291 - 1298
  • [7] Preparation of a reference material for crude oil trace elements: Study of homogeneity and stability
    Bastos Marchezi, Thabita Thiciana
    Vieira, Luiza Valli
    de Sena, Rodrigo C.
    Ribeiro de Castro, Eustaquio Vinicius
    Brandao, Geisamanda Pedrini
    Weitzel Dias Carneiro, Maria Tereza
    MICROCHEMICAL JOURNAL, 2020, 155
  • [8] Application of chemometric tools for homogeneity and stability evaluation during the preparation of a powdered milk laboratory reference material for inorganic analysis
    Santos, Adilson S.
    dos Santos, Liz O.
    dos Santos, Ivanice F.
    dos Santos, Kelly R.
    Ferreira, Sergio L. C.
    ANALYTICAL METHODS, 2020, 12 (08) : 1055 - 1063
  • [9] Preparation of a certified reference material for pesticide control in the cultivation of fruits and vegetables: a homogeneity study
    Wohlers Morelli Cardoso, Maria Helena
    da Nobrega, Armi Wanderley
    Vital, Helio de Carvalho
    Abrantes, Shirley
    CIENCIA E TECNOLOGIA DE ALIMENTOS, 2010, 30 (02): : 429 - 438
  • [10] Evaluation of the sulfur isotopic composition and homogeneity of the Soufre de Lacq reference material
    Carmody, RW
    Seal, RR
    CHEMICAL GEOLOGY, 1999, 153 (1-4) : 289 - 295