Sensitive Electrochemical Non-Enzymatic Detection of Glucose Based on Wireless Data Transmission

被引:12
|
作者
Kim, Young-Joon [1 ]
Chinnadayyala, Somasekhar R. [2 ]
Le, Hien T. Ngoc [1 ]
Cho, Sungbo [1 ,3 ]
机构
[1] Gachon Univ, Dept Elect Engn, 1342 Seongnam Daero, Seongnam 13120, South Korea
[2] Ulsan Natl Inst Sci & Technol UNIST, Dept Mech Engn, Sensors & Aerosols Lab, Ulsan 44919, South Korea
[3] Gachon Univ, Gachon Adv Inst Hlth Sci & Technol, 155 Gaetbeol Ro, Incheon 21999, South Korea
基金
新加坡国家研究基金会;
关键词
microneedle electrode array; continuous glucose monitoring; wireless data transmission; chronoamperometry; platinum black; glucose sensor; SENSOR; ELECTRODE; NANOCOMPOSITE; BIOSENSOR; ARRAY; ACID;
D O I
10.3390/s22072787
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Miniaturization and wireless continuous glucose monitoring are key factors for the successful management of diabetes. Electrochemical sensors are very versatile and can be easily miniaturized for wireless glucose monitoring. The authors report a microneedle-based enzyme-free electrochemical wireless sensor for painless and continuous glucose monitoring. The microneedles (MNs) fabricated consist of a 3 x 5 sharp and stainless-steel electrode array configuration. Each MN in the 3 x 5 array has 575 mu m x 150 mu m in height and width, respectively. A glucose-catalyzing layer, porous platinum black, was electrochemically deposited on the tips of the MNs by applying a fixed cathodic current of 2.5 mA cm(-2) for a period of 200 s. For the non-interference glucose sensing, the platinum (Pt)-black-coated MN was carefully packaged into a biocompatible ionomer, nafion. The surface morphologies of the bare and modified MNs were studied using field-emission scanning electron microscopy (FESEM) and energy-dispersive X-ray analysis (EDX). The wireless glucose sensor displayed a broad linear range of glucose (1 -> 30 mM), a good sensitivity and higher detection limit of 145.33 mu A mM(-1) cm(-2) and 480 mu M, respectively, with bare AuMN as a counter electrode. However, the wireless device showed an improved sensitivity and enhanced detection limit of 445.75, 165.83 mu A mM(-1) cm(-2) and 268 mu M, respectively, with the Pt-black-modified MN as a counter electrode. The sensor also exhibited a very good response time (2 s) and a limited interference effect on the detection of glucose in the presence of other electroactive oxidizing species, indicating a very fast and interference-free chronoamperometric response.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] NON-ENZYMATIC ELECTROCHEMICAL GLUCOSE DETECTION
    TINDALL, CE
    MCMULLEN, JK
    MORROW, DJ
    [J]. IRISH JOURNAL OF MEDICAL SCIENCE, 1985, 154 (07) : 288 - 288
  • [2] CuO nanowires based sensitive and selective non-enzymatic glucose detection
    Zhang, Yuchan
    Liu, Yixin
    Su, Liang
    Zhang, Zhonghua
    Huo, Danqun
    Hou, Changjun
    Lei, Yu
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2014, 191 : 86 - 93
  • [3] Metal oxide based non-enzymatic electrochemical sensors for glucose detection
    Dong, Qiuchen
    Ryu, Heejeong
    Lei, Yu
    [J]. ELECTROCHIMICA ACTA, 2021, 370
  • [4] A highly sensitive non-enzymatic glucose electrochemical sensor based on NiO nanohives
    Thi Oanh Vu
    Thi Xuan Chu
    Duc Hoa Nguyen
    [J]. ADVANCES IN NATURAL SCIENCES-NANOSCIENCE AND NANOTECHNOLOGY, 2021, 12 (04)
  • [5] PtAu/C based bimetallic nanocomposites for non-enzymatic electrochemical glucose detection
    Singh, Baljit
    Laffir, Fathima
    McCormac, Timothy
    Dempsey, Eithne
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2010, 150 (01): : 80 - 92
  • [6] Highly sensitive non-enzymatic electrochemical glucose biosensor based on PANI: β12 Borophene
    Tasaltin, Cihat
    Turkmen, Tulin Ates
    Tasaltin, Nevin
    Karakus, Selcan
    [J]. JOURNAL OF MATERIALS SCIENCE-MATERIALS IN ELECTRONICS, 2021, 32 (08) : 10750 - 10760
  • [7] Highly sensitive non-enzymatic electrochemical glucose biosensor based on PANI: β12 Borophene
    Cihat Taşaltın
    Tülin Ateş Türkmen
    Nevin Taşaltın
    Selcan Karakuş
    [J]. Journal of Materials Science: Materials in Electronics, 2021, 32 : 10750 - 10760
  • [8] Electrochemical Non-Enzymatic Detection of Glucose at Nanostructured Multilayer Electrode
    Maizelis, Antonina
    [J]. 2019 IEEE 39TH INTERNATIONAL CONFERENCE ON ELECTRONICS AND NANOTECHNOLOGY (ELNANO), 2019, : 404 - 408
  • [9] Non-enzymatic electrochemical sensing of glucose
    Guangfeng Wang
    Xiuping He
    Lingling Wang
    Aixia Gu
    Yan Huang
    Bin Fang
    Baoyou Geng
    Xiaojun Zhang
    [J]. Microchimica Acta, 2013, 180 : 161 - 186
  • [10] Non-enzymatic electrochemical sensing of glucose
    Wang, Guangfeng
    He, Xiuping
    Wang, Lingling
    Gu, Aixia
    Huang, Yan
    Fang, Bin
    Geng, Baoyou
    Zhang, Xiaojun
    [J]. MICROCHIMICA ACTA, 2013, 180 (3-4) : 161 - 186