All-Arthroscopic Triangular Fibrocartilage Complex Repair: Safety and Biomechanical Comparison With a Traditional Outside-in Technique in Cadavers

被引:30
|
作者
Yao, Jeffrey [1 ]
机构
[1] Stanford Univ, Med Ctr, Dept Orthopaed Surg, Hand & Upper Limb Ctr, Palo Alto, CA 94304 USA
来源
关键词
Arthroscopy; repair; triangular fibrocartilage complex; wrist; WRIST; ANATOMY; TEARS; MANAGEMENT; INJURIES; LESIONS;
D O I
10.1016/j.jhsa.2009.01.014
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose To compare the biomechanical strength and safety of an all-arthroscopic triangular fibrocartilage (TFCC) repair technique with an outside-in technique in cadavers. Methods Ten matched pairs of fresh-frozen cadaveric wrists were used for this study. The control group TFCC tears were treated with an outside-in technique using 2-0 polydioxane (PDS) sutures. The experimental group was treated with two FasT-Fix suture devices. I measured the location of the implants in relation to the neurovascular structures using a digital caliper. The strength of the repairs was then determined using a tensile testing machine with the load placed across the repair site. I compared results using the Student's t-test. Results The most volar FasT-Fix block averaged 1.8 cm from the ulnar neurovascular bundle, whereas the PDS knots averaged 1.9 cm from it. The most dorsal FasT-Fix averaged 17.1 mm from the dorsal branch of the ulnar nerve, whereas the PDS knot was 4.6 mm. The average load to failure for the FasT-Fix repairs was 3.7 N, compared with 2.4 N for the PDS repairs (p < .05). The mode of failure for the FasT-Fix implants was the suture cutting through the TFCC tissue. The mode of failure for the PDS controls varied between the suture cutting through the tissue and the knots untying. One extensor carpi ulnaris tendon was injured by the PDS technique. No tendons were injured with the FasT-Fix technique. Conclusions This all-arthroscopic technique of TFCC repair is faster and stronger than the inside-out technique and is equally safe. Benefits of this repair are decreased operative time, reduced postoperative immobilization, and decreased irritation from prominent suture knots below the skin. For these reasons, it may be desirable to perform this technique to improve patient satisfaction. (J Hand Surg 2009:34A:671-676. Copyright (c) 2009 by the American Society for Surgery of the Hand. All rights reserved.)
引用
收藏
页码:671 / 676
页数:6
相关论文
共 50 条
  • [1] All-Arthroscopic Triangular Fibrocartilage Complex Repair: Safety and Biomechanical Comparison With a Traditional Outside-in Technique in Cadavers (vol 34, pg 671, 2009)
    Yao, J.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2009, 34A (06): : 1162 - 1162
  • [2] Arthroscopic Repair of Triangular Fibrocartilage Tears: A Biomechanical Comparison of a Knotless Suture Anchor and the Traditional Outside-In Repairs
    Desai, Mihir J.
    Hutton, William C.
    Jarrett, Claudius D.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2013, 38A (11): : 2193 - 2197
  • [3] All-Arthroscopic Repair of Peripheral Triangular Fibrocartilage Complex Tears
    Yao, Jeffrey
    OPERATIVE TECHNIQUES IN SPORTS MEDICINE, 2010, 18 (03) : 168 - 172
  • [4] Arthroscopic-Assisted Outside-In Repair of Triangular Fibrocartilage Complex Tears
    Frank, Rachel M.
    Slikker, William
    Al-Shihabi, Laith
    Wysocki, Robert W.
    ARTHROSCOPY TECHNIQUES, 2015, 4 (05): : E577 - E581
  • [5] A new arthroscopic repair technique for triangular fibrocartilage complex using an intracapsular suture: an outside-in transfer all-inside repair
    Zhao, Jiasong
    Lin, Yanming
    Li, Lang
    Huang, Yong
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2023, 18 (01)
  • [6] A new arthroscopic repair technique for triangular fibrocartilage complex using an intracapsular suture: an outside-in transfer all-inside repair
    Jiasong Zhao
    Yanming Lin
    Lang Li
    Yong Huang
    Journal of Orthopaedic Surgery and Research, 18
  • [7] All-Arthroscopic Triangular Fibrocartilage Complex Ligamentoplasty. Technique and Results
    Carratala, Vicente
    Javier Lucas, Francisco
    Miranda, Ignacio
    Martinez, Cristobal
    Guisasola, Eva
    JOURNAL OF WRIST SURGERY, 2023, 12 (02) : 104 - 112
  • [8] Outside-in Repair of Peripheral Triangular Fibrocartilage Complex Tears
    Wysocki, Robert W.
    Ruch, David S.
    OPERATIVE TECHNIQUES IN SPORTS MEDICINE, 2010, 18 (03) : 163 - 167
  • [9] All-Arthroscopic Foveal Triangular Fibrocartilage Complex Tendon Reconstruction
    Tang, Camelia Q. Y.
    Jiang, Jackson K. H.
    Chin, Andrew Y. H.
    Choudhury, Muntasir M.
    JOURNAL OF WRIST SURGERY, 2024, 13 (01) : 58 - 65
  • [10] All-Inside Versus Outside-in Repair of Triangular Fibrocartilage Complex Peripheral Tears
    Fones, Lilah
    Cole, Keegan P.
    Kwok, Moody
    Gallant, Greg G.
    Tosti, Rick
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2025, 50 (01): : 96e1 - 96e6