Tungsten filled 3D printed field shaping devices for electron beam radiation therapy

被引:17
|
作者
Skinner, Lawrie [1 ]
Fahimian, Benjamin P. [1 ]
Yu, Amy S. [1 ]
机构
[1] Stanford Univ, Dept Radiat Oncol, Palo Alto, CA 94304 USA
来源
PLOS ONE | 2019年 / 14卷 / 06期
关键词
ENERGY;
D O I
10.1371/journal.pone.0217757
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose Electron radiotherapy is a labor-intensive treatment option that is complicated by the need for field shaping blocks. These blocks are typically made from casting Cerrobend alloys containing lead and cadmium. This is a highly toxic process with limited precision. This work aims to provide streamlined and more precise electron radiotherapy by 3D using printing techniques. Methods The 3D printed electron cutout consists of plastic shells filled with 2 mm diameter tungsten ball bearings. Five clinical Cerrobend defined field were compared to the planned fields by measuring the light field edge when mounted in the electron applicator on a linear accelerator. The dose transmitted through the 3D printed and Cerrobend cutouts was measured using an IC profiler ion chamber array with 6 MeV and 16 MeV beams. Dose profiles from the treatment planning system were also compared to the measured dose profiles. Centering and full width half maximum (FWHM) metrics were taken directly from the profiler software. Results The transmission of a 16MeV beam through a 12 mm thick layer of tungsten ball bearings agreed within 1% of a 15 mm thick Cerrobend block (measured with an ion chamber array). The radiation fields shaped by ball bearing filled 3D printed cutout were centered within 0.4 mm of the planned outline, whereas the Cerrobend cutout fields had shift errors of 1-3 mm, and shape errors of 0.5-2 mm. The average shift of Cerrobend cutouts was 2.3 mm compared to the planned fields (n = 5). Beam penumbra of the 3D printed cutouts was found to be equivalent to the 15 mm thick Cerrobend cutout. The beam profiles agreed within 1.2% across the whole 30 cm profile widths. Conclusions This study demonstrates that with a proper quality assurance procedure, 3D-printed cutouts can provide more accurate electron radiotherapy with reduced toxicity compared to traditional Cerrobend methods.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Novel Tungsten Filled 3D Printed Electron Beam Shaping Devices: Design and Evaluation
    Skinner, L.
    Fahimian, B.
    Yu, A.
    MEDICAL PHYSICS, 2018, 45 (06) : E141 - E141
  • [2] Dosimetric Evaluation of 3D Printed Bolus Material for Electron Beam Radiation Therapy
    Wilke, C. T.
    Ferreira, C.
    Sterling, D.
    Mathew, D. C.
    Ehler, E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E480 - E480
  • [3] Shaping success: clinical implementation of a 3D-printed electron cutout program in external beam radiation therapy
    Schulz, Joseph B.
    Gibson, Clinton
    Dubrowski, Piotr
    Marquez, Caroline M.
    Million, Lynn
    Qian, Yushen
    Skinner, Lawrie
    Yu, Amy S.
    FRONTIERS IN ONCOLOGY, 2023, 13
  • [4] FIELD SHAPING FOR ELECTRON-BEAM RADIATION-THERAPY
    PURDY, JA
    ABRATH, FG
    PEREZ, CA
    RADIOLOGY, 1979, 132 (02) : 490 - 490
  • [5] Design and production of 3D printed bolus for electron radiation therapy
    Su, Shiqin
    Moran, Kathryn
    Robar, James L.
    MEDICAL PHYSICS, 2014, 41 (08) : 2 - 2
  • [6] Design and production of 3D printed bolus for electron radiation therapy
    Su, Shiqin
    Moran, Kathryn
    Robar, James L.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (04): : 194 - 211
  • [7] Custom 3D Printed Boluses for Radiation Therapy
    Zhao, B.
    Yang, M.
    Yan, Y.
    Rahimi, A.
    Chopra, R.
    Jiang, S.
    MEDICAL PHYSICS, 2015, 42 (06) : 3188 - 3189
  • [8] 3D printed bolus for chestwall radiation therapy
    Robar, J.
    Allan, J.
    Macdonald, R. L.
    Rutledge, R.
    Joseph, T.
    Clancey, J.
    Moran, K.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S456 - S457
  • [9] FIELD SHAPING IN ELECTRON-BEAM THERAPY
    KHAN, FM
    MOORE, VC
    LEVITT, SH
    BRITISH JOURNAL OF RADIOLOGY, 1976, 49 (586): : 883 - 886
  • [10] Investigation of the possibility of shaping an electron dose field of a clinical accelerator with 3D printed polymer products
    Miloichikova, I.
    Bulavskaya, A.
    Gargioni, E.
    Grigorieva, A.
    Cherepennikov, Yu
    Belousov, D.
    Stuchebrov, S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2024, 1059