Technical Note: Dose effects of 1.5 T transverse magnetic field on tissue interfaces in MRI-guided radiotherapy

被引:49
|
作者
Chen, Xinfeng [1 ]
Prior, Phil [1 ]
Chen, Guang-Pei [1 ]
Schultz, Christopher J. [1 ]
Li, X. Allen [1 ]
机构
[1] Med Coll Wisconsin, Dept Radiat Oncol, 8701 Watertown Plank Rd, Milwaukee, WI 53226 USA
关键词
MRI-guided radiotherapy; dose effect of magnetic field; tissue interface; MONTE-CARLO; LINEAR-ACCELERATOR; SCANNER; SYSTEM; GPUMCD; IMPACT; IMRT;
D O I
10.1118/1.4959534
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The integration of MRI with a linear accelerator (MR-linac) offers great potential for high-precision delivery of radiation therapy (RT). However, the electron deflection resulting from the presence of a transverse magnetic field (TMF) can affect the dose distribution, particularly the electron return effect (ERE) at tissue interfaces. The purpose of the study is to investigate the dose effects of ERE at air-tissue and lung-tissue interfaces during intensity-modulated radiation therapy (IMRT) planning. Methods: IMRT and volumetric modulated arc therapy (VMAT) plans for representative pancreas, lung, breast, and head and neck (HN) cases were generated following commonly used clinical dose volume (DV) criteria. In each case, three types of plans were generated: (1) the original plan generated without a TMF; (2) the reconstructed plan generated by recalculating the original plan with the presence of a TMF of 1.5 T (no optimization); and (3) the optimized plan generated by a full optimization with TMF = 1.5 T. These plans were compared using a variety of DV parameters, including V-100%, D-95%, DHI [dose heterogeneity index: (D-20%-D-80%)/D-prescription], D-max, and D-1cc in OARs (organs at risk) and tissue interface. All the optimizations and calculations in this work were performed on static data. Results: The dose recalculation under TMF showed the presence of the 1.5 T TMF can slightly reduce V-100% and D-95% for PTV, with the differences being less than 4% for all but one lung case studied. The TMF results in considerable increases in D-max and D-1cc on the skin in all cases, mostly between 10% and 35%. The changes in D-max and D-1cc on air cavity walls are dependent upon site, geometry, and size, with changes ranging up to 15%. The VMAT plans lead to much smaller dose effects from ERE compared to fixed-beam IMRT in pancreas case. When the TMF is considered in the plan optimization, the dose effects of the TMF at tissue interfaces (e.g., air-cavity wall, lung-tissue interfaces, skin) are significantly reduced in most cases. Conclusions: The doses on tissue interfaces can be significantly changed by the presence of a TMF during MR-guided RT when the magnetic field is not included in plan optimization. These changes can be substantially reduced or even eliminated during VMAT/IMRT optimization that specifically considers the TMF, without deteriorating overall plan quality. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:4797 / 4802
页数:6
相关论文
共 50 条
  • [1] Dose Effects of a 1.5T Magnetic Field On Air-Tissue and Lung-Tissue Interfaces in MRI-Guided Radiotherapy
    Chen, Xinfeng
    Prior, Phillip
    Chen, Guangpei
    Schultz, Christopher
    Li, X. Allen
    MEDICAL PHYSICS, 2016, 43 (06) : 3660 - 3660
  • [2] Impact of a 1.5 T magnetic field on DNA damage in MRI-guided HDR brachytherapy
    Tambasco, Mauro
    Pang, Geordi
    Fuller, Loni
    Brescia, Erika L.
    Mardirossian, George
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 76 : 85 - 91
  • [3] A 1.5 T transverse magnetic field in radiotherapy of rectal cancer: Impact on the dose distribution
    Uilkema, Sander
    van der Heide, Uulke
    Sonke, Jan-Jakob
    Moreau, Michel
    van Triest, Baukelien
    Nijkamp, Jasper
    MEDICAL PHYSICS, 2015, 42 (12) : 7182 - 7189
  • [4] Investigation of the dose perturbation effect for therapeutic beams with the presence of a 1.5 T transverse magnetic field in magnetic resonance imaging-guided radiotherapy
    Shao, Wencheng
    Tang, Xiaobin
    Bai, Yanling
    Shu, Diyun
    Geng, Changran
    Gong, Chunhui
    Guan, Fada
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2018, 14 (01) : 184 - 195
  • [5] Simulation of tissue dependent magnetic field susceptibility effects in MRI guided radiotherapy
    Kroll, C.
    Opel, M.
    Paganelli, C.
    Kamp, F.
    Neppl, S.
    Baroni, G.
    Dietrich, O.
    Belka, C.
    Parodi, K.
    Riboldi, M.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S554 - S555
  • [6] Effects of rotating gantry on magnetic field and eddy currents in 0.35 T MRI-guided radiotherapy (MR-IGRT) system
    Gach, H. Michael
    Curcuru, Austen N.
    Kim, Taeho
    Yang, Deshan
    MEDICAL PHYSICS, 2021, 48 (11) : 7228 - 7235
  • [7] MRI-Guided Adaptive Head and Neck Radiotherapy On a 1.5T MR Linac System
    Aliotta, E.
    Hesse, J.
    Subashi, E.
    Scripes, P.
    Aristophanous, M.
    Mechalakos, J.
    Cervino, L.
    Tyagi, N.
    Zakeri, K.
    Lee, N.
    MEDICAL PHYSICS, 2022, 49 (06) : E761 - E761
  • [8] Technical Note: Cumulative dose modeling for organ motion management in MRI-guided radiation therapy
    Mirzapour, Seyed Ali
    Mazur, Thomas R.
    Harold Li, H.
    Salari, Ehsan
    Sharp, Gregory C.
    MEDICAL PHYSICS, 2021, 48 (02) : 597 - 604
  • [9] The Impact of the 1.5T Transverse Magnetic Field in Radiotherapy of Esophageal Cancer
    Li, Z.
    Yin, Y.
    MEDICAL PHYSICS, 2019, 46 (06) : E241 - E241
  • [10] Early Health Economic Analysis of 1.5T MRI-guided Radiotherapy for Localized Prostate Cancer
    Hehakaya, C.
    van Zyp, J. R. van der Voort
    Vanneste, B. G.
    Grutters, J. P.
    Grobbee, D. E.
    Verkooijen, H. M.
    Frederix, G. W.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1082 - S1083