Femur: MR imaging-guided radio-frequency ablation in porcine model - Feasibility study

被引:16
|
作者
Aschoff, AJ
Merkle, EM
Emancipator, SN
Petersilge, CA
Duerk, JL
Lewin, JS
机构
[1] Case Western Reserve Univ, Univ Hosp Cleveland, Dept Radiol, Cleveland, OH 44106 USA
[2] Case Western Reserve Univ, Univ Hosp Cleveland, Dept Pathol, Cleveland, OH 44106 USA
[3] Case Western Reserve Univ, Univ Hosp Cleveland, Dept Oncol, Cleveland, OH 44106 USA
[4] Univ Hosp Ulm, Dept Diagnost Radiol, Ulm, Germany
[5] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
关键词
animals femur; MR; femur; neoplasms; magnetic resonance (MR); guidance; radiofrequency (RF) ablation;
D O I
10.1148/radiol.2252001500
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To determine the feasibility of magnetic resonance (MR) imaging-guided and -monitored radio-frequency (RF) ablation of bone. MATERIALS AND METHODS: Seven femurs were treated in five pigs with use of a 0.2-T open MR imager. An 11-gauge bone marrow needle was percutaneously inserted into the distal femur metaphysis with MR fluoroscopy (fast-imaging with steady-state precession, or FISP, sequences) to introduce an RF electrode into the bone with further image guidance. Thermal ablation was performed for 10 minutes (90degreesC 2 [mean +/- SD]). MR follow-up was performed immediately after ablation and again at 7 and 14 days after the procedure (with contrast material-enhanced T1-weighted, T2-weighted, and fast short inversion time inversion-recovery, or STIR, sequences). The animals were sacrificed at day 14. The femurs were sliced, decalcified, and stained. image analysis was performed to measure lesion diameter and contrast-to-noise ratio (CNR) and to evaluate complications. RESULTS: Technical success was obtained in all animals. The lesion diameter perpendicular to the electrode was 15.4 mm +/- 2.7. No significant complications were noted. The thermal lesions displayed low signal intensity with a sharp rim of high signal intensity. T2-weighted images demonstrated the highest CNR and the lowest error in predicting the lesion size immediately after ablation (2.7 mm +/- 1.3). Contrast-enhanced T1-weighted images demonstrated the highest accuracy at day 14 (1.0 mm +/- 1.0). CONCLUSION: RF ablation of bone with MR imaging as the sole imaging modality is feasible and allows monitoring of the ablation. (C) RSNA, 2002.
引用
收藏
页码:471 / 478
页数:8
相关论文
共 50 条
  • [1] MR imaging-guided radio-frequency thermal ablation of the lumbar vertebrae in porcine models
    Nour, SG
    Aschoff, AJ
    Mitchell, ICS
    Emancipator, SN
    Duerk, JL
    Lewin, JS
    RADIOLOGY, 2002, 224 (02) : 452 - 462
  • [2] MR imaging-guided radio-frequency thermal ablation in the pancreas in a porcine model with a modified clinical C-arm system
    Merkle, EM
    Haaga, JR
    Duerk, JL
    Jacobs, GH
    Brambs, HJ
    Lewin, JS
    RADIOLOGY, 1999, 213 (02) : 461 - 467
  • [3] Perfusion-modulated MR imaging-guided radiofrequency ablation of the kidney in a porcine model
    Aschoff, AJ
    Sulman, A
    Martinez, M
    Duerk, JL
    Resnick, MI
    MacLennan, GT
    Lewin, JS
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (01) : 151 - 158
  • [4] MR imaging-guided radiofrequency thermal ablation in the porcine brain at 0.2 T
    Merkle, EM
    Shonk, JR
    Zheng, L
    Duerk, JL
    Lewin, JS
    EUROPEAN RADIOLOGY, 2001, 11 (05) : 884 - 892
  • [5] MR imaging-guided radiofrequency thermal ablation in the porcine brain at 0.2 T
    Elmar M. Merkle
    Jason R. Shonk
    L. Zheng
    Jeffrey L. Duerk
    Jonathan S. Lewin
    European Radiology, 2001, 11 : 884 - 892
  • [6] MR imaging-guided percutaneous tumor ablation
    Silverman, SG
    Tuncali, K
    Morrison, PR
    ACADEMIC RADIOLOGY, 2005, 12 (09) : 1100 - 1109
  • [7] Radio-frequency in vitro ablation of porcine tongues
    Hammerschmitt, N
    Huber, K
    Farin, G
    Schäller, D
    Voigtländer, M
    Bergler, W
    Hörmann, K
    HNO, 2004, 52 (01) : 18 - 24
  • [8] Imaging of the temporal evolution of hepatic tissue destruction during MR-guided radio-frequency thermal ablation in a rabbit model
    Boaz, TL
    Lewin, JS
    Chung, Y
    Duerk, JL
    Clampitt, M
    Haaga, JR
    RADIOLOGY, 1996, 201 : 1412 - 1412
  • [9] MR imaging-guided radiofrequency thermal ablation in the porcine brain at 0.2T
    Merkle, EM
    Shonk, JR
    Zheng, L
    Duerk, JL
    Lewin, JS
    RADIOLOGY, 1999, 213P : 359 - 359
  • [10] Postbiopsy bleeding in a porcine model: Reduction with radio-frequency ablation preliminary results
    Laeseke, PF
    Winter, TC
    Davis, CL
    Stevens, KR
    Johnson, CD
    Fronczak, FJ
    Webster, JG
    Lee, FT
    RADIOLOGY, 2003, 227 (02) : 493 - 499