MRI is proving to be a very useful tool for sodium quantification in animal models of stroke, ischemia, and cancer. In this work, we present the practical design of a dual-frequency RF surface coil that provides H-1 and Na-23 images of the rat head at 4 T. The dual-frequency RF surface coil comprised of a large loop tuned to the H-1 frequency and a smaller co-planar loop tuned to the Na-23 frequency. The mutual coupling between the two loops was eliminated by the use of a trap circuit inserted in the smaller coil. This independent-loop design was versatile since it enabled a separate optimisation of the sensitivity and RF field distributions of the two coils. To allow for an easy extension of this simple double-tuned coil design to other frequencies (nuclei) and dimensions, we describe in detail the practical aspects of the workbench design and MRI testing using a phantom that mimics in vivo conditions. A comparison between our independent-loop, double-tuned coil and a single-tuned Na-23 coil of equal size obtained with a phantom matching in vivo conditions, showed a reduction of the Na-23 sensitivity (about 28%) because of signal losses in the trap inductance. Typical congruent H-1 and Na-23 rat brain images showing good SNR (Na-23: brain 7, ventricular cerebrospinal fluid 11) and spatial resolution (Na-23: 1.25 x 1.25 x 5 mm(3)) are also reported. The in vivo SNR values obtained with this coil were comparable to, if not better than, other contemporary designs in the literature. (c) 2006 Elsevier Inc. All rights reserved.
引用
收藏
页码:203 / 211
页数:9
相关论文
共 42 条
  • [1] Design, construction, and use of a tapered-spiral, quadrature 1H/23Na double-tuned coil for in ovo MRI at 7 T
    Choi, Chang-Hoon
    Hong, Suk-Min
    Felder, Jörg
    Bruch, Maximilian
    Worthoff, Wieland A.
    Krause, Sandra
    Shah, N. Jon
    Medical Physics, 2024, 51 (12) : 8761 - 8767
  • [2] A double-tuned 1H/23Na dual resonator system for tissue sodium concentration measurements in the rat brain via Na-MRI
    Wetterling, Friedrich
    Tabbert, Martin
    Junge, Sven
    Gallagher, Lindsay
    Macrae, I. Mhairi
    Fagan, Andrew J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (24): : 7681 - 7695
  • [3] Numerical and Workbench Design of 2.35 T Double-Tuned (1H/23Na) Nested RF Birdcage Coils Suitable for Animal Size MRI
    Fantasia, Marco
    Galante, Angelo
    Maggiorelli, Francesca
    Retico, Alessandra
    Fontana, Nunzia
    Monorchio, Agostino
    Alecci, Marcello
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2020, 39 (10) : 3175 - 3186
  • [4] Cardiorenal 23Na MRI at 7.0 Tesla using a 4/4 channel 1H/23Na RF antenna array
    Boehmert, Laura
    Kuehne, Andre
    Waiczies, Helmar
    Wenz, Daniel
    Eigentler, Thomas Wilhelm
    Funk, Stephanie
    von Knobelsdorff-Brenkenhoff, Florian
    Schulz-Menger, Jeanette
    Nagel, Armin M.
    Seeliger, Erdmann
    Niendorf, Thoralf
    MAGNETIC RESONANCE IN MEDICINE, 2019, 82 (06) : 2343 - 2356
  • [5] Design and testing of a 1.5 Tesla double-tuned (1H/31P) RF surface coil with intrinsic geometric isolation
    Alfonsetti, Maria
    Sotgiu, Antonello
    Alecci, Marcello
    MEASUREMENT, 2010, 43 (09) : 1266 - 1276
  • [6] Quantitative 23Na MRI of Human Knee Cartilage Using Dual-Tuned 1H/23Na Transceiver Array Radiofrequency Coil at 7 Tesla
    Moon, Chan Hong
    Kim, Jung-Hwan
    Zhao, Tiejun
    Bae, Kyongtae Ty
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 38 (05) : 1063 - 1072
  • [7] A dual-tuned multichannel bilateral RF coil for 1H/23Na breast MRI at 7 T
    Ianniello, Carlotta
    Madelin, Guillaume
    Moy, Linda
    Brown, Ryan
    MAGNETIC RESONANCE IN MEDICINE, 2019, 82 (04) : 1566 - 1575
  • [8] A double-tuned 1H/23Na resonator allows 1H-guided 23Na-MRI in ischemic stroke patients in one session
    Neumaier-Probst, Eva
    Konstandin, Simon
    Ssozi, Judith
    Groden, Christoph
    Hennerici, Michael
    Schad, Lothar R.
    Fatar, Marc
    INTERNATIONAL JOURNAL OF STROKE, 2015, 10 : 56 - 61
  • [9] Double-Tuned Surface 1H–23Na Radio Frequency Coils at 7 T: Comparison of Three Decoupling Methods
    Francesca Maggiorelli
    Eddy B. Boskamp
    Gianluigi Tiberi
    Alessandra Retico
    Mark R. Symms
    Michela Tosetti
    Fraser Robb
    Applied Magnetic Resonance, 2019, 50 : 649 - 661
  • [10] Design of a Folded, Double-Tuned Loop Coil for 1H/X-Nuclei MRI Applications
    Hong, Suk-Min
    Choi, Chang-Hoon
    Shah, N. Jon
    Felder, Jorg
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2023, 42 (05) : 1424 - 1430