Staff eye lens dose in interventional radiology and cardiology in Finland

被引:6
|
作者
Pekkarinen, Antti [1 ,2 ,3 ,4 ,5 ]
Lindholm, Carita [1 ]
Kortesniemi, Mika [2 ,3 ,4 ]
Siiskonen, Teemu [1 ]
机构
[1] Radiat & Nucl Safety Author STUK, POB 14, Helsinki 00811, Finland
[2] Univ Helsinki, HUS Diagnost Ctr, Haartmaninkatu 4, Helsinki 00290, Finland
[3] Helsinki Univ Hosp, Haartmaninkatu 4, Helsinki 00290, Finland
[4] Univ Helsinki, Dept Phys, Helsinki, Finland
[5] Kymenlaakso Cent Hosp, Dept Med Phys, Kymsote, Kotkantie 41, Kotka 48210, Finland
关键词
PROTECTION; DOSIMETERS;
D O I
10.1016/j.ejmp.2022.04.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The aim of this study was to investigate the eye lens and whole-body radiation doses to interventional radiology and cardiology staff in two Finnish hospitals. Methods: Simultaneous measurements of personal dose equivalent quantities H-p(3) and H-p(10) were conducted in clinical conditions during different radiological and cardiological interventional procedures. In order to study the feasibility to estimate eye lens dose with H-p(10) measured over the protective apron or thyroid shield, the ratio between measured H-p(3) and H-p(10) was investigated. Results and conclusions: Applying the obtained ratio on H-p(10) records from national dose register showed that only a small number of interventional radiologists and cardiologists in Finland may exceed eye lens equivalent dose levels of 20 mSv per year or 100 mSv in five consecutive years, but likely do not exceed 50 mSv in a single year. For the most Finnish interventionalists, the eye lens dose is well below 10 mSv per year. Nurses and radiographers assisting in interventions are, on average, less exposed than interventionalists, and will not exceed 20 mSv per year. Based on our results, H-p(10) measured over the protective apron or thyroid shield provides a conservative estimate of the eye lens dose for interventional radiologists and cardiologists, provided that appropriate protective glasses are used.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 50 条
  • [1] Assessment of the occupational eye lens dose for clinical staff in interventional radiology, cardiology and neuroradiology
    Omar, Artur
    Kadesjo, Nils
    Palmgren, Charlotta
    Marteinsdottir, Maria
    Segerdahl, Tony
    Fransson, Annette
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2017, 37 (01) : 145 - 159
  • [2] EYE LENS DOSE IN INTERVENTIONAL CARDIOLOGY
    Principi, S.
    Delgado Soler, C.
    Ginjaume, M.
    Beltran Vilagrasa, M.
    Rovira Escutia, J. J.
    Duch, M. A.
    [J]. RADIATION PROTECTION DOSIMETRY, 2015, 165 (1-4) : 289 - 293
  • [3] Exposure of the eye lens and brain for interventional cardiology staff
    Kidon, Joanna
    Polaczek-Grelik, Kinga
    Zurek, Przemyslaw
    Wojakowski, Wojciech
    Ochala, Andrzej
    [J]. POSTEPY W KARDIOLOGII INTERWENCYJNEJ, 2021, 17 (03): : 298 - 304
  • [4] A study of the dose distribution in the region of the eye lens and extremities for staff working in interventional cardiology
    Domienik, J.
    Brodecki, M.
    Rusicka, D.
    [J]. RADIATION MEASUREMENTS, 2012, 47 (02) : 130 - 138
  • [5] EXTREMITY AND EYE LENS DOSES OF INTERVENTIONAL RADIOLOGY AND CARDIOLOGY WORKERS
    Urboniene, Ausra
    Griciene, Birute
    [J]. MEDICAL PHYSICS IN THE BALTIC STATES, 2010, (02): : 109 - 111
  • [6] Eye lens dosimetry in interventional cardiology: results of staff dose measurements and link to patient dose levels
    Antic, V.
    Ciraj-Bjelac, O.
    Rehani, M.
    Aleksandric, S.
    Arandjic, D.
    Ostojic, M.
    [J]. RADIATION PROTECTION DOSIMETRY, 2013, 154 (03) : 276 - 284
  • [7] Recommendations to reduce extremity and eye lens doses in interventional radiology and cardiology
    Carinou, E.
    Brodecki, M.
    Domienik, J.
    Donadille, L.
    Koukorava, C.
    Krim, S.
    Nikodemova, D.
    Ruiz-Lopez, N.
    Sans-Merce, M.
    Struelens, L.
    Vanhavere, F.
    [J]. RADIATION MEASUREMENTS, 2011, 46 (11) : 1324 - 1329
  • [8] A practical method for routine eye lens dosimetry of staff in interventional radiology
    Bergans, Niki
    Vandermaesen, Alexander
    Vanheule, Jeroen
    Bogaerts, Ria
    [J]. RADIATION PROTECTION DOSIMETRY, 2023, 199 (15-16) : 1779 - 1784
  • [9] Radiation dose to physicians' eye lens during interventional radiology
    Bahruddin, N. A.
    Hashim, S.
    Karim, M. K. A.
    Sabarudin, A.
    Ang, W. C.
    Salehhon, N.
    Bakar, K. A.
    [J]. 13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [10] COMPARISON OF TWO TYPES EYE LENS DOSEMETERS USED IN INTERVENTIONAL CARDIOLOGY AND RADIOLOGY
    Martincik, Jiri
    Cechak, Tomas
    Urban, Tomas
    Tamasova, Jana
    Sikova, Lenka
    Jakubcova, Iva
    Balogova, Zdenka
    [J]. RADIATION PROTECTION DOSIMETRY, 2022, 198 (9-11) : 580 - 585