A computer system for the numerical analysis of nonstress tests

被引:156
|
作者
Pardey, J
Moulden, M
Redman, CWG
机构
[1] Oxford Instruments Med Ltd, Technol Dev Grp, Surrey GU22 9JU, England
[2] Univ Oxford, John Radcliffe Hosp, Dept Obstet & Gynaecol, Oxford OX3 9DU, England
关键词
antepartum; computer analysis; fetal heart rate; nonstress test; Sonicaid FetalCare;
D O I
10.1067/mob.2002.122447
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
The development and clinical validation of a computer system for the numerical analysis of nonstress tests is reviewed, and recent improvements are reported. The analysis was developed by using a database of 73,802 nonstress test readings to provide a numerical definition of reactivity that is tailored to the gestational age of the fetus and independent of the presence of accelerations. When used at the bedside, the analysis minimizes monitoring time by alerting the operator when monitoring can be safely stopped because the fetal heart rate is normal. It also detects potentially sinister sinusoidal patterns and improves the quality of nonstress test readings by quantifying signal loss and, if significant, alerting the operator.