Improving the stability of cryogenic current comparator setups

被引:23
|
作者
Drung, D. [1 ]
Goetz, M. [2 ]
Pesel, E. [2 ]
Storm, J. -H [1 ]
Assmann, C. [1 ]
Peters, M. [1 ]
Schurig, Th [1 ]
机构
[1] Phys Tech Bundesanstalt, D-10587 Berlin, Germany
[2] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
来源
SUPERCONDUCTOR SCIENCE & TECHNOLOGY | 2009年 / 22卷 / 11期
关键词
D O I
10.1088/0953-2048/22/11/114004
中图分类号
O59 [应用物理学];
学科分类号
摘要
We have realized an improved resistance calibration setup based on a cryogenic current comparator (CCC). The comparator, with 18 windings and 4647 turns in total, is well-suited for all the necessary comparisons with a 100 Omega standard resistor with the quantum Hall effect and inside the range of standard resistance values from 1 Omega to 1 M Omega. The new state-of-the-art setup is equipped with a low-noise dc SQUID (superconducting quantum interference device) and a digital double current source. Proper damping of the CCC resonance, careful electronics design ('box-in-box') and application of internal wideband feedback to the SQUID sensor improve the dynamic stability considerably and lower the overall measurement time as compared against our previous setup.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Aspects of Application and Calibration of a Binary Compensation Unit for Cryogenic Current Comparator Setups
    Drung, Dietmar
    Goetz, Martin
    Pesel, Eckart
    Barthelmess, Henry J.
    Hinnrichs, Colmar
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2013, 62 (10) : 2820 - 2827
  • [2] Stability and Performance of the Binary Compensation Unit for Cryogenic Current Comparator Bridges
    Gotz, Martin
    Drung, Dietmar
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2017, 66 (06) : 1467 - 1474
  • [3] Ultimate current resolution of a cryogenic current comparator
    Sesé, J
    Camón, A
    Rillo, C
    Rietveld, G
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1999, 48 (06) : 1306 - 1313
  • [4] CRYOGENIC CURRENT COMPARATOR WITH INCREASED RESOLUTION
    VANDERWEL, W
    MOOIJ, JE
    HARMANS, CJPM
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (04): : 624 - 626
  • [5] Cryogenic current comparator implementation at CENAM
    Sánchez, C
    Avilés, D
    Hernández, D
    Durán, G
    [J]. 2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 96 - 97
  • [6] LEAKAGE CURRENT DETECTION IN CRYOGENIC CURRENT COMPARATOR BRIDGES
    ELMQUIST, RE
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1993, 42 (02) : 167 - 169
  • [7] On forward estimate of the sensitivity of a cryogenic current comparator
    Frantsuz, ET
    [J]. METROLOGIA, 1998, 35 (06) : 847 - 851
  • [8] Simplified calculus for the design of a cryogenic current comparator
    Sesé, J
    Bartolomé, E
    Camón, A
    Flokstra, J
    Rietveld, G
    Rillo, C
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2003, 52 (02) : 612 - 616
  • [9] Method for the Calibration of Cryogenic Current Comparator Ratios
    Lu, Yunfeng
    Zhao, Jianting
    He, Qing
    [J]. 2020 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM), 2020,
  • [10] Thin-film cryogenic current comparator
    Seppä, H
    Satrapinksi, A
    Kiviranta, M
    Virkki, V
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1999, 48 (02) : 365 - 369