Early failure for wear after ceramic-on-highly cross-linked polyethylene total hip arthroplasty: a case report

被引:2
|
作者
Li, Chenkai [1 ]
Zhang, Haining [1 ]
机构
[1] Qingdao Univ, Dept Joint Surg, Affiliated Hosp, Qingdao 266000, Shandong, Peoples R China
基金
中国国家自然科学基金;
关键词
Dysplasia dysfunction of the hip; Ceramic-on-highly cross-linked polyethylene; Total hip arthroplasty; Case report; YOUNGER; BEARING;
D O I
10.1186/s12891-020-03697-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
BackgroundHighly cross-linked polyethylene (HXLPE) enhances the anti-wear characteristics of the conventional polyethylene (PE). Early failure for wear after ceramic-on-highly cross-linked polyethylene (CoHXLPE) total hip arthroplasty (THA) is extremely rare.Case presentationWe described the case of a 60-year-old man who underwent right CoHXLPE THA because of the developmental dysplasia hip (DDH) complained pain 32months after this procedure. Plain radiographs showed that eccentric wear existed at the polyethylene insert. However, the patient refused surgery at that time and did not stop weight-bearing. The right hip pain continued for 7months. Plain radiographs of the pelvis showed that the HXLPE liner was penetrated and partial inner wall of acetabular shell was worn. Acetabular cup revision was performed, and the ceramic head and HXLPE were exchanged.ConclusionsDifficult reduction during primary THA, especially for DDH, can result in higher abductor tension, which may lead to early eccentric wear of the prosthesis. Whenever eccentric wear of HXLPE liner was found, weight-bearing must be stopped to avoid the accelerated wear and adverse reactions to metal debris (ARMD).
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Early failure for wear after ceramic-on-highly cross-linked polyethylene total hip arthroplasty: a case report
    Chenkai Li
    Haining Zhang
    [J]. BMC Musculoskeletal Disorders, 21
  • [2] Ceramic on highly cross-linked polyethylene in cementless total hip arthroplasty
    Kong, J. -S.
    Moon, K. -H.
    [J]. BIOCERAMICS AND ALTERNATIVE BEARINGS IN JOINT ARTHROPLASTY, 2007, : 67 - 70
  • [3] A randomised prospective evaluation of ceramic-on-ceramic and ceramic-on-highly cross-linked polyethylene bearings in the same patients with primary cementless total hip arthroplasty
    Young-Hoo Kim
    Jang-Won Park
    Sourabh S. Kulkarni
    Yoon-Hong Kim
    [J]. International Orthopaedics, 2013, 37 : 2131 - 2137
  • [4] A randomised prospective evaluation of ceramic-on-ceramic and ceramic-on-highly cross-linked polyethylene bearings in the same patients with primary cementless total hip arthroplasty
    Kim, Young-Hoo
    Park, Jang-Won
    Kulkarni, Sourabh S.
    Kim, Yoon-Hong
    [J]. INTERNATIONAL ORTHOPAEDICS, 2013, 37 (11) : 2131 - 2137
  • [5] Early failure of a highly cross-linked polyethylene inlay after total hip arthroplasty probably due to insufficient irradiation
    Wahl, Peter
    Mossu-Haas, Christine
    Dommann-Scherrer, Corina
    Wei, Kongchang
    Eschbach, Lukas
    Gehr, Peter
    Benninger, Emanuel
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2022, 236 (12) : 1711 - 1719
  • [6] Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty?
    Glyn-Jones, Sion
    Saac, Sherif
    Hauptfleisch, Jennifer
    McLardy-Smith, Peter
    Murray, David W.
    Singh, Harinderjit
    [J]. JOURNAL OF ARTHROPLASTY, 2008, 23 (03): : 337 - 343
  • [7] Wear Rate of Highly Cross-Linked Polyethylene in Total Hip Arthroplasty A Randomized Controlled Trial
    McCalden, Richard W.
    MacDonald, Steven J.
    Rorabeck, Cecil H.
    Bourne, Robert B.
    Chess, David G.
    Charron, Kory D.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (04): : 773 - 782
  • [8] In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty
    Manning, DW
    Chiang, PP
    Martell, JM
    Galante, JO
    Harris, WH
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (07): : 880 - 886
  • [9] Durability of Highly Cross-Linked Polyethylene in Total Hip and Total Knee Arthroplasty
    Dion, Neil T.
    Bragdon, Charles
    Muratoglu, Orhun
    Freiberg, Andrew A.
    [J]. ORTHOPEDIC CLINICS OF NORTH AMERICA, 2015, 46 (03) : 321 - +
  • [10] Highly cross-linked polyethylene in total hip arthroplasty, present and future
    Mori, Toshiharu
    Tsukamoto, Manabu
    [J]. ANNALS OF JOINT, 2018, 3 (08):