摘要
This paper aims to provide guidance and a framework for commissioning tests and tolerances for the ExacTrac Dynamic image-guided and surface-guided radiotherapy (SGRT) system. ExacTrac Dynamic includes a stereoscopic X-ray system, a structured light projector, stereoscopic cameras, thermal camera for SGRT, and has the capability to track breath holds and internal markers. The system provides fast and accurate image guidance and intrafraction guidance for stereotactic radiosurgery and stereotactic ablative radiotherapy. ExacTrac Dynamic was commissioned on a recently installed Elekta Versa HD. Commissioning tests are described including safety, isocenter calibration, dosimetry, image quality, data transfer, SGRT stability, SGRT localization, gating, fusion, implanted markers, breath hold, and end-to-end testing. Custom phantom designs have been implemented for assessment of the deep inspiration breath-hold workflow, the implanted markers workflow, and for gating tests where remote-controlled movement of a phantom is required. Commissioning tests were all found to be in tolerance, with maximum translational and rotational deviations in SGRT of 0.3 mm and 0.4 degrees, respectively, and X-ray image fusion reproducibility standard deviation of 0.08 mm. Tolerances were based on published documents and upon the performance characteristics of the system as specified by the vendor. The unique configuration of ExacTrac Dynamic requires the end user to design commissioning tests that validate the system for use in the clinical implementation adopted in the department. As there are multiple customizable workflows available, tests should be designed around these workflows, and can be ongoing as workflows are progressively introduced into departmental procedures.
引用
收藏
页码:398 / 408
页数:11
相关论文
共 50 条
  • [1] Commissioning and initial stereotactic ablative radiotherapy experience with Vero
    Solberg, Timothy D.
    Medin, Paul M.
    Ramirez, Ezequiel
    Ding, Chuxiong
    Foster, Ryan D.
    Yordy, John
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02): : 205 - 225
  • [2] Commissioning a system for stereotactic radiosurgery/radiotherapy of the head and neck
    Clark, BG
    Robar, J
    Vollans, E
    Audet, C
    Sheehan, F
    RADIOSURGERY, VOL 4, 2002, 4 : 228 - 238
  • [3] Delivery validation of VMAT stereotactic ablative body radiotherapy at commissioning
    Olding, T.
    Alexander, K. M.
    Jechel, C.
    Nasr, A. T.
    Joshi, C.
    8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573
  • [4] Implementation of Stereotactic Ablative Radiotherapy (Stereotactic Body Radiotherapy)
    Kirkbride, P.
    Hoskin, P. J.
    CLINICAL ONCOLOGY, 2012, 24 (09) : 627 - 628
  • [5] Comparing the Accuracy of AlignRT and ExacTrac at Intrafractional Time Points During Cranial Stereotactic Radiosurgery and Stereotactic Radiotherapy Treatments
    Wolken, T.
    Junell, S.
    MEDICAL PHYSICS, 2018, 45 (06) : E224 - E225
  • [6] Further Considerations in Regards to Commissioning of Spine Stereotactic Ablative Body Radiotherapy
    Barker, C. L.
    King, J.
    Colaco, R.
    CLINICAL ONCOLOGY, 2023, 35 (01)
  • [7] Commissioning of a micro-multileaf collimator for conformal stereotactic radiosurgery and radiotherapy
    Wurm, RE
    Cosgrove, VP
    Schlenger, L
    Kaiser, A
    Zeunert, G
    Dinges, S
    Bohsung, J
    Groll, J
    Pfaender, M
    Bauer, S
    Leonhardt, G
    Jahn, U
    Stuschke, A
    Budach, V
    CONTROVERSIES IN NEURO-ONCOLOGY, 1999, 33 : 64 - 77
  • [8] Stereotactic Radiosurgery for trigeminal neuralgia using Exactrac Dynamic. First experience
    Corro Verde, U. A.
    Navarrete Solano, P. A.
    Fabregat Borras, R.
    Raba Diez, J. I.
    Canon Garcia, V.
    Vazquez Rodriguez, J. A.
    Albendea Roch, J.
    Gutierrez, M.
    Astudillo Olalla, R.
    Rivero Perez, A. L.
    Arrojo Alvarez, E.
    Pinto Guevera, F.
    Pacheco Baldor, M. T.
    Diaz De Cerio, I.
    Prada Gomez, P. J.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S960 - S961
  • [9] Stereotactic Proton Ablative Radiosurgery of the Spine
    Dougherty, M.
    Mullikin, T. C.
    Park, S. S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E552 - E553
  • [10] A feasibility study of stereotactic radiosurgery/stereotactic body radiotherapy/stereotactic ablative radiotherapy practice using tomoedge in helical tomotherapy for lung, liver, and spine targets
    Sresty, N. V. N. Madhusudhana
    Raju, A. Krishnam
    Kumar, G. Deleep
    Rohit, S.
    Reddy, B. Nagarjuna
    Sahithya, V. C.
    Reddy, B. Devender
    Mohd, Yakub
    Rushdi, Tasneem
    Bajwa, Harjoth
    Aparna, S.
    JOURNAL OF MEDICAL PHYSICS, 2021, 46 (03) : 204 - 210