Evaluation of a novel high-resolution mapping system for catheter ablation of ventricular arrhythmias

被引:47
|
作者
Viswanathan, Karthik [1 ]
Mantziari, Lilian [1 ]
Butcher, Charles [1 ]
Hodkinson, Emily [1 ]
Lim, Eric [1 ]
Khan, Habib [1 ]
Panikker, Sandeep [1 ]
Haldar, Shouvik [1 ]
Jarman, Julian W. E. [1 ]
Jones, David G. [1 ]
Hussain, Wajid [1 ]
Foran, John P. [1 ]
Markides, Vias [1 ]
Wong, Tom [1 ]
机构
[1] Imperial Coll, Royal Brompton & Harefield NHS Fdn Trust, NIHR Cardiovasc Biomed Res Unit, London, England
关键词
Ventricular tachycardia; High-resolution mapping; High-density mapping; Catheter ablation; Minibasket catheter; Automatic mapping system; TACHYCARDIA;
D O I
10.1016/j.hrthm.2016.11.018
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND The mapping of ventricular arrhythmias in humans using a minibasket 64-electrode catheter paired with a novel automatic mapping system (Rhythmia) has not been evaluated. OBJECTIVE The purpose of this study was to evaluate the safety and efficacy of mapping ventricular arrhythmias and clinical outcomes after ablation using this system. METHODS Electroanatomic maps for ventricular arrhythmias were obtained during 20 consecutive procedures in 19 patients (12 with ventricular tachycardia [VT] and 2 with ventricular ectopy [VE]). High-density maps were acquired using automatic beat acceptance and automatic system annotation of electrograms. RESULTS Forty-seven electroanatomic maps (including 3 right ventricular and 9 epicardial maps) were obtained. Left ventricular endocardial mapping by transseptal (n = 13) and/or transaortic (n = 11) access was safe with no complications related to the minibasket catheter. VT substrate maps (n = 14; median 10,184 points) consistently demonstrated late potentials with high resolution. VT activation maps (n = 25; median 6401 points) obtained by automatic annotation included 7 complete maps (covering >= 90% of the tachycardia cycle length) in 5 patients in whom the entire VT circuit was accurately visualized. VE timing maps (n = 8) successfully localized the origin of VEs in all, with all accepted beats consistent with clinical VEs. Over a median follow-up of 10 months, no arrhythmia recurrence was noted in 75% after VT ablation and 86% after VE ablation. CONCLUSION In this first human experience for ventricular arrhythmias using this system, ultra-high-density maps were created rapidly and safely, with a reliable automatic annotation of VT and consistent recording of abnormal electrograms. Medium-term outcomes after ablation were encouraging. Further larger studies are needed to validate these findings.
引用
收藏
页码:176 / 183
页数:8
相关论文
共 50 条
  • [1] The utility, efficacy and safety of a new rapid high-resolution mapping system in the catheter ablation of atrial and ventricular arrhythmias in humans
    Mantziari, L.
    Butcher, C.
    Kontogeorgis, A.
    Panikker, S.
    Roy, K.
    Markides, V.
    Wong, T.
    EUROPEAN HEART JOURNAL, 2015, 36 : 1058 - 1059
  • [2] Catheter ablation of a left posterior fascicular ventricular tachycardia guided by a novel high-resolution multipolar mapping catheter
    Giulio Conte
    Maria Luce Caputo
    François Regoli
    Tiziano Moccetti
    Angelo Auricchio
    Journal of Interventional Cardiac Electrophysiology, 2017, 49 : 101 - 102
  • [3] Catheter ablation of a left posterior fascicular ventricular tachycardia guided by a novel high-resolution multipolar mapping catheter
    Conte, Giulio
    Caputo, Maria Luce
    Regoli, Francois
    Moccetti, Tiziano
    Auricchio, Angelo
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2017, 49 (01) : 101 - 102
  • [4] UTILITY OF HIGH-RESOLUTION MAPPING TO GUIDE ABLATION OF VENTRICULAR ARRHYTHMIAS FROM THE AORTIC CUSPS
    Matto, Faisal
    Giacomino, Bria
    Venugopal, Dev
    Bhave, Prashant
    Rhodes, Troy
    Mazur, Alexander
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 75 (11) : 290 - 290
  • [5] EPICARDIAL VENTRICULAR TACHYCARDIA ABLATION GUIDED BY A NOVEL HIGH-RESOLUTION CONTACT MAPPING SYSTEM: A MULTICENTRE STUDY
    Shi, Rui
    Chen, Zhong
    Kontogeorgis, Andrianos
    Sacher, Fred
    Della Bella, Paolo
    Bisceglia, Caterina
    Martin, Ruairidh
    Meyer, Christian
    Willems, Stephan
    Markides, Vias
    Maury, Philippe
    Wong, Tom
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 73 (09) : 325 - 325
  • [6] Catheter ablation of ventricular outflow tract arrhythmias using an electroanatomic mapping system
    Horlitz, M
    Schley, P
    Sause, A
    Müller, M
    Shin, DI
    Klein, RM
    Gülker, H
    MEDIZINISCHE KLINIK, 2005, 100 (10) : 611 - 616
  • [7] Epicardial Ventricular Tachycardia Ablation Guided by a Novel High-Resolution Contact Mapping System: A Multicenter Study
    Shi, Rui
    Chen, Zhong
    Kontogeorgis, Andrianos
    Sacher, Frederic
    Della Bella, Paolo
    Bisceglia, Caterina
    Martin, Ruairidh
    Meyer, Christian
    Willems, Stephan
    Markides, Vias
    Maury, Philippe
    Wong, Tom
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2018, 7 (21):
  • [8] Evaluation and Catheter Ablation of Ventricular Arrhythmias in Cardiac Sarcoidosis
    Ezzeddine, Fatima M.
    Tan, Nicholas
    Siontis, Konstantinos C.
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (22)
  • [9] Catheter Ablation for Ventricular Arrhythmias
    Nof, Eyal
    Stevenson, William G.
    John, Roy M.
    ARRHYTHMIA & ELECTROPHYSIOLOGY REVIEW, 2013, 2 (01) : 45 - 52
  • [10] Catheter Ablation of Ventricular Arrhythmias
    Shivkumar, Kalyanam
    NEW ENGLAND JOURNAL OF MEDICINE, 2019, 380 (16): : 1555 - 1564