The Calibration Laboratory (LCI) at the Instituto de Pesquisas Energeticas e Nucleares (IPEN) is going to establish a Bohm extrapolation chamber as a primary standard system for the dosimetry and calibration of beta radiation sources and detectors. This chamber was already tested in beta radiation beams with an aluminized Mylar entrance window, and now, it was characterized with an original Hostaphan entrance window. A comparison between the results of the extrapolation chamber with the two entrance windows was performed. The results showed that this extrapolation chamber presents the same effectiveness in beta radiation fields as a primary standard system with both entrance windows, showing that any one of them may be utilized. (C) 2014 Elsevier Ltd. All rights reserved.