A simple cryogenic holder for tensile testing of soft biological tissues

被引:13
|
作者
Lepetit, J [1 ]
Favier, R
Grajales, A
Skjervold, PO
机构
[1] INRA, Biophys Res Lab, F-63122 St Genes Champanelle, France
[2] Univ Autonoma San Luis Potosi, Fac Ciencias Quim, San Luis Potosi 78210, SLP, Mexico
[3] Agr Univ Norway, Dept Agr Engn, N-1432 As, Norway
关键词
cryogenic grip; muscle; fish; tensile test; mechanical properties;
D O I
10.1016/j.jbiomech.2003.08.006
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
To overcome the difficulty of gripping soft biological materials for tensile test, a simple inexpensive cryogenic holder was developed which allows rapid (3 min) preparation of samples. It is made of 6 parts, built in a bakelite cloth, which is an excellent thermal isolant, and is used with rectangular (8 x 10(-2) x 10(-2) x 10-2 m) samples. The holder with the sample inside is completely immersed in liquid nitrogen for 50 s. This duration allows the freezing of the sample ends on a 10(-2) M length and gives a very flat freezing surface throughout the sample cross section. The 6 x 10(-2) M central part of the sample remained at ambient temperature. Two parts of the holder help maintain the sample until its ends are vertically gripped in the tensile machine thus avoiding any sample deformation during this step. No pressure was applied on the frozen part of the sample by grips of the tensile machine and this avoids breaks in this region. The sample is fixed by adhesion forces (> 1 kN) between its frozen parts and 2 pieces of the holder. The procedure has been successfully tested with bovine and salmon muscle sample s and results show tensile breaks randomly distributed in the unfrozen region of the samples. Particular attention has been paid to obtain a very flat freezing surface so that the axial strain is equal throughout the sample and therefore any strain-related mechanical parameters can be accurately determined. The dimensions of the holder can be easily modified to fit other sample geometries and can be used with other biological materials. (C) 2003 Elsevier Ltd. All rights reserved.
引用
收藏
页码:557 / 562
页数:6
相关论文
共 50 条
  • [1] Simple shearing of soft biological tissues
    Horgan, Cornelius O.
    Murphy, Jeremiah G.
    PROCEEDINGS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2011, 467 (2127): : 760 - 777
  • [2] A magnetically actuated, optically sensed tensile testing method for mechanical characterization of soft biological tissues
    Rosalia, Luca
    Hallou, Adrien
    Cochrane, Laurence
    Savin, Thierry
    SCIENCE ADVANCES, 2023, 9 (02)
  • [3] Dual cryogenic fixation for mechanical testing of soft musculoskeletal tissues
    Ramachandran, N
    Koike, Y
    Poitras, P
    Backman, D
    Uhthoff, HK
    Trudel, G
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (10) : 1792 - 1795
  • [4] Characterization of mechanical properties of soft tissues using sub-microscale tensile testing and 3D-Printed sample holder
    Kang, John S.
    Navindaran, Kishev
    Phillips, J.
    Kenny, K.
    Moon, Kee S.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2023, 138
  • [5] Method of testing very soft biological tissues in compression
    Miller, K
    JOURNAL OF BIOMECHANICS, 2005, 38 (01) : 153 - 158
  • [6] Effect of fiber exclusion in uniaxial tensile tests of soft biological tissues
    Breslavsky, Ivan
    Franchini, Giulio
    Amabili, Marco
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2020, 112
  • [7] Biomechanical Characterization of Human Soft Tissues Using Indentation and Tensile Testing
    Griffin, Michelle
    Premakumar, Yaami
    Seifalian, Alexander
    Butler, Peter Edward
    Szarko, Matthew
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2016, (118):
  • [8] CRYOGENIC HOMOGENIZATION OF BIOLOGICAL TISSUES
    ZEISLER, R
    LANGLAND, JK
    HARRISON, SH
    ANALYTICAL CHEMISTRY, 1983, 55 (14) : 2431 - 2434
  • [9] Advanced 3D-Printed Capstan Clamping System for Accurate Uniaxial Tensile Testing of Biological Soft Tissues
    Horvath, Paul
    Glaeser, Franz
    Antipova, Veronica
    Klug, Corina
    Lin, Alvin C.
    Scholze, Mario
    Hammer, Niels
    APPLIED SCIENCES-BASEL, 2024, 14 (24):
  • [10] Indentation Versus Tensile Measurements of Young's Modulus for Soft Biological Tissues
    McKee, Clayton T.
    Last, Julie A.
    Russell, Paul
    Murphy, Christopher J.
    TISSUE ENGINEERING PART B-REVIEWS, 2011, 17 (03) : 155 - 164