Precision high-value resistance scaling with a two-terminal cryogenic current comparator

被引:16
|
作者
Hernandez-Marquez, F. L. [1 ]
Bierzychudek, M. E. [2 ]
Jones, G. R., Jr. [3 ]
Elmquist, R. E. [3 ]
机构
[1] Ctr Nacl Metrol, Queretaro 76900, Mexico
[2] Inst Nacl Tecnol Ind, Buenos Aires, DF, Argentina
[3] NIST, Gaithersburg, MD 20899 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2014年 / 85卷 / 04期
关键词
M-OMEGA; SERIES; SYSTEM; BRIDGE;
D O I
10.1063/1.4869240
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe a cryogenic two-terminal high-resistance bridge and its application in precision resistance scaling from the quantized Hall resistance (QHR) at R-H = R-K/2 = 12 906.4035 Omega to decade resistance standards with values between 1 M Omega and 1 G Omega. The design minimizes lead resistance errors with multiterminal connections to the QHR device. A single variable voltage source and resistive ratio windings are utilized to achieve excellent dynamic stability, which is not readily obtained in low-current measurements with conventional cryogenic current comparators (CCCs). Prototypes of this bridge have been verified by a successful international comparison of high-resistance scaling using two-terminal CCCs in the national metrology institutes of Argentina, Mexico, and the United States. (C) 2014 AIP Publishing LLC.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Uncertainty Evaluation in a Two-Terminal Cryogenic Current Comparator
    Bierzychudek, M. E.
    Elmquist, R. E.
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2009, 58 (04) : 1170 - 1175
  • [2] Robust Control of a Two-Terminal Cryogenic Current Comparator
    Bierzychudek, M. E.
    Sanchez-Pena, R. S.
    Tonina, Alejandra
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2013, 62 (06) : 1736 - 1742
  • [3] Uncertainty evaluation in a two-terminal cryogenic current comparator
    Bierzychudek, M. E.
    Elmquist, R. E.
    [J]. 2008 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2008, : 152 - +
  • [4] SIM.EM-S10 RMO COMPARISON FINAL REPORT High value resistance comparison with two-terminal cryogenic current comparators
    Bierzychudek, Marcos E.
    Elmquist, Randolph
    Hernandez, Felipe
    [J]. METROLOGIA, 2014, 51
  • [5] High resistance scaling from 10kΩ and QHR standards using a cryogenic current comparator
    Elmquist, R.
    Jones, G. R., Jr.
    Pritchard, B.
    Bierzychudek, M.
    Hemandez, F.
    [J]. 2008 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2008, : 268 - 269
  • [6] Resistance scaling from 10 mΩ to kΩ at MIKES with cryogenic current comparator and guildline bridges
    Satrapinski, A
    Rautiainen, A
    Rajala, R
    [J]. 2002 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CONFERENCE DIGEST, 2002, : 56 - 57
  • [7] RESISTANCE SCALING-UP TO 1 M-OMEGA AT PTB WITH A CRYOGENIC CURRENT COMPARATOR
    PESEL, E
    SCHUMACHER, B
    WARNECKE, P
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1995, 44 (02) : 273 - 275
  • [8] AC resistance bridge based on the cryogenic current comparator
    Seppa, H
    Satrapinski, A
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (02) : 463 - 466
  • [9] A quantum Hall cryogenic current comparator resistance bridge
    Jones, K
    Early, MD
    [J]. 2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 92 - 93
  • [10] AN AUTOMATED CRYOGENIC CURRENT COMPARATOR RESISTANCE RATIO BRIDGE
    WILLIAMS, JM
    HARTLAND, A
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1991, 40 (02) : 267 - 270