Ultrasound imaging algorithm based on generalized sidelobe canceller

被引:1
|
作者
Wang Ping [1 ]
Cheng Na [1 ]
Gong Zhi-Hui [1 ]
Wang Lin-Hong [1 ]
机构
[1] Chongqing Univ, State Key Lab Power Transmiss Equip & Syst Secur, Chongqing 400044, Peoples R China
基金
中国国家自然科学基金;
关键词
minimum variance; coherence factor; generalized side lobe canceller; sign coherence factor;
D O I
10.7498/aps.64.238701
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The traditional delay and sum (DAS) algorithm is the most widely adopted method in medical ultrasound imaging; although it can produce images quickly, it sacrifices the resolution and the contrast ratio. The adaptive method such as the minimum variance (MV) continuously updates the apodization weighting vectors according to the received signals, so that the variance of the weighted signals is minimized, and thus the quality of the ultrasound imaging can be improved, especially its resolution. Although the image quality may be improved in the contrast ratio as well as the resolution after combining the minimum variance with the coherence factor (MV-CF), it complicates the algorithm, and the robustness against noise is enhanced but a little. An improved ultrasound imaging algorithm based on the generalized side lobe canceller (GSC) is proposed, which is constructed according to the minimum variance principle. The canceller is designed to classify the signal into desired and noise signals, combined with wiping off the big interferential eigenvectors, so that the robustness against noise can be enhanced. Firstly, the canceller divides the weighting vector into non-adaptive and adaptive weights, then the eigenstructure subspace is established according to the covariance matrix of the received signals, and the renewed weighting vector is achieved finally by projecting the weighting vector into the left singular space of the eigenstructure subspace. Simulations of the point targets and the cyst phantom through the simulation tool Field II demonstrate that the ultrasound image acquired through the proposed method is better than the traditional DAS and MV-CF algorithms in terms of the contrast ratio and resolution. In practice, the contrast ratio increases by roughly 7 dB compared to DAS and 5 dB to MV-CF. Furthermore, the proposed method gives a more satisfactory lateral resolution as well as the lowest side lobe peak level. From the sound-absorbing speckle simulation, the contrast ratio increases by 3 dB more than that of DAS and over 4 dB than that of MV-CF when noise exists. In addition, MV-CF performs the worst in the robustness aspect while the proposed GSC method makes improvement on the basis of it. Besides, the image quality can be further improved by combining the proposed method with sign coherence factor (GSC-SCF). After such a combination, the noise added to the data sets is almost invisible in point targets simulation. It also possesses the maximum mean power in cyst region in sound-absorbing speckle simulation. Finally, an experiment is conducted on the basis of the complete data sets which are offered by the University of Michigan. Results indicate that the proposed methods can perform better than the conventional DAS and MV-CF in resolution, contrast ratio and the robustness against noise.
引用
收藏
页数:7
相关论文
共 17 条
  • [1] Aboulnasr H, 2006, IEEE T SIGNAL PROCES, V54, P1587
  • [2] Minimum Variance Beamforming Combined with Adaptive Coherence Weighting Applied to Medical Ultrasound Imaging
    Asl, Babak Mohammadzadeh
    Mahloojifar, Ali
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2009, 56 (09) : 1923 - 1931
  • [3] Phase Coherence Imaging
    Camacho, Jorge
    Parrilla, Montserrat
    Fritsch, Carlos
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2009, 56 (05) : 958 - 974
  • [4] In vivo hyperthermia effect induced by high-intensity pulsed ultrasound
    Cui, Wei-Cheng
    Tu, Juan
    Hwang, Joo-Ha
    Li, Qian
    Fan, Ting-Bo
    Zhang, Dong
    Chen, Jing-Hai
    Chen, Wei-Zhong
    [J]. CHINESE PHYSICS B, 2012, 21 (07)
  • [5] CALCULATION OF PRESSURE FIELDS FROM ARBITRARILY SHAPED, APODIZED, AND EXCITED ULTRASOUND TRANSDUCERS
    JENSEN, JA
    SVENDSEN, NB
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1992, 39 (02) : 262 - 267
  • [6] Sequential beamforming for synthetic aperture imaging
    Kortbek, Jacob
    Jensen, Jorgen Arendt
    Gammelmark, Kim Lokke
    [J]. ULTRASONICS, 2013, 53 (01) : 1 - 16
  • [7] Doubly constrained robust Capon beamformer
    Li, J
    Stoica, P
    Wang, ZS
    [J]. IEEE TRANSACTIONS ON SIGNAL PROCESSING, 2004, 52 (09) : 2407 - 2423
  • [8] Adaptive imaging using the generalized coherence factor
    Li, PC
    Li, ML
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2003, 50 (02) : 128 - 141
  • [9] Adaptive beamforming for photoacoustic imaging
    Park, Suhyun
    Karpiouk, Andrei B.
    Aglyamov, Salavat R.
    Emelianov, Stanislav Y.
    [J]. OPTICS LETTERS, 2008, 33 (12) : 1291 - 1293
  • [10] A decimated minimum variance beamformer applied to ultrasound imaging
    Sakhaei, Sayed Mahmoud
    [J]. ULTRASONICS, 2015, 59 : 119 - 127