Laplacian electrode to record small bowel myoelectrical activity from abdominal surface

被引:0
|
作者
Prats-Boluda, G [1 ]
Martínez, JL [1 ]
Silvestre, J [1 ]
Sáiz, FJ [1 ]
Ponce, JL [1 ]
García-Casado, J [1 ]
机构
[1] Univ Politecn Valencia, Grp Bioelect, Dept Ingn Elect, E-46071 Valencia, Spain
关键词
electroenterogram; laplacian; electrode; bowel;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of this study is to develop a system to record small intestine myoelectric activity (electroenterogram) from abdominal surface. The difficult anatomic access is a technical problem for monitoring intestinal activity, therefore most methods are invasive. It results in many problems like the low amplitude of the signal recorded and physiological interferences captured. In order to avoid these troubles, and to estimate the location of the small intestine point whose activity is being recorded, it has been implemented a tripolar concentric ring electrode in bipolar configuration (TCB). This electrode is based on laplacian theory. It was analysed the signals recorded by TCB electrode and contrasted to the internal myoelectric signal and to the signals obtained from the body surface by using two unipolar electrodes. It was carried out a surgical intervention in a Beagle dog to suture a bipolar electrode at the bowel serosa. The signals recorded were acquired and preprocessed to carry out a density power spectral analysis (periodogram). The electroenterogram captured by TCB electrode were free of ECG interferences. Its periodogram revealed a peak near 0.25 Hz, corresponding to the slow wave frequency (13 to 18 cpm), while it rejected a peak at 0.4 Hz due to breath interference. These results show the utility of TCB electrode in order to remove interferences in abdominal surface electroenterogram acquisitions.
引用
收藏
页码:3070 / 3073
页数:4
相关论文
共 50 条
  • [1] Identification of the slow wave of small bowel myoelectrical activity by surface recording
    Martínez-de-Juan, JL
    Saiz, FJ
    Silvestre, J
    Ponce, JL
    PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 2024 - 2027
  • [2] EFFECT OF SOMATOSTATIN ON MYOELECTRICAL ACTIVITY OF SMALL BOWEL
    THOR, P
    KROL, R
    KONTUREK, SJ
    COY, DH
    SCHALLY, AV
    AMERICAN JOURNAL OF PHYSIOLOGY, 1978, 235 (03): : E249 - E254
  • [3] INFLUENCE OF SOMATOSTATIN ON MYOELECTRICAL ACTIVITY OF SMALL BOWEL
    THOR, PJ
    KROL, R
    KONTUREK, SJ
    GASTROENTEROLOGY, 1977, 72 (05) : 1169 - 1169
  • [4] Model predictions of myoelectrical activity of the small bowel
    Miftakhov, RN
    Abdusheva, GR
    Wingate, DL
    BIOLOGICAL CYBERNETICS, 1996, 74 (02) : 167 - 179
  • [6] CHANGES IN SMALL BOWEL MYOELECTRICAL ACTIVITY FOLLOWING LAPAROTOMY
    MORRIS, IR
    DARBY, CF
    HAMMOND, P
    TAYLOR, I
    BRITISH JOURNAL OF SURGERY, 1983, 70 (09) : 547 - 548
  • [7] EFFECT OF REVERSED JEJUNAL SEGMENTS ON MYOELECTRICAL ACTIVITY OF SMALL BOWEL
    OLEARY, JF
    HENNESSY, TPJ
    BRITISH JOURNAL OF SURGERY, 1976, 63 (02) : 159 - 159
  • [8] EFFECT OF REVERSED JEJUNAL SEGMENTS ON MYOELECTRICAL ACTIVITY OF SMALL BOWEL
    TANNER, WA
    OLEARY, JF
    BYRNE, PJ
    HENNESSY, TPJ
    BRITISH JOURNAL OF SURGERY, 1978, 65 (08) : 567 - 571
  • [9] EFFECT OF REVERSED JEJUNAL SEGMENTS ON MYOELECTRICAL ACTIVITY OF SMALL BOWEL
    BYRNE, PJ
    HENNESSY, TPJ
    MARKS, P
    OLEARY, JF
    TANNER, WA
    JOURNAL OF PHYSIOLOGY-LONDON, 1976, 263 (02): : P276 - P277
  • [10] A model of myoelectrical activity recording after small bowel transplantation
    Minioli, AM
    Lopasso, FP
    Forno, ERS
    Larangeira, LLS
    Goldenberg, S
    Gomes, PP
    Neto, AB
    TRANSPLANTATION PROCEEDINGS, 1998, 30 (06) : 2611 - 2612