Comparative analysis of radiometer systems using non-stationary processes

被引:0
|
作者
Racette, P [1 ]
Lang, R [1 ]
机构
[1] NASA, Goddard Space Flight Ctr, Microwave Instrument Technol Branch, Greenbelt, MD 20771 USA
关键词
D O I
暂无
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Radiometers require periodic calibration to correct for instabilities in the receiver response. Various calibration techniques exist that minimize the effect of instabilities in the receivers. The optimal technique depends upon many parameters. Some parameters are constrained by the particular application and others can be chosen in the system design. For example, the measurement uncertainty may be reduced to the limits of the resolution of the measurement (sensitivity) if periodic absolute calibration can be performed with sufficient frequency. However if the period between calibrations is long, a reference-differencing technique, i.e. Dicke-type design, can yield better performance. The measurement uncertainty not only depends upon the detection scheme but also on the number of pixels between calibrations, the integration time per pixel, integration time per calibration reference measurement, calibration reference temperature, and the brightness temperature of what is being measured. The best scheme for reducing the measurement uncertainty also depends, in large part, on the stability of the receiver electronics. In this presentation a framework for evaluating calibration schemes for a wide range of system architectures is presented. Two methods for treating receiver non-stationarity are compared with radiometer measurements.
引用
收藏
页码:565 / 567
页数:3
相关论文