Determination of the impurities elements in high-purity cadmium by inductively coupled plasma mass spectrometry

被引:0
|
作者
Cao, SQ [1 ]
Chen, HT [1 ]
Zeng, XJ [1 ]
机构
[1] Chinese Acad Sci, Changchun Inst Appl Chem, Natl Electrochem & Spect Res & Anal Ctr, Changchun 130022, Peoples R China
关键词
inductively coupled plasma mass spectrometry; high purity cadmium; impurities;
D O I
暂无
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A method for the determination of impurity elements in high purity cadmium by inductively coupled plasma mass spectrometry was developed. The spectral interference arising from Cd was discussed and the magnitude of interference by the polyatomic ion of Cd was compared. The effect of the high concentration Cd on the analyte signal was studied. The results showed that Cd possessed both significant suppression effects on the light mass elements signal and enhancement effects on the heavy mass elements signal,and the effects can be corrected by the use of internal standard elements. In this paper, Y and Bi was employed for the light mass elements and the heavy mass elements,respectively. As, Be, Co, Cu, Ga, Ge, Mn, Mo, Pb, Ni, Sr, Au, Tl, Th, V and U in high purity cadmium were determined. Detection limits are 0.005-0.052 mu g L-1,and the recoveries of standard addition are 82%-108%.
引用
收藏
页码:854 / 857
页数:4
相关论文
共 5 条