A comparison of coulometric titration and potentiometric determination of chloride concentration in rumen fluid

被引:4
|
作者
Cebra, CK
Tornquist, SJ
Vap, LM
Dodson, LA
机构
[1] Oregon State Univ, Coll Vet Med, Dept Large Anim Clin Sci, Corvallis, OR 97331 USA
[2] Oregon State Univ, Coll Vet Med, Dept Biomed Sci, Corvallis, OR 97331 USA
[3] Colorado State Univ, Vet Teaching Hosp, Coll Vet Med, Ft Collins, CO 80523 USA
关键词
acetate; camelid; chloride; rumen fluid; ruminant;
D O I
10.1111/j.1939-165X.2001.tb00433.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The concentration of chloride ions in rumen fluid is a useful measure of obstructive gastrointestinal disease in ruminants and camelids. However, rumen fluid is very different from other biological fluids in its bacterial populations, consistency, and concentrations of various anions. Two methods of determining the chloride concentration in biological fluids were compared using centrifuged and filtered rumen fluid containing different amounts of sodium chloride. Although coulometric titration and potentiometric electrode analysis yielded results that had a strong linear relationship, the results of potentiometry were consistently and significantly higher, by about 20 mEq/L. This difference was investigated further by analyzing a series of fluids containing different concentrations of sodium acetate. Acetate was detected as chloride (0.21 chloride molecules per acetate molecule) by potentiometry but not by coulometric titration. Therefore, the acetate concentration of rumen fluid was the most likely cause of the discrepancy between tests in the original trial. In conclusion, the coulometric procedure may be more accurate than the potentiometric procedure for measuring rumen chloride when the concentrations of possible confounding ions are unknown. (C) 2001 American Society for Veterinary Clinical Pathology.
引用
收藏
页码:211 / 213
页数:3
相关论文
共 50 条