Nonlinear S-Parameters Inversion for Stroke Imaging

被引:64
|
作者
Fedeli, Alessandro [1 ]
Schenone, Valentina [1 ]
Randazzo, Andrea [1 ]
Pastorino, Matteo [1 ]
Henriksson, Tommy [2 ]
Semenov, Serguei [3 ]
机构
[1] Univ Genoa, Dept Elect Elect Telecommunicat Engn & Naval Arch, I-16145 Genoa, Italy
[2] EMTensor GmbH, Electromagnet Tomog Device Dev Team, A-1220 Vienna, Austria
[3] EMTensor GmbH, A-1220 Vienna, Austria
关键词
Scattering parameters; Imaging; Antenna measurements; Microwave theory and techniques; Microwave imaging; Dielectrics; Mathematical model; Finite elements; inverse problems; Lebesgue spaces; S-parameters; stroke imaging; BREAST-CANCER DETECTION; ACUTE ISCHEMIC-STROKE; NEURAL-NETWORK; MICROWAVE; SCATTERING; TIME; LOCALIZATION; ALGORITHM; DIAGNOSIS; RADAR;
D O I
10.1109/TMTT.2020.3040483
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Stroke identification by means of microwave tomography requires a very accurate reconstruction of the dielectric properties inside patient's head. This is possible when a precise measurement system is combined with a full nonlinear inversion method. In this article, the inversion of S-parameter data collected in a metallic chamber is performed with a nonlinear inversion strategy in Lebesgue spaces with nonconstant exponents. This is the first time that this kind of nonlinear S-parameter electromagnetic formulation has been applied to this problem. The inverse-scattering method incorporates a 2-D electromagnetic model of the imaging chamber based on a finite-element formulation, which has led to a complete redefinition of the solving procedure with respect to previous works. This allows a suitable description of the multistatic S-parameters due to the interactions between the incident radiation and the structure under test. The developed inversion procedure is first assessed by means of numerical simulations. The experimental results, obtained with a clinical microwave system prototype containing a liquid-filled 3-D SAM phantom with an inhomogeneity mimicking a hemorrhagic stroke, further prove the effectiveness of the proposed approach.
引用
收藏
页码:1760 / 1771
页数:12
相关论文
共 50 条
  • [1] S-Parameters
    Smith, JA
    MICROWAVES & RF, 2004, 43 (09) : 13 - 13
  • [2] Nonlinear S-Parameters And SDR Impact Test And Measurement Equipment
    Friedrich, Nancy
    MICROWAVES & RF, 2010, 49 (01) : 33 - +
  • [3] A Comparison of the Volterra Series-based Nonlinear S-parameters and X-parameters
    Widemann, Christian
    Weber, Harry
    Schatz, Soeren
    Mathis, Wolfgang
    2015 22ND INTERNATIONAL CONFERENCE MIXED DESIGN OF INTEGRATED CIRCUITS & SYSTEMS (MIXDES), 2015, : 453 - 457
  • [4] S-PARAMETERS - A PRACTICAL EDUCATION
    DAVIS, WA
    OVERSTREET, WP
    IEEE TRANSACTIONS ON EDUCATION, 1989, 32 (01) : 18 - 25
  • [5] Software finesses S-parameters
    Smolyansky, D
    Mayevskiy, E
    CONNECTOR SPECIFIER, 2005, 21 (08) : 11 - 13
  • [6] Microwave Inversion of Measured S-Parameters using a Thin-Wire Antenna Model
    Banting, Lucas
    Jeffrey, Ian
    LoVetri, Joe
    2024 18TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION, EUCAP, 2024,
  • [7] Large Signal S-parameters
    Odyniec, Michal
    67TH ARFTG MICROWAVE MEASUREMENTS CONFERENCE, 2007, : 55 - 60
  • [8] Analytical Calculation of Transformer Parameters by S-Parameters
    Lou, Jianquan
    Bhobe, Alpesh
    Shu, Yingchun
    Yu, Jinghan
    2018 JOINT IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY AND 2018 IEEE ASIA-PACIFIC SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY (EMC/APEMC), 2018, : 1310 - 1313
  • [9] S-parameters made simple
    Green, L
    ELECTRONICS WORLD, 2000, 106 (1776): : 928 - 933
  • [10] Measurement technique for the extraction of differential s-parameters from single-ended s-parameters
    Vaz, K
    Caggiano, M
    27th International Spring Seminar on Electronics Technology, Books 1-3, Conference Proceedings: MEETING THE CHALLENGES OF ELECTRONICS TECHNOLOGY PROGRESS, 2004, : 313 - 317