Effects of luminance and resolution on observer performance with chest radiographs

被引:53
|
作者
Herron, JM
Bender, TM
Campbell, WL
Sumkin, JH
Rockette, HE
Gur, D
机构
[1] Univ Pittsburgh, Dept Radiol, Pittsburgh, PA 15261 USA
[2] Univ Pittsburgh, Dept Biostat, Pittsburgh, PA 15261 USA
关键词
diagnostic radiology; observer performance; images; display; lung; diseases; neoplasms; pneumothorax; ribs; fracture;
D O I
10.1148/radiology.215.1.r00ap34169
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To examine the combined effects of image resolution and display luminance on observer performance for detection of abnormalities depicted on posteroanterior chest radiographs. MATERIALS AND METHODS: A total of 529 radiographs were displayed on a specially constructed view box at three luminance levels (770, 260, and 85 cd/m(2)) and three resolutions (100-mu m, 200-mu m, and 400-mu m pixels). Each image was reviewed nine times by six radiologists who participated in this study. The abnormalities included nodule, pneumothorax, interstitial disease, alveolar infiltrates, and rib fracture. Negative (normal) radiographs were also included. REULTS: Receiver operating characteristic curves indicated that the effect of image luminance was greater than that of resolution. The detection of pneumothorax, interstitial disease, and rib fracture showed statistically significant differences (P < .05) due to luminance. The detection of pneumothorax was the only abnormality with a statistically significant difference due to resolution. There was no evidence that luminance was related to image resolution for any of the abnormalities. CONCLUSION: At a resolution of 400-mu m pixels or higher across the field of view and a luminance of 260 cd/m(2) or more, primary diagnosis with posteroanterior chest radiographs is not likely to be affected by the quality of display.
引用
收藏
页码:169 / 174
页数:6
相关论文
共 50 条
  • [1] Influence of monitor luminance change on observer performance for detection of abnormalities depicted on chest radiographs
    Ikeda, M
    Ishigaki, T
    Shimamoto, K
    Kodera, Y
    Koyama, S
    Usami, H
    Kato, K
    INVESTIGATIVE RADIOLOGY, 2003, 38 (01) : 57 - 63
  • [2] Effect of monitor luminance and ambient light on observer performance in soft-copy reading of digital chest radiographs
    Goo, JM
    Choi, AY
    Im, JG
    Lee, HL
    Chung, MJ
    Han, DH
    Park, SH
    Kim, JH
    Nam, SH
    RADIOLOGY, 2004, 232 (03) : 762 - 766
  • [3] SPATIAL-RESOLUTION REQUIREMENTS FOR DIGITAL CHEST RADIOGRAPHS - AN ROC STUDY OF OBSERVER PERFORMANCE IN SELECTED CASES
    LAMS, PM
    COCKLIN, ML
    RADIOLOGY, 1986, 158 (01) : 11 - 19
  • [4] CHEST RADIOGRAPHS OBTAINED WITH SHAPED FILTERS - EVALUATION BY OBSERVER PERFORMANCE TESTS
    KELSEY, CA
    LANE, RG
    MOSELEY, RD
    METTLER, FA
    ROSENBERG, RD
    WILLIAMS, AG
    GARCIA, JF
    FELDMAN, BS
    BOARDMAN, RE
    RADIOLOGY, 1986, 159 (03) : 653 - 655
  • [5] Effects of monitor luminance change on observer detection performance
    Hidano, S
    Ikeda, M
    Ishigaki, T
    Usami, H
    Shimamoto, K
    Kato, K
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2005, 29 (01) : 35 - 41
  • [6] EFFECTS OF IMAGE-PROCESSING ON NODULE DETECTION RATES IN DIGITIZED CHEST RADIOGRAPHS - ROC STUDY OF OBSERVER PERFORMANCE
    SHERRIER, RH
    CHILES, C
    WILKINSON, WE
    JOHNSON, GA
    RAVIN, CE
    RADIOLOGY, 1988, 166 (02) : 447 - 450
  • [7] Observer performance evaluation of the feasibility of a deep learning model to detect cardiomegaly on chest radiographs
    Ajmera, Pranav
    Kharat, Amit
    Gupte, Tanveer
    Pant, Richa
    Kulkarni, Viraj
    Duddalwar, Vinay
    Lamghare, Purnachandra
    ACTA RADIOLOGICA OPEN, 2022, 11 (07)
  • [8] Computer-aided Detection of Lung Cancer on Chest Radiographs: Effect on Observer Performance
    de Hoop, Bartjan
    De Boo, Diederik W.
    Gietema, Hester A.
    van Hoorn, Frans
    Mearadji, Banafsche
    Schijf, Laura
    van Ginneken, Bram
    Prokop, Mathias
    Schaefer-Prokop, Cornelia
    RADIOLOGY, 2010, 257 (02) : 532 - 540
  • [9] COMPARISON OF SCREEN FILM AND DIGITAL CHEST RADIOGRAPHS FOR DETECTION OF PNEUMOTHORAX - ROC STUDY OF OBSERVER PERFORMANCE
    FAJARDO, LL
    HILLMAN, BJ
    INVESTIGATIVE RADIOLOGY, 1988, 23 (09) : S19 - S19
  • [10] OBSERVER EVALUATION OF SCATTER SUBTRACTION FOR DIGITAL PORTABLE CHEST RADIOGRAPHS