A novel physician-assembled endograft for the repair of pararenal, paravisceral, Crawford type IV thoracoabdominal aortic aneurysms, and aneurysms requiring treatment after prior repair

被引:7
|
作者
Jorgensen, Benjamin D. [1 ]
Malek, Mathew [1 ]
VandenHull, Angela [2 ]
Remund, Tyler [2 ]
Truong, Khang-Christopher [2 ]
Pohlson, Katie [2 ]
Kelly, Patrick W. [3 ,4 ]
机构
[1] Univ South Dakota Sanford, Sch Med, Sioux Falls, SD USA
[2] Univ South Dakota Sanford, Sanford Res, Sioux Falls, SD USA
[3] Sanford Res, Dept Vasc Surg, Sanford Vasc Associates, Sioux Falls, SD 57117 USA
[4] Sanford Hlth, Sanford Vasc Associates, Dept Vasc Surg, Sioux Falls, SD 57117 USA
关键词
Branched; Complex aneurysm repair; Endovascular aortic repair; Thoracoabdominal aneurysm; STENT-GRAFT; UNITED-STATES;
D O I
10.1016/j.jvs.2020.03.045
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: In the present study, we reviewed the 30-day and 1-year clinical results of the use of the investigational unitary manifold (UM) stent graft system (Sanford Health, Sioux Falls SDak) for the repair of Crawford type IV, pararenal, paravisceral, juxtarenal, and short-neck infrarenal aneurysms (<10 mm). Methods: The present study was a single-center, mu ltiarm, prospective review of the first 44 patients who had undergone repair of Crawford type IV, pararenal, juxtarenal, and short-neck infrarenal aneurysms (<10 mm) using the physician modified UM under a physician-sponsored investigational device exemption. The primary end point was freedom from major adverse events at 30 days, including all-cause mortality, myocardial infarction, stroke, paraplegia, bowel ischemia, respiratory failure, and renal failure. Results: Technical success was achieved in all 44 patients (100%), with a large number of these patients having undergone previous aortic repair (20 of 44; 45.5%). All the intended 170 visceral vessels (100%) had been successfully cannulated and stent grafted. No episodes of paraplegia or in-hospital deaths were recorded. One patient had died of aneurysm-related ischemic stroke (2.3%). The rate of transient nonclinically significant spinal cord ischemia was 4.5%. At the last follow-up, one reintervention had been required owing to branch patency from a thrombotic event. Of the 170 bridging stent grafts, 169 have remained patent through a mean follow-up of 8.8 months (range, 0-36 months). No type I or III endoleaks, migration, or component separation in the investigational device has occurred. Conclusions: The early and midterm results with the use of the UM suggest it could be a viable option for the repair of Crawford type IV, pararenal, paravisceral, juxtarenal, and short-neck infrarenal aneurysms (<10 mm) without exposing patients to the increased risk of permanent spinal cord ischemia, renal failure, visceral vessel ischemia, or aneurysm related mortality that results from open thoracoabdominal aortic aneurysm repair. The high technical success rate, in native and previous repairs, supports the utility of this device as a bail-out technique for failed endovascular aneurysm repair or proximal extension of disease after previous aortic repair. However, experience is limited, and this approach requires further study before widespread adoption.
引用
收藏
页码:1897 / +
页数:11
相关论文
共 48 条
  • [1] Survival after repair of pararenal and paravisceral abdominal aortic aneurysms
    Kabbani, Loay S.
    West, Charles A.
    Viau, David
    Nypaver, Timothy J.
    Weaver, Mitchell R.
    Barth, Charlotte
    Lin, Judith C.
    Shepard, Alexander D.
    JOURNAL OF VASCULAR SURGERY, 2014, 59 (06) : 1488 - 1494
  • [2] A novel endovascular debranching technique using physician-assembled endografts for repair of thoracoabdominal aneurysms
    Anderson, Joe
    Nykamp, Madeline
    Danielson, Laura
    Remund, Tyler
    Kelly, Patrick W.
    JOURNAL OF VASCULAR SURGERY, 2014, 60 (05) : 1177 - 1184
  • [4] Assessment of postoperative paraplegia after endograft repair of thoracic aortic aneurysms and thoracoabdominal aortic aneurysms
    Kuratani, T
    Kaneko, M
    Ueda, H
    Nishimura, T
    Shimamura, K
    CIRCULATION, 2003, 108 (17) : 525 - 525
  • [5] The Impact of Prior Aortic Surgery on Outcomes After Multibranched Endovascular Repair of Thoracoabdominal and Pararenal Aortic Aneurysms
    Kaushik, Smita
    Ramanan, Bala
    Gasper, Warren
    Vartanian, Mdshant
    Reilly, Linda
    Chuter, Timothy
    Hiramoto, Jade
    JOURNAL OF VASCULAR SURGERY, 2017, 65 (06) : 29S - 30S
  • [6] Improvement in results of repair of type IV thoracoabdominal aortic aneurysms
    Schwartz, LB
    Belkin, M
    Donaldson, MC
    Mannick, JA
    Whittemore, AD
    JOURNAL OF VASCULAR SURGERY, 1996, 24 (01) : 74 - 81
  • [7] Independent associations with early mortality after open repair of Crawford extent IV thoracoabdominal aortic aneurysms
    Koksoy, Cuneyt
    Rebello, Kimberly R.
    Green, Susan Y.
    Amarasekara, Hiruni S.
    Moon, Marc R.
    LeMaire, Scott A.
    Coselli, Joseph S.
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2024, 168 (01):
  • [8] Endovascular stent-graft repair of pararenal and type IV thoracoabdominal aortic aneurysms with adjunctive visceral reconstruction
    Fulton, JJ
    Farber, MA
    Marston, WA
    Mendes, R
    Mauro, MA
    Keagy, BA
    JOURNAL OF VASCULAR SURGERY, 2005, 41 (02) : 191 - 198
  • [9] TRANSABDOMINAL REPAIR OF TYPE-IV THORACOABDOMINAL AORTIC-ANEURYSMS
    GILLINGSMITH, GL
    WOLFE, JHN
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1995, 9 (01) : 112 - 113
  • [10] Contemporary results for open repair of suprarenal and type IV thoracoabdominal aortic aneurysms
    Richards, J. M. J.
    Nimmo, A. F.
    Moores, C. R.
    Hansen, P. A.
    Murie, J. A.
    Chalmers, R. T. A.
    BRITISH JOURNAL OF SURGERY, 2010, 97 (01) : 45 - 49