First World Report of Internal Power Cable Repair in Left Ventricular Assist Device Jarvik 2000: Case Report

被引:0
|
作者
Sassi, C. G. [1 ]
Cameli, M. [2 ]
Dokollari, A. [1 ]
Diciolla, F. [1 ]
Scolletta, S. [3 ]
Ricci, C. [4 ]
Lucatelli, P. [4 ]
Mondillo, S. [2 ]
Maccherini, M. [1 ]
机构
[1] Univ Siena, Dept Cardiac Surg, Viale Bracci 16, Siena, Italy
[2] Univ Siena, Dept Cardiovasc Dis, Siena, Italy
[3] Univ Hosp Siena, Dept Med Biotechnol Anesthesiol & Intens Care, Siena, Italy
[4] Azienda Osped Univ Senese, Policlin Santa Maria Scotte, Intervent Radiol, Siena, Italy
关键词
D O I
10.1016/j.transproceed.2017.02.029
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Background. There are limited clinical reports concerning internal power cable fixing in left ventricular assist device (L-VAD) patients. Actually there are no reports in the literature about Jarvik 2000 internal cable repair. We show the first description of a technique for surgical reparation of such a fatal complication. Patient History. The patient was a 62-year-old woman who had L-VAD implantation (Jarvik 2000) with outflow graft apposition in descending thoracic aorta through left thoracotomy access, in 2009. She arrived urgently on January 25, 2014 for Jarvik 2000 dysfunction correlated with head movements. The neck X-rays revealed the rupture of one of the nine power cables located inside the neck and the damaging of two more cables nearby to be ruptured. On the same day she got pump failure due to the final interruption of the remaining two cables, we were obliged to install femoro-femoral extracorporeal membrane oxygenation (ECMO) assistance, to repair the power cables, approaching them through a pacemaker extension cable. The L-VAD outflow was occluded with vascular ball occluder inserted via right axillary artery under fluoroscopy before ECMO installation. At the end the ECMO assistance was interrupted and the Jarvik 2000 was turned back on. The patient was dismissed from the hospital 12 days after the procedure. Discussion. At the moment the international literature is poor regarding this issue. This case provides evidence that in emergency conditions ECMO assistance is mandatory and a hybrid surgical and radiological approach could help to repair the damage in safe conditions.
引用
收藏
页码:743 / 746
页数:4
相关论文
共 50 条
  • [1] Implantation of the Jarvik 2000® left ventricular assist device using the miniaturized extracorporeal circulation system - a case report
    Formica, Francesco
    Bamodu, Oluwaseun Adebayo
    Mariani, Serena
    Caruso, Rosa
    Sangalli, Fabio
    Paolini, Giovanni
    [J]. PERFUSION-UK, 2016, 31 (06): : 518 - 520
  • [2] Stapled explant of the Jarvik 2000 left ventricular assist device
    Banbury, MK
    [J]. ANNALS OF THORACIC SURGERY, 2003, 75 (01): : 297 - 297
  • [3] The first clinical case in Japan of destination therapy using the Jarvik 2000 left ventricular assist device
    Sokichi Kamata
    Taichi Sakaguchi
    Shigeru Miyagawa
    Yasushi Yoshikawa
    Takashi Yamauchi
    Koji Takeda
    Shunsuke Saito
    Takayoshi Ueno
    Toru Kuratani
    Yoshiki Sawa
    [J]. Journal of Artificial Organs, 2010, 13 : 170 - 173
  • [4] Evolutionary Improvements in the Jarvik 2000 Left Ventricular Assist Device
    Selzman, Craig H.
    Koliopoulou, Antigone
    Glotzbach, Jason P.
    McKellar, Stephen H.
    [J]. ASAIO JOURNAL, 2018, 64 (06) : 827 - 830
  • [5] The first clinical case in Japan of destination therapy using the Jarvik 2000 left ventricular assist device
    Kamata, Sokichi
    Sakaguchi, Taichi
    Miyagawa, Shigeru
    Yoshikawa, Yasushi
    Yamauchi, Takashi
    Takeda, Koji
    Saito, Shunsuke
    Ueno, Takayoshi
    Kuratani, Toru
    Sawa, Yoshiki
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2010, 13 (03) : 170 - 173
  • [6] Suicide by cutting the electrical cable of left ventricular assist device (A case report)
    Zagelidou, E.
    Chatzinikolaou, F.
    Enache, A.
    Paliouras, D.
    Leodari, R.
    [J]. IALM 2012: PRESENTATIONS FROM THE 22ND CONGRESS OF THE INTERNATIONAL ACADEMY OF LEGAL MEDICINE, 2013, : 115 - 118
  • [7] Midterm experience with the Jarvik 2000 axial flow left ventricular assist device
    Haj-Yahia, Saleem
    Birks, Emma J.
    Rogers, Paula
    Bowles, Christopher
    Hipkins, Mandy
    George, Robert
    Amrani, Mohammed
    Petrou, Mario
    Pepper, John
    Dreyfus, Gilles
    Khaghani, Asghar
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2007, 134 (01): : 199 - 203
  • [8] Pump Exchange from a Left Ventricular Assist Device to a Jarvik 2000 with a Postauricular Cable Due to Abdominal Driveline Infection
    Kishimoto, Yuichiro
    Yoshikawa, Yasushi
    Onohara, Takeshi
    Sumi, Naoki
    Kumagai, Kunitaka
    Sakaguchi, Yuki
    Kishimoto, Nozomi
    Ikeda, Yosuke
    Yoshikawa, Yuki
    Yamane, Kazuma
    Nishimura, Motonobu
    [J]. YONAGO ACTA MEDICA, 2023, 66 (03) : 389 - 393
  • [9] Central aortic valve closure successfully treated aortic insufficiency of the patient with Jarvik 2000 continuous flow left ventricular assist device: a case report
    Itoda, Yoshifumi
    Nawata, Kan
    Yamauchi, Haruo
    Kinoshita, Osamu
    Kimura, Mitsutoshi
    Ono, Minoru
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2017, 20 (01) : 99 - 101
  • [10] Intermediate experience with the Jarvik 2000 axial flow left ventricular assist device (LVAD)
    Haj-Yahya, S. M.
    Birks, E. J.
    Hardy, J.
    Hipkin, M.
    Bowles, C.
    Dreyfus, G.
    Banner, N.
    Khaghani, A.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2005, 24 (02): : S58 - S58