Reasons for failure of a mammography unit at clinical image review in the American College of Radiology Mammography Accreditation Program

被引:47
|
作者
Bassett, LW
Farria, DM
Bansal, S
Farquhar, MA
Wilcox, PA
Feig, SA
机构
[1] Thomas Jefferson Univ, Breast Imaging Ctr, Philadelphia, PA 19107 USA
[2] Univ Calif Los Angeles, Sch Med, Iris Cantor Ctr Breast Imaging, Los Angeles, CA USA
[3] Amer Coll Radiol, Reston, VA USA
[4] CUNY Mt Sinai Sch Med, Dept Radiol, New York, NY 10029 USA
关键词
breast radiography; quality assurance;
D O I
10.1148/radiology.215.3.r00jn32698
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To identify the most common deficiencies in the quality of mammograms submitted for clinical image evaluation (evaluation of image from actual patient referred for mammography). MATERIALS AND METHODS: In 1997, the American College of Radiology Mammography Accreditation Program reviewed clinical images for 2,341 mammography units. For each mammography unit, the facility submitted bilateral mediolateral oblique and craniocaudal mammograms obtained in a woman with fatty breasts and a woman with dense breasts. Images were reviewed independently by two experienced radiologists. Reviewers listed the general categories and specific deficiencies that led to a decision to fail the unit that produced the clinical images. RESULTS: Of the 2,341 mammography units, 1,034 (44%) failed the clinical image evaluation process. Of 6,128 categories cited by reviewers as deficient, 1,250 (20%) involved problems in positioning; 944 (15%), exposure; 887 (14%), compression; 806 (13%), sharpness; 785 (13%), contrast; 703 (11%), labeling; 465 (8%), artifacts; and 288 (5%), noise. A significantly higher proportion of failures was attributed to positioning deficiencies for fatty breasts than for dense breasts (P = .028). Higher proportions of failures in dense breasts were related to compression (P <.001) and exposure (P <. 001) deficiencies. CONCLUSION: Common problems in clinical image quality have been identified. This information should be useful for educators and facilities striving to improve the quality of mammography.
引用
收藏
页码:698 / 702
页数:5
相关论文
共 33 条