Do Titanium Mini-Implants Have the Same Quality of Finishing and Degree of Contamination before and after Different Manipulations? An In Vitro Study

被引:4
|
作者
Zogheib, Tatiana [1 ]
Walter-Solana, Andre [1 ]
de la Iglesia, Fernando [1 ]
Espinar, Eduardo [2 ]
Gil, Javier [1 ,3 ]
Puigdollers, Andreu [1 ]
机构
[1] Univ Int Catalunya, Dept Orthodont, Fac Dent, Barcelona 08195, Spain
[2] Univ Seville, Dept Orthodont, Fac Dent, Seville 41009, Spain
[3] Univ Int Catalunya, Bioengn Inst Technol, Barcelona 08195, Spain
关键词
contamination; manipulation; orthodontic; mini-implant; SEM; surface treatment;
D O I
10.3390/met11020245
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Evaluate the quality of finishing and degree of contamination before and after handling and surface treatment of titanium (Ti) orthodontic mini-implants (OMIs). A scanning electron microscope (SEM) study on ninety-six titanium OMIs was done. Energy-Dispersive X-ray Analysis (EDX) identified the present particles on manufactured OMIs surfaces. Then, OMIs were manipulated with gauze (dry sterile, soaked in chlorhexidine) and gloves (latex, nitrile) to evaluate the contamination of these handling materials. Finally, OMIs underwent surface treatments and were placed in bone to observe the contaminants they released. Roughness (Ra) and wettability with contact angle parameter (CA) were measured on these treated OMIs. Machined OMIs presented surface irregularities and were contaminated with manufacturing-process particles (carbon, plastic Polyvinyl Chloride PVC, aluminum). Hand-manipulated OMIs were also contaminated by the handling materials. OMIs surface characteristics were as follows: acid-etched (Ra approximate to 1.3 mu m, CA approximate to 66 degrees), machined (Ra approximate to 0.3 mu m, CA approximate to 68 degrees), SB (Ra approximate to 3.3 mu m, CA approximate to 78 degrees), and SBAO (Ra approximate to 3.1 mu m, CA approximate to 92 degrees). Bone was contaminated by OMIs surface defects and extra particles. Manufactured OMIs have surface contaminants that increase with clinical handling. Surface treatments (SBAO, a combination of sandblasting and anodic oxidation) increase the roughness and contact angle, which play an important role in osseointegration. Surface-treated OMIs leave titanium particles in the bone during their insertion-removal. The use of a gauze soaked in chlorhexidine is recommended when handling OMIs. Further investigations would be interesting to study more variables and confirm the present results.
引用
收藏
页码:1 / 11
页数:11
相关论文
共 7 条
  • [1] Do Chlorhexidine and Probiotics Solutions Provoke Corrosion of Orthodontic Mini-implants? An In Vitro Study
    Pavlic, Andrej
    Perissinotto, Fabio
    Turco, Gianluca
    Contardo, Luca
    Stjepan, Spalj
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2019, 34 (06) : 1379 - 1388
  • [2] Mechanical strength of stainless steel and titanium alloy mini-implants with different diameters: an experimental laboratory study
    Barros, Sergio Estelita
    Vanz, Viviane
    Chiqueto, Kelly
    Janson, Guilherme
    Ferreira, Eduardo
    PROGRESS IN ORTHODONTICS, 2021, 22 (01)
  • [3] Mechanical strength of stainless steel and titanium alloy mini-implants with different diameters: an experimental laboratory study
    Sérgio Estelita Barros
    Viviane Vanz
    Kelly Chiqueto
    Guilherme Janson
    Eduardo Ferreira
    Progress in Orthodontics, 22
  • [4] A comparative evaluation of ion release from different commercially-available orthodontic mini-implants - an in-vitro study
    Ananthanarayanan, Venkateswaran
    Padmanabhan, Sridevi
    Chitharanjan, Arun B.
    AUSTRALIAN ORTHODONTIC JOURNAL, 2016, 32 (02) : 165 - 174
  • [5] Bacterial contamination of ultrasound probes in different radiological institutions before and after specific hygiene training: do we have a general hygienical problem?
    Thomas Sartoretti
    Elisabeth Sartoretti
    Candid Bucher
    Aleksis Doert
    Christoph Binkert
    Klaus Hergan
    Matthias Meissnitzer
    Johannes Froehlich
    Orpheus Kolokythas
    Simon Matoori
    Christina Orasch
    Sebastian Kos
    Sabine Sartoretti-Schefer
    Andreas Gutzeit
    European Radiology, 2017, 27 : 4181 - 4187
  • [6] Bacterial contamination of ultrasound probes in different radiological institutions before and after specific hygiene training: do we have a general hygienical problem?
    Sartoretti, Thomas
    Sartoretti, Elisabeth
    Bucher, Candid
    Doert, Aleksis
    Binkert, Christoph
    Hergan, Klaus
    Meissnitzer, Matthias
    Froehlich, Johannes
    Kolokythas, Orpheus
    Matoori, Simon
    Orasch, Christina
    Kos, Sebastian
    Sartoretti-Schefer, Sabine
    Gutzeit, Andreas
    EUROPEAN RADIOLOGY, 2017, 27 (10) : 4181 - 4187
  • [7] Adhesion of Resin to Lithium Disilicate with Different Surface Treatments before and after Salivary Contamination-An In-Vitro Study
    Harouny, Ryan
    Hardan, Louis
    Harouny, Elie
    Kassis, Cynthia
    Bourgi, Rim
    Lukomska-Szymanska, Monika
    Kharouf, Naji
    Ball, Vincent
    Khairallah, Carlos
    BIOENGINEERING-BASEL, 2022, 9 (07):