Design and development of wireless power transfer system for implantable cardioverter defibrillator

被引:7
|
作者
Aditya, Kuragayala [1 ]
Bobba, Phaneendra Babu [1 ]
Satyavani, Kuragayala [1 ]
机构
[1] Gokaraju Rangaraju Inst Engn & Technol, Dept Elect & Elect Engn, Hyderabad, Telangana, India
关键词
Implantable cardioverter defibrillator (ICD); LCC-C compensation; magnetic coupling resonance (MCR); wireless power transfer (WPT);
D O I
10.3233/JAE-201524
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Wireless power transfer system is playing an important role in biomedical applications by transferring power wirelessly to the implants present in the body. In this paper, magnetic coupling resonance wireless power transfer system is used to recharge battery of implantable cardioverter defibrillator. The proposed system operates at 300 KHz frequency. The LCC-C compensation topology is used to tune the coils for operating the system at stable resonant frequency and to obtain higher efficiency. In this paper, we have investigated the performance of different coil structures in multiple layer configurations and we offer recommendations on the best suitable configuration for effective transfer of power.
引用
收藏
页码:703 / 714
页数:12
相关论文
共 50 条
  • [1] Design and development of wireless power transfer system for implantable cardioverter defibrillator (vol 65, pg 703, 2021)
    Aditya, Kuragayala
    Bobba, Phaneendra Babu
    Satyavani, Yadlapalli
    [J]. INTERNATIONAL JOURNAL OF APPLIED ELECTROMAGNETICS AND MECHANICS, 2021, 66 (01) : 203 - 203
  • [2] RECENT ADVANCES IN THE DEVELOPMENT OF A POWER SOURCE FOR THE IMPLANTABLE DEFIBRILLATOR CARDIOVERTER
    TAKEUCHI, ES
    MUFFOLETTO, BC
    GREENWOOD, JM
    HOLMES, CF
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (03): : 689 - 689
  • [3] Development of the Implantable Cardioverter-Defibrillator
    Maron, Barry J.
    Estes, N. A. Mark
    Rowin, Ethan J.
    Maron, Martin S.
    Reynolds, Matthew R.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2023, 82 (04) : 353 - 373
  • [4] THE IMPLANTABLE CARDIOVERTER/DEFIBRILLATOR
    MEINERTZ, T
    KUCK, KH
    [J]. HERZ, 1994, 19 (05) : 243 - 245
  • [5] The Design of Ultra Low Power SAR ADC for Implantable Cardioverter Defibrillator (ICD)
    Mohapatra, Satyajit
    Mohapatra, Nihar Ranjan
    [J]. 2020 33RD INTERNATIONAL CONFERENCE ON VLSI DESIGN AND 2020 19TH INTERNATIONAL CONFERENCE ON EMBEDDED SYSTEMS (VLSID), 2020, : 72 - 77
  • [6] The implantable cardioverter defibrillator
    Glikson, M
    Friedman, PA
    [J]. LANCET, 2001, 357 (9262): : 1107 - 1117
  • [7] Leadless implantable cardioverter defibrillator system for pediatrics
    Berul, CI
    Walsh, EP
    Alexander, ME
    Bevilacqua, LM
    Triedman, JK
    Dahlby, D
    Gilkerson, JO
    [J]. CIRCULATION, 2001, 104 (17) : 753 - 753
  • [8] The implantable cardioverter defibrillator
    Rinaldi, CA
    Gill, JS
    [J]. HOSPITAL MEDICINE, 2002, 63 (11): : 672 - 675
  • [9] THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
    FURMAN, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (02): : 249 - 249
  • [10] A completely subcutaneous implantable cardioverter defibrillator system functioning simultaneously with an endocardial implantable cardioverter defibrillator programmed as pacemaker
    van Opstal, Jurren
    Geskes, Gijs
    Debie, Luuk
    [J]. EUROPACE, 2011, 13 (01): : 141 - 142