Cadmium determination in biological samples by direct solid sampling flame atomic absorption spectrometry

被引:0
|
作者
Flores, TMD [1 ]
Paniz, JNG
Martins, AF
Dressler, VL
Müller, EI
da Costa, AB
机构
[1] Univ Fed Santa Maria, Dept Quim, BR-97105900 Santa Maria, RS, Brazil
[2] Univ Santa Cruz Sul, Dept Biol, BR-96815900 Santa Cruz Do Sul, RS, Brazil
关键词
direct solid analysis; flame atomic absorption spectrometry; cadmium determination; biological samples;
D O I
暂无
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A direct solid sampling flame atomic absorption spectrometric procedure for trace determination of cadmium in biological samples has been developed. Test samples (0.05-2.00 mg) were ground and weighed into small polyethylene vials, which were connected to the device for solid sample introduction into a conventional air/acetylene flame. Test samples were carried as a dry aerosol to a quartz cell, placed between the burner and the optical path, which had a perpendicular entrance and a slit in the upper part. The atomic vapor generated in the flame produced a transient signal that was totally integrated within 1 s. The effect of operating conditions and the extent of grinding on the analytical signal were evaluated. Background signals were always low and a characteristic mass of 0.29 ng Cd was obtained. Calibration was performed using different masses of solid certified reference materials. Results obtained for certified and in-house reference materials were typically within the 95% confidence interval of the certified and/or reference value, and the precision, expressed as relative standard deviation, was between 3.9 and 6.7%. The proposed system is simple and it might be adapted to conventional atomic absorption spectrometers allowing the determination of Cd in more than 80 test samples per hour, excluding weighing. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:2187 / 2193
页数:7
相关论文
共 50 条
  • [1] Use of paper capsules for cadmium determination in biological samples by solid sampling flame atomic absorption spectrometry
    da Costa, AB
    de Mattos, JCP
    Müller, EI
    Paniz, JNG
    Dressler, VL
    Flores, TMD
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2005, 60 (05) : 583 - 588
  • [2] Direct solid sampling by flame atomic absorption spectrometry:: Determination of manganese in coal samples
    Flores, EMM
    Paniz, JNG
    Saidelles, APF
    Barin, JS
    Dressler, VL
    Müller, EI
    Costa, AB
    JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2004, 15 (02) : 199 - 204
  • [3] Direct determination of cadmium in geological samples by slurry sampling electrothermal atomic absorption spectrometry
    Cui, Hao
    Guo, Wei
    Cheng, Mengting
    Zhang, Ping
    Jin, Lanlan
    Guo, Qinghai
    Hu, Shenghong
    ANALYTICAL METHODS, 2015, 7 (20) : 8970 - 8976
  • [4] Direct flame solid sampling for atomic absorption spectrometry:: determination of copper in bovine liver
    Flores, ÉMD
    da Costa, AB
    Barin, JS
    Dressler, VL
    Paniz, JNG
    Martins, AF
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2001, 56 (10) : 1875 - 1882
  • [5] Use of the median in the direct determination of cadmium in solid samples by electrothermal atomic absorption spectrometry
    Belarra, MA
    Lavilla, I
    Castillo, JR
    ANALYST, 1995, 120 (12) : 2813 - 2817
  • [6] Determination of cadmium and lead in fresh fish samples by direct sampling electrothermal atomic absorption spectrometry
    Zmozinski, Ariane V.
    Passos, Leonardo D.
    Damin, Isabel C. F.
    Espirito Santo, Maria Aparecida B.
    Vale, Maria G. R.
    Silva, Marcia M.
    ANALYTICAL METHODS, 2013, 5 (22) : 6416 - 6424
  • [7] Direct determination of cadmium in solid biological materials by slurry sampling electrothermal atomic absorption spectrometry with polytetrafluoroethylene as a chemical modifier
    Wang, FY
    Jiang, ZC
    ANALYTICA CHIMICA ACTA, 1999, 391 (01) : 89 - 94
  • [8] Direct cadmium determination in sediment samples by flame atomic absorption spectrometry using multivariate calibration procedures
    Flores, EMD
    Paniz, JNG
    Saidelles, APF
    Müller, EI
    da Costa, AB
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2003, 18 (07) : 769 - 774
  • [10] Determination of Lead and Manganese in Biological Samples and Sediment Using Slurry Sampling and Flame Atomic Absorption Spectrometry
    Vieira, Daniel Rodrigues
    Castro, Jacira Teixeira
    Lemos, Valfredo Azevedo
    JOURNAL OF AOAC INTERNATIONAL, 2011, 94 (02) : 645 - 649