摘要
A lack of standardized evaluation procedures for dopamine transporter (DAT) SPECT investigations impairs both intra- and interindividual comparisons as well as multicenter trials-for example, for assessment of disease progression or the response to various drug treatments. Therefore, the aim of this study was to evaluate a novel automated method, which has been specifically developed for a standardized quantification of N-(3-fluoropropyl)-2 beta-carbomethoxy-3 beta-(4-iodophenyl)nortro- pane (I-123-FP-CIT) SPECT studies. Methods: DAT binding ratios of 155 I-123-FP-CIT SPECT studies in 14 control subjects and 141 patients referred to confirm or exclude a presynaptic dopaminergic deficit were determined manually and by a fully automated technique. The latter included coregistration of patient studies to an I-123-FP-CIT mean template of controls with specialized, nonrigid adjustment for variation in striatal location, followed by calculation of specific striatal DAT binding using a standardized 3-dimensional volume-of-interest (VOI) map. The map is based on a MR scan and covers the striatum (S), caudate (C), putamen (P), and an occipital reference region. The semi-quantitative ratios of both methods were compared with the visual findings. Results: Excellent linear correlations were observed between manually and automatically determined results (S: r = 0.99; C: r = 0.99; P: r = 0.99; P < 0.001, respectively). Automated evaluations delivered highly reproducible and visually exact coregistrations. Individual variations in striatal anatomy (e.g., atrophy) were considered and VOI positions were corrected before quantification. Both the manual and the automated method showed identical accuracy in supporting the visual diagnoses. Conclusion: In a large patient population, excellent agreement was observed between quantitative DAT results using a time-consuming, observer-dependent, conventional manual method and the objective, automated evaluation specifically developed for a standardized evaluation of I-123-FP-CIT SPECT studies. It is suggested that the novel automated technique may substantially facilitate both intra- and interindividual comparisons as well as multicenter trials.
引用
收藏
页码:1109 / 1118
页数:10
相关论文
共 14 条
  • [1] Clinical testing of an optimized software solution for an automated, observer-independent evaluation of dopamine transporter SPECT studies.
    Koch, W
    Hamann, C
    Radau, P
    Tatsch, K
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (05) : 255P - 255P
  • [2] Objective, observer-independent evaluation of myocardial perfusion and function: role of SPECT
    Szot, Wojciech
    Kwiecien, Ewa
    Tekieli, Lukasz
    Czyz, Lukasz
    Borkowska, Eliza
    Dabrowski, Wladyslaw
    Drabik, Leszek
    Dabrowski, Maciej
    Kostkiewicz, Magdalena
    Musialek, Piotr
    POSTEPY W KARDIOLOGII INTERWENCYJNEJ, 2022, 18 (04): : 366 - 372
  • [3] Automated, objective software for 3D registration and analysis of dopamine transporter SPECT studies.
    Radau, P
    Koch, W
    Holtmannspoetter, M
    Poepperl, G
    Tatsch, K
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (05) : 12P - 12P
  • [4] Automated evaluation of dopamine D2 receptor SPECT studies: Benefit in clinical routine?
    Hamann, CS
    Koch, W
    Radau, P
    Tatsch, K
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (05) : 261P - 262P
  • [5] External evaluation of an automated quantification software for dopamine transporter PET without MRI
    Kang, Seung Kwan
    Kim, Daewoon
    Shin, Seong A.
    Choi, Hongyoon
    Kim, Yu Kyeong
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2023, 64
  • [6] 18F-FDG PET studies in patients with extratemporal and temporal epilepsy:: Evaluation of an observer-independent analysis
    Drzezga, A
    Arnold, S
    Minoshima, S
    Noachtar, S
    Szecsi, J
    Winkler, P
    Römer, W
    Tatsch, K
    Weber, W
    Bartenstein, P
    JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (05) : 737 - 746
  • [7] Optimized image acquisition for dopamine transporter imaging with ultra-high resolution clinical pinhole SPECT
    Chen, Yuan
    Vastenhouw, Brendan
    Wu, Chao
    Goorden, Marlies C.
    Beekman, Freek
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (22):
  • [8] Automated identification of uncertain cases in deep learning-based classification of dopamine transporter SPECT to improve clinical utility and acceptance
    Thomas Budenkotte
    Ivayla Apostolova
    Roland Opfer
    Julia Krüger
    Susanne Klutmann
    Ralph Buchert
    European Journal of Nuclear Medicine and Molecular Imaging, 2024, 51 : 1333 - 1344
  • [9] Automated identification of uncertain cases in deep learning-based classification of dopamine transporter SPECT to improve clinical utility and acceptance
    Budenkotte, Thomas
    Apostolova, Ivayla
    Opfer, Roland
    Krueger, Julia
    Klutmann, Susanne
    Buchert, Ralph
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 (05) : 1333 - 1344
  • [10] Experimental evaluation of absolute quantification in 99mTc-TRODAT-1 SPECT/CT brain dopamine transporter (DAT) studies
    Zainudin, Norasma Amira
    Zulkifli, Nadiah
    Hamid, Khadijah
    Hashim, Hazlin
    Mansor, Syahir
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (08):